New Frontiers in Muon-Spin Spectroscopy Using Si-Pixel Detectors
Authors:
Heiko Augustin,
Niklaus Berger,
Andrin Doll,
Pascal Isenring,
Marius Köppel,
Jonas A. Krieger,
Hubertus Luetkens,
Lukas Mandok,
Thomas Prokscha,
Thomas Rudzki,
André Schöning,
Zaher Salman
Abstract:
The study of novel quantum materials relies on muon-spin rotation, relaxation, or resonance (\mSR) measurements. Yet, a fundamental limitation persists: many of these materials can only be synthesized in extremely small quantities, often at sub-millimeter scales. While \mSR ~offers unique insights into electronic and magnetic properties, existing spectrometers lack a sub-millimeter spatial resolut…
▽ More
The study of novel quantum materials relies on muon-spin rotation, relaxation, or resonance (\mSR) measurements. Yet, a fundamental limitation persists: many of these materials can only be synthesized in extremely small quantities, often at sub-millimeter scales. While \mSR ~offers unique insights into electronic and magnetic properties, existing spectrometers lack a sub-millimeter spatial resolution and the possibility of triggerless pump-probe data acquisition, which would enable more advanced measurements. The General Purpose Surface-muon instrument (GPS) at the Paul Scherrer Institute (PSI) is currently limited to a muon stopping rate of \SI{40}{\kilo\hertz} to \SI{120}{\kilo\hertz}, a constraint that will become more pressing with the upcoming High-Intensity Muon Beam (HIMB) project. To overcome these challenges, we demonstrate the feasibility of employing ultra-thin monolithic Si-pixel detectors to reconstruct the stopping position of muons within the sample, thereby significantly enhancing the capability of measuring at higher muon rate. Additionally, we explore the first steps toward a triggerless pump-probe \mSR ~measurement scheme. Unlike conventional pump-probe techniques that require external triggers, a triggerless readout system can continuously integrate stimuli pulses into the data stream, allowing real-time tracking of ultra-fast dynamics in quantum materials. This approach will enable the study of transient states, spin dynamics, and quantum coherence under external stimuli.
△ Less
Submitted 17 April, 2025;
originally announced April 2025.
-
Advanced muon-spin spectroscopy with high lateral resolution using Si-pixel detectors
Authors:
Lukas Mandok,
Pascal Isenring,
Heiko Augustin,
Marius Köppel,
Jonas A. Krieger,
Hubertus Luetkens,
Thomas Prokscha,
Thomas Rudzki,
André Schöning,
Zaher Salman
Abstract:
Muon-spin spectroscopy at continuous sources has stagnated at a stopped muons rate of ~40 kHz for the last few decades. The major limiting factor is the requirement of a single muon in the sample during the typical 10 μs data gate window. To overcome this limit and to be able to perform muon-spin relaxation (μSR) measurements on millimeter-sized samples, one can use vertex reconstruction methods t…
▽ More
Muon-spin spectroscopy at continuous sources has stagnated at a stopped muons rate of ~40 kHz for the last few decades. The major limiting factor is the requirement of a single muon in the sample during the typical 10 μs data gate window. To overcome this limit and to be able to perform muon-spin relaxation (μSR) measurements on millimeter-sized samples, one can use vertex reconstruction methods to construct μSR spectra. This is now possible thanks to the availability of very thin monolithic Si-pixel chips, which offer minimal particle scattering and high count rate. Here we present results from a Si-pixel based spectrometer that utilizes vertex reconstruction schemes for the incoming muons and emitted positrons. With this spectrometer we were able to obtain a first vertex reconstructed μSR (VR-μSR) spectrum. The unique capabilities and benefits of such a spectrometer are discussed.
△ Less
Submitted 11 March, 2025;
originally announced March 2025.
-
GermanIum Array for Non-destructive Testing (GIANT) setup for Muon Induced X-ray Emission (MIXE) at the Paul Scherrer Institute
Authors:
Lars Gerchow,
Sayani Biswas,
Gianluca Janka,
Carlos Vigo,
Andreas Knecht,
Stergiani Marina Vogiatzi,
Narongrit Ritjoho,
Thomas Prokscha,
Hubertus Luetkens,
Alex Amato
Abstract:
The usage of muonic X-rays to study elemental properties like nuclear radii ranges back to the seventies. This triggered the pioneering work at the Paul Scherrer Institute (PSI), during the eighties, on the Muon Induced X-ray Emission (MIXE) technique for a non-destructive assessment of elemental compositions. In the recent years, this method has seen a rebirth, improvement and adoption at most mu…
▽ More
The usage of muonic X-rays to study elemental properties like nuclear radii ranges back to the seventies. This triggered the pioneering work at the Paul Scherrer Institute (PSI), during the eighties, on the Muon Induced X-ray Emission (MIXE) technique for a non-destructive assessment of elemental compositions. In the recent years, this method has seen a rebirth, improvement and adoption at most muon facilities around the world. Hereby, the PSI offers unique capabilities with its high-rate continuous muon beam at the Swiss Muon Source (S$μ$S). We report here the decision making, construction and commissioning of a dedicated MIXE spectrometer at PSI, the GermanIum Array for Non-destructive Testing (GIANT) setup. Multiple campaigns highlighted the outstanding capabilities of MIXE at PSI, e.g. resolving down to 1 at% elemental concentrations with as little as 1 h data taking, measuring isotopic ratios for elements from iron to lead, and characterizing gamma rays induced by muon nuclear capture. On-target beamspots were characterized with a dedicated charged particle tracker to be 22.06$\pm$0.18 and 14.45$\pm$0.06 mm for 25 and 45 MeV/c, respectively. Advanced analysis of the High Purity Germanium (HPGe) signals further allows to improve energy and timing resolutions to ~1 keV and 20 ns at 1 MeV, respectively. Within the GIANT setup, an average detector has a photopeak efficiency of $\overline{ε_E}$ =0.11% and an energy resolution of $\overline{σ_E}$ = 0.8 keV at E=1000 keV. The overall performance of the GIANT setup at S$μ$S allowed to start a rich user program with archaeological samples, Li-ion battery research, and collaboration with industry. Future improvements will include a simulation based analysis and a higher degree of automation, e.g. automatic scans of a series of muon momenta and automatic sample changing.
△ Less
Submitted 28 October, 2022;
originally announced October 2022.
-
In-situ uniaxial pressure cell for X-ray and neutron scattering experiments
Authors:
G. Simutis,
A. Bollhalder,
M. Zolliker,
J. Küspert,
Q. Wang,
D. Das,
F. Van Leeuwen,
O. Ivashko,
O. Gutowski,
J. Philippe,
T. Kracht,
P. Glaevecke,
T. Adachi,
M. Von Zimmermann,
S. Van Petegem,
H. Luetkens,
Z. Guguchia,
J. Chang,
Y. Sassa,
M. Bartkowiak,
M. Janoschek
Abstract:
We present an in-situ uniaxial pressure device optimized for small angle X-ray and neutron scattering experiments at low-temperatures and high magnetic fields. A stepper motor generates force, which is transmitted to the sample via a rod with integrated transducer that continuously monitors the force. The device has been designed to generate forces up to 200 N in both compressive and tensile confi…
▽ More
We present an in-situ uniaxial pressure device optimized for small angle X-ray and neutron scattering experiments at low-temperatures and high magnetic fields. A stepper motor generates force, which is transmitted to the sample via a rod with integrated transducer that continuously monitors the force. The device has been designed to generate forces up to 200 N in both compressive and tensile configurations and a feedback control allows operating the system in a continuous-pressure mode as the temperature is changed. The uniaxial pressure device can be used for various instruments and multiple cryostats through simple and exchangeable adapters. It is compatible with multiple sample holders, which can be easily changed depending on the sample properties and the desired experiment and allow rapid sample changes.
△ Less
Submitted 26 July, 2022;
originally announced July 2022.
-
Characterization of a continuous muon source for the Muon-Induced X-ray Emission (MIXE) Technique
Authors:
Sayani Biswas,
Lars Gerchow,
Hubertus Luetkens,
Thomas Prokscha,
Aldo Antognini,
Niklaus Berger,
Thomas Elias Cocolios,
Rugard Dressler,
Paul Indelicato,
Klaus Jungmann,
Klaus Kirch,
Andreas Knecht,
Angela Papa,
Randolf Pohl,
Maxim Pospelov,
Elisa Rapisarda,
Peter Reiter,
Narongrit Ritjoho,
Stephanie Roccia,
Nathal Severijns,
Alexander Skawran,
Stergiani Marina Vogiatzi,
Frederik Wauters,
Lorenz Willmann,
Alex Amato
Abstract:
The toolbox for material characterization has never been richer than today. Great progress with all kinds of particles and interaction methods provide access to nearly all properties of an object under study. However, a tomographic analysis of the subsurface region remains still a challenge today. In this regard, the Muon-Induced X-ray Emission (MIXE) technique has seen rebirth fueled by the avail…
▽ More
The toolbox for material characterization has never been richer than today. Great progress with all kinds of particles and interaction methods provide access to nearly all properties of an object under study. However, a tomographic analysis of the subsurface region remains still a challenge today. In this regard, the Muon-Induced X-ray Emission (MIXE) technique has seen rebirth fueled by the availability of high intensity muon beams. We report here a study conducted at the Paul Scherrer Institute (PSI). It demonstrates that the absence of any beam time-structure leads to low pile-up events and a high signal-to-noise ratio (SNR) with less than one hour acquisition time per sample or data point. This performance creates the perspective to open this technique to a wider audience for the routine investigation of non-destructive and depth-sensitive elemental compositions, for example in rare and precious samples. Using a hetero-structured sample of known elements and thicknesses, we successfully detected the characteristic muonic X-rays, emitted during the capture of a negative muon by an atom, and the gamma-rays resulting from the nuclear capture of the muon, characterizing the capabilities of MIXE at PSI. This sample emphasizes the quality of a continuous beam, and the exceptional SNR at high rates. Such sensitivity will enable totally new statistically intense aspects in the field of MIXE, e.g. elemental 3D-tomography and chemical analysis. Therefore, we are currently advancing our proof-of-concept experiments with the goal of creating a full fledged permanently operated user station to make MIXE available to the wider scientific community as well as industry.
△ Less
Submitted 8 February, 2022;
originally announced February 2022.
-
Piezoelectric-driven uniaxial pressure cell for muon spin relaxation and neutron scattering experiments
Authors:
Shreenanda Ghosh,
Felix Brückner,
Artem Nikitin,
Vadim Grinenko,
Matthias Elender,
Andrew P. Mackenzie,
Hubertus Luetkens,
Hans-Henning Klauss,
Clifford W. Hicks
Abstract:
We present a piezoelectric-driven uniaxial pressure cell that is optimized for muon spin relaxation and neutron scattering experiments, and that is operable over a wide temperature range including cryogenic temperatures. To accommodate the large samples required for these measurement techniques, the cell is designed to generate forces up to 1000 N, and to minimize the background signal the space a…
▽ More
We present a piezoelectric-driven uniaxial pressure cell that is optimized for muon spin relaxation and neutron scattering experiments, and that is operable over a wide temperature range including cryogenic temperatures. To accommodate the large samples required for these measurement techniques, the cell is designed to generate forces up to 1000 N, and to minimize the background signal the space around the sample is kept as open as possible. We demonstrate here that by mounting plate-like samples with epoxy, a uniaxial stress exceeding 1 GPa can be achieved in an active volume of 5 mm3. We show that for practical operation it is important to monitor both the force and displacement applied to the sample. Also, because time is critical during facility experiments, samples are mounted in detachable holders that can be rapidly exchanged. The piezoelectric actuators are likewise contained in an exchangeable cartridge.
△ Less
Submitted 12 October, 2020;
originally announced October 2020.
-
The new versatile general purpose surface-muon instrument (GPS) based on silicon photomultipliers for $μ$SR measurements on a continuous-wave beam
Authors:
A. Amato,
H. Luetkens,
K. Sedlak,
A. Stoykov,
R. Scheuermann,
M. Elender,
A. Raselli,
D. Graf
Abstract:
We report on the design and commissioning of a new spectrometer for muon-spin relaxation/rotation studies installed at the Swiss Muon Source (S$μ$S) of the Paul Scherrer Institute (PSI, Switzerland). This new instrument is essentially a new design and replaces the old general-purpose surface-muon instrument (GPS) which has been for long the workhorse of the $μ$SR user facility at PSI. By making us…
▽ More
We report on the design and commissioning of a new spectrometer for muon-spin relaxation/rotation studies installed at the Swiss Muon Source (S$μ$S) of the Paul Scherrer Institute (PSI, Switzerland). This new instrument is essentially a new design and replaces the old general-purpose surface-muon instrument (GPS) which has been for long the workhorse of the $μ$SR user facility at PSI. By making use of muon and positron detectors made of plastic scintillators read out by silicon photomultipliers (SiPMs), a time resolution of the complete instrument of about 160 ps (standard deviation) could be achieved. In addition, the absence of light guides, which are needed in traditionally built $μ$SR instrument to deliver the scintillation light to photomultiplier tubes located outside magnetic fields applied, allowed us to design a compact instrument with a detector set covering an increased solid angle compared to the old GPS.
△ Less
Submitted 2 October, 2017; v1 submitted 30 May, 2017;