Non-contact acoustic micro-tapping optical coherence elastography for evaluating biomechanical changes in the cornea following UV/riboflavin collagen cross linking: ex vivo human study
Authors:
Mitchell A. Kirby,
Ivan Pelivanov,
Gabriel Regnault,
John J. Pitre,
Ryan T. Wallace,
Matthew O'Donnell,
Ruikang Wang,
Tueng T. Shen
Abstract:
Purpose: To evaluate changes in the anisotropic elastic properties of ex vivo human cornea treated with UV cross-linking (CXL) using non-contact acoustic micro-tapping Optical Coherence Elastography (AuT-OCE) Design: AuT performed on normal and CXL ex vivo human donor cornea Methods: Elastic properties of normal and UV CXL treated human corneas were quantified using non-contact acoustic micro-tapp…
▽ More
Purpose: To evaluate changes in the anisotropic elastic properties of ex vivo human cornea treated with UV cross-linking (CXL) using non-contact acoustic micro-tapping Optical Coherence Elastography (AuT-OCE) Design: AuT performed on normal and CXL ex vivo human donor cornea Methods: Elastic properties of normal and UV CXL treated human corneas were quantified using non-contact acoustic micro-tapping Optical Coherence Elastography (AuT-OCE) Main Outcome Measures: Corneal elastic moduli (in-plane Young's, E, and out-of-plane shear, G) can be evaluated in both normal and CXL treated tissues, as well as during the CXL procedure using non-contact AuT-OCE. Results: CXL induced a significant increase in both the tensile and shear moduli in human cornea. The mean in the paired study (pre- and post-, n=7) of the in-plane Young's modulus, E=3u, increased from 19 MPa to 43 MPa while the out-of-plane shear modulus, G, increased from 188 kPa to 673 kPa. Mechanical tests in a subgroup support CXL-induced cornea moduli changes and generally agree with AuT-OCE. Conclusions: The human cornea is a highly anisotropic material where in-plane mechanical properties are very different from those out-of-plane. Non-contact AuT-OCE can measure changes in the anisotropic elastic properties in human cornea as a result of UV-CXL.
△ Less
Submitted 15 September, 2022; v1 submitted 28 June, 2022;
originally announced June 2022.
-
Delineating corneal elastic anisotropy in a porcine model using non-contact optical coherence elastography and ex vivo mechanical tests
Authors:
Mitchell A. Kirby,
John J. Pitre,
Hong-Cin Liou,
David S. Li,
Ruikang Wang,
Ivan Pelivanov,
Matthew O'Donnell,
Tueng T. Shen
Abstract:
Objective: To compare non-contact acoustic micro-tapping optical coherence elastography (AuT-OCE) with destructive mechanical tests to confirm corneal elastic anisotropy.
Design: Ex vivo, laboratory study with non-contact AuT-OCE followed by mechanical rheometry and extensometry.
Subjects: Inflated cornea of whole-globe porcine eyes.
Methods: A non-contact transducer was used to launch mecha…
▽ More
Objective: To compare non-contact acoustic micro-tapping optical coherence elastography (AuT-OCE) with destructive mechanical tests to confirm corneal elastic anisotropy.
Design: Ex vivo, laboratory study with non-contact AuT-OCE followed by mechanical rheometry and extensometry.
Subjects: Inflated cornea of whole-globe porcine eyes.
Methods: A non-contact transducer was used to launch mechanical waves in the cornea that were imaged with phase-sensitive OCT at physiologically relevant pressures. Reconstruction of both Young's modulus (E) and out-of-plane shear modulus (G) in the cornea from experimental data was performed using a model of a nearly incompressible transversally isotropic (NITI) medium. Samples were then excised and parallel plate rheometry was performed to measure the shear modulus G. Corneal samples were then subjected to strip extensomety to measure the Young's modulus.
Main Outcome Measures: Strong corneal anisotropy was confirmed with both AuT-OCE and mechanical tests, with the Young's and shear moduli differing by over an order of magnitude. These results show that AuT-OCE can quantify both moduli with a non-contact, non-invasive, clinically translatable technique.
△ Less
Submitted 27 May, 2021;
originally announced May 2021.
-
Nearly-incompressible transverse isotropy (NITI) of cornea elasticity: model and experiments with acoustic micro-tapping OCE
Authors:
John J Pitre Jr,
Mitchell A Kirby,
David S Li,
Tueng T Shen,
Ruikang K Wang,
Matthew O'Donnell,
Ivan Pelivanov
Abstract:
The cornea provides the largest refractive power for the human visual system. Its stiffness, along with intraocular pressure (IOP), are linked to several pathologies, including keratoconus and glaucoma. Although mechanical tests can quantify corneal elasticity ex vivo, they cannot be used clinically. Optical coherence elastography (OCE), which launches and tracks shear waves to estimate stiffness,…
▽ More
The cornea provides the largest refractive power for the human visual system. Its stiffness, along with intraocular pressure (IOP), are linked to several pathologies, including keratoconus and glaucoma. Although mechanical tests can quantify corneal elasticity ex vivo, they cannot be used clinically. Optical coherence elastography (OCE), which launches and tracks shear waves to estimate stiffness, provides an attractive non-contact probe of corneal elasticity. To date, however, OCE studies report corneal moduli around tens of kPa, orders-of-magnitude less than those (few MPa) obtained by tensile/inflation testing. This large discrepancy impedes OCE's clinical adoption. Based on corneal microstructure, we introduce and fully characterize a nearly-incompressible transversally isotropic (NITI) model depicting corneal biomechanics. We show that the cornea must be described by two shear moduli, contrary to current single-modulus models, decoupling tensile and shear responses. We measure both as a function of IOP in ex vivo porcine cornea, obtaining values consistent with both tensile and shear tests. At pressures above 30 mmHg, the model begins to fail, consistent with non-linear changes in cornea at high IOP.
△ Less
Submitted 18 June, 2020;
originally announced June 2020.