Traceable thermal imaging in harsh environments
Authors:
Jamie Luke McMillan
Abstract:
Despite being regarded as a well-established field, temperature measurement continues to pose significant challenges for many professionals in the metrology industry. Thermal imagers enable fast, non-contact and a full field measurement, however there is a lack of metrological development to support their use. Here, thermal imagers have been examined for the monitoring of special nuclear material…
▽ More
Despite being regarded as a well-established field, temperature measurement continues to pose significant challenges for many professionals in the metrology industry. Thermal imagers enable fast, non-contact and a full field measurement, however there is a lack of metrological development to support their use. Here, thermal imagers have been examined for the monitoring of special nuclear material containers; the surface temperature is an important parameter for store management decisions. Throughout this research: a selection of thermal imagers were calibrated and made traceable to the International Temperature Scale of 1990; laboratory observations of a proxy steel plate were made; initial measurement of nuclear material storage containers were made; then a deployment to an inactive store was demonstrated. For this technique to be feasible, uncertainties less than 10$^\circ$C would be required.
During the laboratory calibration of an uncooled and cooled thermal imager against blackbody reference sources, across the measured temperature range of 10$^\circ$C to 100$^\circ$C the uncertainties were less than 3.20$^\circ$C ($k=2$) and 0.50$^\circ$C ($k=2$) respectively. Here $k$ is the uncertainty coverage factor. When these calibrations were applied to the plate, regions of steel and higher emissivity coating were evaluated. These uncoated regions were measured with a thermal imager to demonstrate temperature differences compared to surface mounted thermocouples of 8.3$^\circ$C and uncertainties up to 30.1$^\circ$C ($k=2$). For the coated regions this temperature difference was reduced to 1.8$^\circ$C with uncertainties up to 6.8$^\circ$C ($k=2$).
△ Less
Submitted 10 February, 2023;
originally announced February 2023.
-
Thermal and dimensional evaluation of a test plate for assessing the measurement capability of a thermal imager within nuclear decommissioning storage
Authors:
Jamie Luke McMillan,
Michael Hayes,
Rob Hornby,
Sofia Korniliou,
Christopher Jones,
Daniel O'Connor,
Rob Simpson,
Graham Machin,
Robert Bernard,
Chris Gallagher
Abstract:
In this laboratory-based study, a plate was designed, manufactured and then characterised thermally and dimensionally using a thermal imager. This plate comprised a range of known scratch, dent, thinning and pitting artefacts as mimics of possible surface anomalies, as well as an arrangement of higher emissivity targets. The thermal and dimensional characterisation of this plate facilitated surfac…
▽ More
In this laboratory-based study, a plate was designed, manufactured and then characterised thermally and dimensionally using a thermal imager. This plate comprised a range of known scratch, dent, thinning and pitting artefacts as mimics of possible surface anomalies, as well as an arrangement of higher emissivity targets. The thermal and dimensional characterisation of this plate facilitated surface temperature determination. This was verified through thermal models and successful defect identification of the scratch and pitting artefacts at temperatures from \SIrange{30}{170}{\celsius}.
These laboratory measurements demonstrated the feasibility of deploying in-situ thermal imaging to the thermal and dimensional characterisation of special nuclear material containers. Surface temperature determination demonstrated uncertainties from \SIrange{1.0}{6.8}{\celsius} (\(k = 2\)). The principle challenges inhibiting successful deployment are a lack of suitable emissivity data and a robust defect identification algorithm suited to both static and transient datasets.
△ Less
Submitted 26 April, 2022;
originally announced April 2022.
-
Thermometry of intermediate level nuclear waste containers in multiple environmental conditions
Authors:
J Norman,
A Sposito,
J L McMillan,
W Bond,
M Hayes,
R Simpson,
G Sutton,
V Panicker,
G Machin,
J Jowsey,
A Adamska
Abstract:
Intermediate level nuclear waste must be stored until it is safe for permanent disposal. Temperature monitoring of waste packages is important to the nuclear decommissioning industry to support management of each package. Phosphor thermometry and thermal imaging have been used to monitor the temperature of intermediate level waste containers within the expected range of environmental storage condi…
▽ More
Intermediate level nuclear waste must be stored until it is safe for permanent disposal. Temperature monitoring of waste packages is important to the nuclear decommissioning industry to support management of each package. Phosphor thermometry and thermal imaging have been used to monitor the temperature of intermediate level waste containers within the expected range of environmental storage conditions at the Sellafield Ltd site: temperatures from 10 °C to 25 °C and relative humidities from 60 %rh to 90 %rh. The feasibility of determining internal temperature from external surface temperature measurement in the required range of environmental conditions has been demonstrated.
△ Less
Submitted 7 October, 2020;
originally announced October 2020.
-
Thermometry of intermediate level waste containers using phosphor thermometry and thermal imaging
Authors:
J L McMillan,
A Greenen,
W Bond,
M Hayes,
R Simpson,
G Sutton,
G Machin
Abstract:
Intermediate level waste containers are used for the storage of an assortment of radioactive waste. This waste is heat-generating and needs monitoring and so this work was undertaken to determine whether the mean internal container temperature can be inferred from the temperature of the vent. By using two independent thermometry techniques, phosphor thermometry and thermal imaging, the internal te…
▽ More
Intermediate level waste containers are used for the storage of an assortment of radioactive waste. This waste is heat-generating and needs monitoring and so this work was undertaken to determine whether the mean internal container temperature can be inferred from the temperature of the vent. By using two independent thermometry techniques, phosphor thermometry and thermal imaging, the internal temperature was demonstrated to be proportional to the vent temperature as measured by both methods. The correlation is linear and given suitable characterisation could provide robust indication of the internal bulk temperature.
△ Less
Submitted 20 July, 2018;
originally announced July 2018.