The new MRTOF mass spectrograph following the ZeroDegree spectrometer at RIKEN's RIBF facility
Authors:
M. Rosenbusch,
M. Wada,
S. Chen,
A. Takamine,
S. Iimura,
D. Hou,
W. Xian,
S. Yan,
P. Schury,
Y. Hirayama,
Y. Ito,
H. Ishiyama,
S. Kimura,
T. Kojima,
J. Lee,
J. Liu,
S. Michimasa,
H. Miyatake,
M. Mukai,
J. Y. Moon,
S. Nishimura,
S. Naimi,
T. Niwase,
T. Sonoda,
Y. X. Watanabe
, et al. (1 additional authors not shown)
Abstract:
A newly assembled multi-reflection time-of-flight mass spectrograph (MRTOF-MS) at RIKEN's RIBF facility became operational for the first time in spring 2020; further modifications and performance tests using stable ions were completed in early 2021. By using a pulsed-drift-tube technique to modify the ions' kinetic energy in a wide range, we directly characterize the dispersion function of the sys…
▽ More
A newly assembled multi-reflection time-of-flight mass spectrograph (MRTOF-MS) at RIKEN's RIBF facility became operational for the first time in spring 2020; further modifications and performance tests using stable ions were completed in early 2021. By using a pulsed-drift-tube technique to modify the ions' kinetic energy in a wide range, we directly characterize the dispersion function of the system for use in a new procedure for optimizing the voltages applied to the electrostatic mirrors. Thus far, a mass resolving power of $R_m > 1\,000\,000$ is reached within a total time-of-flight of only $12.5\,\mathrm{ms}$, making the spectrometer capable of studying short-lived nuclei possessing low-lying isomers. Detailed information about the setup and measurement procedure is reported, and an alternative in-MRTOF ion selection scheme to remove molecular contaminants in the absence of a dedicated deflection device is introduced. The setup underwent an initial on-line commissioning at the BigRIPS facility at the end of 2020, where more than 70 nuclear masses have been measured. A summary of the commissioning experiments and results from a test of mass accuracy will be presented.
△ Less
Submitted 2 November, 2022; v1 submitted 22 October, 2021;
originally announced October 2021.
Accurately accounting for effects on times-of-flight caused by finite field-transition times during the ejection of ions from a storage trap: A study for TOF and MRTOF mass spectrometry
Authors:
M. Rosenbusch,
P. Schury,
M. Wada,
S. Iimura,
Y. Ito,
H. Wollnik
Abstract:
In applied forms of time-of-flight mass spectrometry utilizing ion storage devices prior to an analysis device, a non instantaneous electric ejection pulse applied in the region of ion storage is used to accelerate ions into the time-of-flight analyzer. The calculated mass value of the ions from the time-of-flight is dependent on the duration of the field transition up to full strength. For novel…
▽ More
In applied forms of time-of-flight mass spectrometry utilizing ion storage devices prior to an analysis device, a non instantaneous electric ejection pulse applied in the region of ion storage is used to accelerate ions into the time-of-flight analyzer. The calculated mass value of the ions from the time-of-flight is dependent on the duration of the field transition up to full strength. For novel applications dedicated to precision measurements, such as multi-reflection time-of-flight mass spectrometry of short-lived isotopes, the goal is to continuously decrease the measurement uncertainty while providing a mass accuracy on the same order. Even though dynamic-field models for time-of-flight mass spectrometry have been considered in the past for technological advances, it is important to study the accuracy of the measured mass in this context. Using a simplified linear model for the field transition, we provide a basic investigation of the scenario, and discuss the deviation from the classical "mass-over-charge" dependency of the ions' time-of-flight, which becomes violated. The emerging mass discrepancy depends on the distance between the mass of the ion used for calibration and that of the ion of interest and, in extreme cases, can increase to about one percent for systems with short times-of-flight. However, for typical conditions in single-reference multi-reflection time-of-flight mass spectrometry, mass deviations caused by this effect typically remain below the 1 ppm level. If a mass calibration using two or more ion species is possible during the measurement, the effect becomes negligible for appropriate choices of reference masses.
△ Less
Submitted 27 April, 2020; v1 submitted 22 October, 2019;
originally announced October 2019.