High-pressure Ion Trap
Authors:
Evgeny V. Krylov
Abstract:
A High-Pressure Ion Trap operating at pressure ~1 Torr is a core component of the portable hand-held mass-spectrometric gas analyzer. A comprehensive mathematical model of the HPIT is described in this paper. The influence of the instrumental parameters (gas composition, pressure, and temperature; applied voltages; ion trap size and geometry) and ion properties (mass, diffusion, and mobility) on t…
▽ More
A High-Pressure Ion Trap operating at pressure ~1 Torr is a core component of the portable hand-held mass-spectrometric gas analyzer. A comprehensive mathematical model of the HPIT is described in this paper. The influence of the instrumental parameters (gas composition, pressure, and temperature; applied voltages; ion trap size and geometry) and ion properties (mass, diffusion, and mobility) on the ion trap analytical parameters (mass-spectral peak position, height, and width) is examined. The model explained the difference between a high-pressure and regular low-pressure ion trap.
△ Less
Submitted 31 March, 2025;
originally announced April 2025.
Ion Trap Geometry
Authors:
Evgeny V Krylov
Abstract:
This work aims to find ion trap geometry for a high-quality ion trap mass analyzer that minimizes mass shift and is easy to fabricate. The theoretical procedure of the ion trap mass shift evaluation is developed, which provides a quantitative criterion of the mass shift (ion ejection delay in comparison with ideal ion trap) for the correct comparison of the different ion trap geometries. This proc…
▽ More
This work aims to find ion trap geometry for a high-quality ion trap mass analyzer that minimizes mass shift and is easy to fabricate. The theoretical procedure of the ion trap mass shift evaluation is developed, which provides a quantitative criterion of the mass shift (ion ejection delay in comparison with ideal ion trap) for the correct comparison of the different ion trap geometries. This procedure is validated by comparison with known experimental facts about existing mass analyzers: 3D hyperbolic ion trap (Finnigan), 2D hyperbolic ion trap (Thermo), Rectilinear Ion Trap (Purdue), and quadrupole mass filter. The developed theoretical procedure optimizes known ion trap geometries for low mass shift. Applying this procedure to various ion trap geometries reveals that in terms of the mass shift existing ion traps are close to the optimum in a framework of chosen geometry. Alternative ion trap geometry of very low mass shift, simple in fabrication, and satisfying the requirement of high mechanical accuracy is proposed. The proposed alternative ion trap geometry provides a lower ejection delay than linear ion traps with hyperbolic rods.
△ Less
Submitted 20 January, 2025;
originally announced January 2025.
A gapless micro-dielectric-barrier-discharge ion source for analytical applications
Authors:
Stephen L. Coy,
Evgeny V. Krylov,
Gary A. Eiceman,
Isik Kanik
Abstract:
Use of dielectric barrier discharge (DBD) as an ion source for sensitive chemical analysis is uncommon because barrier discharges generate excess noise due to spatial and temporal instability. This design uses contacted, crossed glass-coated micro-wires to focus the field into a gradually vanishing gap, suppressing spatial and temporal variability, reducing pressure, temperature, and humidity effe…
▽ More
Use of dielectric barrier discharge (DBD) as an ion source for sensitive chemical analysis is uncommon because barrier discharges generate excess noise due to spatial and temporal instability. This design uses contacted, crossed glass-coated micro-wires to focus the field into a gradually vanishing gap, suppressing spatial and temporal variability, reducing pressure, temperature, and humidity effects, stabilizing discharge initiation and limiting chemical fragmentation. Positive-ion-mode proton transfer, chemical fragmentation from a micro-discharge, and NO+ adducts combine to allow broad chemical sensitivity. We analyze noise properties of the ion source and report chemical responsivity for a wide range of volatile organic compounds. Source noise spectral density is compared for three systems: the contacted coated wires source, a gapped dielectric barrier discharge source, and a 5 mCi Ni-63 radioactive source. The crossed-wires source shows noise properties approaching those of the white-noise Ni-63 source, while gapped discharge exhibits 1/f noise from area-discharge random path and intensity variations. For chemical sensitivity testing, dilute samples are delivered by vapor flow injection or by gas chromatography, and then detected by differential ion mobility spectrometry (DMS / FAIMS) and, in a few cases by mass spectrometry. The compounds tested in positive ion mode include ketones and alcohols, simple aromatics (benzene, toluene, xylenes, ethyl and propyl benzene), chlorinated and nitrated solvents and aliphatics (hexanes and n-octane), with sensitivities from ppt to ppb levels. The wires source for trace vapor detection is stable under a range of environmental conditions from very low (ppb) humidity levels upward, has wider chemical coverage than the radioactive nickel source with nearly equivalent noise properties, and is adjustable to higher intensity.
△ Less
Submitted 19 February, 2016;
originally announced February 2016.