Investigation of Novel Preclinical Total Body PET Designed With J-PET Technology:A Simulation Study
Authors:
M. Dadgar,
S. Parzych,
F. Tayefi Ardebili,
J. Baran,
N. Chug,
C. Curceanu,
E. Czerwinski,
K. Dulski,
K. Eliyan,
A. Gajos,
B. C. Hiesmayr,
K. Kacprzak,
L. Kaplon,
K. Klimaszewski,
P. Konieczka,
G. Korcyl,
T. Kozik,
W. Krzemien,
D. Kumar,
S. Niedzwiecki,
D. Panek,
E. Perez del Rio,
L. Raczynski,
S. Sharma,
Shivani
, et al. (7 additional authors not shown)
Abstract:
The growing interest in human-grade total body positron emission tomography (PET) systems has also application in small animal research. Due to the existing limitations in human-based studies involving drug development and novel treatment monitoring, animal-based research became a necessary step for testing and protocol preparation. In this simulation-based study two unconventional, cost-effective…
▽ More
The growing interest in human-grade total body positron emission tomography (PET) systems has also application in small animal research. Due to the existing limitations in human-based studies involving drug development and novel treatment monitoring, animal-based research became a necessary step for testing and protocol preparation. In this simulation-based study two unconventional, cost-effective small animal total body PET scanners (for mouse and rat studies) have been investigated in order to inspect their feasibility for preclinical research. They were designed with the novel technology explored by the Jagiellonian-PET (J-PET) Collaboration. Two main PET characteristics: sensitivity and spatial resolution were mainly inspected to evaluate their performance. Moreover, the impact of the scintillator dimension and time-of-flight on the latter parameter was examined in order to design the most efficient tomographs. The presented results show that for mouse TB J-PET the achievable system sensitivity is equal to 2.35% and volumetric spatial resolution to 9.46 +- 0.54 mm3, while for rat TB J-PET they are equal to 2.6% and 14.11 +- 0.80 mm3, respectively. Furthermore, it was shown that the designed tomographs are almost parallax-free systems, hence, they resolve the problem of the acceptance criterion tradeoff between enhancing spatial resolution and reducing sensitivity.
△ Less
Submitted 6 August, 2024; v1 submitted 1 August, 2024;
originally announced August 2024.
-
Feasibility studies for imaging e$^{+}$e$^{-}$ annihilation with modular multi-strip detectors
Authors:
S. Sharma,
L. Povolo,
S. Mariazzi,
G. Korcyl,
K. Kacprzak,
D. Kumar,
S. Niedzwiecki,
J. Baran,
E. Beyene,
R. S. Brusa,
R. Caravita,
N. Chug,
A. Coussat,
C. Curceanu,
E. Czerwinski,
M. Dadgar,
M. Das,
K. Dulski,
K. Eliyan,
A. Gajos,
N. Gupta,
B. C. Hiesmayr,
L. Kaplon,
T. Kaplanoglu,
K. Klimaszewski
, et al. (19 additional authors not shown)
Abstract:
Studies based on imaging the annihilation of the electron (e$^{-}$) and its antiparticle positron (e$^{+}$) open up several interesting applications in nuclear medicine and fundamental research. The annihilation process involves both the direct conversion of e$^{+}$e$^{-}$ into photons and the formation of their atomically bound state, the positronium atom (Ps), which can be used as a probe for fu…
▽ More
Studies based on imaging the annihilation of the electron (e$^{-}$) and its antiparticle positron (e$^{+}$) open up several interesting applications in nuclear medicine and fundamental research. The annihilation process involves both the direct conversion of e$^{+}$e$^{-}$ into photons and the formation of their atomically bound state, the positronium atom (Ps), which can be used as a probe for fundamental studies. With the ability to produce large quantities of Ps, manipulate them in long-lived Ps states, and image their annihilations after a free fall or after passing through atomic interferometers, this purely leptonic antimatter system can be used to perform inertial sensing studies in view of a direct test of Einstein equivalence principle. It is envisioned that modular multistrip detectors can be exploited as potential detection units for this kind of studies. In this work, we report the results of the first feasibility study performed on a e$^{+}$ beamline using two detection modules to evaluate their reconstruction performance and spatial resolution for imaging e$^{+}$e$^{-}$ annihilations and thus their applicability for gravitational studies of Ps.
△ Less
Submitted 12 September, 2023;
originally announced September 2023.
-
Comparative studies of the sensitivities of sparse and full geometries of Total-Body PET scanners built from crystals and plastic scintillators
Authors:
Meysam Dadgar,
Szymon Parzych,
Jakub Baran,
Neha Chug,
Catalina Curceanu,
Eryk Czerwiński,
Kamil Dulski,
Kavya Elyan,
Aleksander Gajos,
Beatrix Hiesmayr,
Łukasz Kapłon,
Konrad Klimaszewski,
Paweł Konieczka,
Grzegorz Korcyl,
Tomasz Kozik,
Wojciech Krzemień,
Deepak Kumar,
Szymon Niedźwiecki,
Domonik Panek,
Eleną Perez del Rio,
Lech Raczyński,
Sushil Sharma,
Shivani,
Roman Shopa,
Magdalena Skurzok
, et al. (6 additional authors not shown)
Abstract:
Background: Total-Body imaging offers high sensitivity, single-bed position, and low dose, but high construction costs limit worldwide utilization. This study compares existing and developing tomographs using plastic scintillators via simulations to propose a cost-efficient Total-Body PET scanner.
Methods: Simulations of eight uEXPLORER tomographs with different scintillator materials, axial fie…
▽ More
Background: Total-Body imaging offers high sensitivity, single-bed position, and low dose, but high construction costs limit worldwide utilization. This study compares existing and developing tomographs using plastic scintillators via simulations to propose a cost-efficient Total-Body PET scanner.
Methods: Simulations of eight uEXPLORER tomographs with different scintillator materials, axial field-of-view, and detector configuration, and eight J-PET scanners with various field-of-view, plastic scintillator cross-sections, and layers were performed. Biograph Vision was also simulated. Two types of simulations were conducted with a central source and a water-filled phantom.
Results: BGO crystal-based scanners showed the best sensitivity (350 cps/kBq at the center). Sparse geometry or LYSO crystals reduced sensitivity. J-PET design showed similar sensitivity to sparse LYSO detectors, with full body coverage and additional gain for brain imaging.
Conclusion: The J-PET tomography system using plastic scintillators could be a cost-efficient alternative for Total-Body PET scanners, overcoming high construction costs while maintaining sensitivity
△ Less
Submitted 12 April, 2023;
originally announced April 2023.
-
Feasibility of the J-PET to monitor range of therapeutic proton beams
Authors:
Jakub Baran,
Damian Borys,
Karol Brzeziński,
Jan Gajewski,
Michał Silarski,
Neha Chug,
Aurélien Coussat,
Eryk Czerwiński,
Meysam Dadgar,
Kamil Dulski,
Kavya V. Eliyan,
Aleksander Gajos Krzysztof Kacprzak,
Łukasz Kapłon,
Konrad Klimaszewski,
Paweł Konieczka,
Renata Kopeć,
Grzegorz Korcyl,
Tomasz Kozik,
Wojciech Krzemień,
Deepak Kumar,
Antony J. Lomax,
Keegan McNamara,
Szymon Niedźwiecki,
Paweł Olko,
Dominik Panek
, et al. (18 additional authors not shown)
Abstract:
Objective: The aim of this work is to investigate the feasibility of the Jagiellonian Positron Emission Tomography (J-PET) scanner for intra-treatment proton beam range monitoring. Approach: The Monte Carlo simulation studies with GATE and PET image reconstruction with CASToR were performed in order to compare six J-PET scanner geometries (three dual-heads and three cylindrical). We simulated prot…
▽ More
Objective: The aim of this work is to investigate the feasibility of the Jagiellonian Positron Emission Tomography (J-PET) scanner for intra-treatment proton beam range monitoring. Approach: The Monte Carlo simulation studies with GATE and PET image reconstruction with CASToR were performed in order to compare six J-PET scanner geometries (three dual-heads and three cylindrical). We simulated proton irradiation of a PMMA phantom with a Single Pencil Beam (SPB) and Spread-Out Bragg Peak (SOBP) of various ranges. The sensitivity and precision of each scanner were calculated, and considering the setup's cost-effectiveness, we indicated potentially optimal geometries for the J-PET scanner prototype dedicated to the proton beam range assessment. Main results: The investigations indicate that the double-layer cylindrical and triple-layer double-head configurations are the most promising for clinical application. We found that the scanner sensitivity is of the order of 10$^{-5}$ coincidences per primary proton, while the precision of the range assessment for both SPB and SOBP irradiation plans was found below 1 mm. Among the scanners with the same number of detector modules, the best results are found for the triple-layer dual-head geometry. Significance: We performed simulation studies demonstrating that the feasibility of the J-PET detector for PET-based proton beam therapy range monitoring is possible with reasonable sensitivity and precision enabling its pre-clinical tests in the clinical proton therapy environment. Considering the sensitivity, precision and cost-effectiveness, the double-layer cylindrical and triple-layer dual-head J-PET geometry configurations seem promising for the future clinical application. Experimental tests are needed to confirm these findings.
△ Less
Submitted 28 February, 2023;
originally announced February 2023.
-
J-PET detection modules based on plastic scintillators for performing studies with positron and positronium beams
Authors:
S. Sharma,
J. Baran,
R. S. Brusa,
R. Caravita,
N. Chug,
A. Coussat,
C. Curceanu,
E. Czerwinski,
M. Dadgar,
K. Dulski,
K. Eliyan,
A. Gajos,
B. C. Hiesmayr,
K. Kacprzak,
L. Kaplon,
K. Klimaszewski,
P. Konieczka,
G. Korcyl,
T. Kozik,
W. Krzemien,
D. Kumar,
S. Mariazzi,
S. Niedzwiecki,
L. Panasa,
S. Parzych
, et al. (11 additional authors not shown)
Abstract:
The J-PET detector, which consists of inexpensive plastic scintillators, has demonstrated its potential in the study of fundamental physics. In recent years, a prototype with 192 plastic scintillators arranged in 3 layers has been optimized for the study of positronium decays. This allows performing precision tests of discrete symmetries (C, P, T) in the decays of positronium atoms. Moreover, than…
▽ More
The J-PET detector, which consists of inexpensive plastic scintillators, has demonstrated its potential in the study of fundamental physics. In recent years, a prototype with 192 plastic scintillators arranged in 3 layers has been optimized for the study of positronium decays. This allows performing precision tests of discrete symmetries (C, P, T) in the decays of positronium atoms. Moreover, thanks to the possibility of measuring the polarization direction of the photon based on Compton scattering, the predicted entanglement between the linear polarization of annihilation photons in positronium decays can also be studied. Recently, a new J-PET prototype was commissioned, based on a modular design of detection units. Each module consists of 13 plastic scintillators and can be used as a stand-alone, compact and portable detection unit. In this paper, the main features of the J-PET detector, the modular prototype and their applications for possible studies with positron and positronium beams are discussed. Preliminary results of the first test experiment performed on two detection units in the continuous positron beam recently developed at the Antimatter Laboratory (AML) of Trento are also reported.
△ Less
Submitted 7 January, 2023;
originally announced January 2023.