-
Reducing Artifacts in Grating Interferometry Using Multiple Harmonics and Phase Step Corrections
Authors:
Hunter C. Meyer,
Conner B. Dooley,
Victoria L. Fontenot,
Kyungmin Ham,
Leslie G. Butler,
Alexandra Noel,
Joyoni Dey
Abstract:
X-ray interferometry is an emerging imaging modality with a wide variety of potential clinical applications, including lung and breast imaging, as well as in non-destructive testing, such as additive manufacturing and porosimetry. A grating interferometer uses a diffraction grating to produce a periodic interference pattern and measures how a patient or sample perturbs the pattern, producing three…
▽ More
X-ray interferometry is an emerging imaging modality with a wide variety of potential clinical applications, including lung and breast imaging, as well as in non-destructive testing, such as additive manufacturing and porosimetry. A grating interferometer uses a diffraction grating to produce a periodic interference pattern and measures how a patient or sample perturbs the pattern, producing three unique images that highlight X-ray absorption, refraction, and small angle scattering, known as the transmission, differential-phase, and dark-field images, respectively. Image artifacts that are unique to X-ray interferometry are introduced when assuming the fringe pattern is perfectly sinusoidal and the phase steps are evenly spaced. Inaccuracies in grating position, coupled with multi-harmonic fringes, lead to remnant oscillations and phase wraparound artifacts. We have developed an image recovery algorithm that uses additional harmonics, direct relative phase fitting, and phase step corrections to prevent them. The direct relative phase fitting removes the phase wraparound artifact. Correcting the phase step positions and introducing the additional harmonic removes the grating remnant artifact present in the transmission, differential-phase, and dark-field images. By modifying existing algorithms, the fit to the fringe pattern is greatly improved and artifacts are minimized, as we demonstrate with the imaging of several samples, including PMMA microspheres, ex vivo formalin fixed mouse lungs, and porous alumina.
△ Less
Submitted 29 March, 2025;
originally announced March 2025.
-
X-ray Interferometry Using a Modulated Phase Grating: Theory and Experiments
Authors:
Hunter Meyer,
Joyoni Dey,
Sydney Carr,
Kyungmin Ham,
Leslie G. Butler,
Kerry M. Dooley,
Ivan Hidrovo,
Markus Bleuel,
Tamas Varga,
Joachim Schulz,
Thomas Beckenbach,
Konradin Kaiser
Abstract:
X-ray grating interferometry allows for the simultaneous acquisition of attenuation, differential-phase contrast, and dark-field images, resulting from X-ray attenuation, refraction, and small-angle scattering, respectively. The modulated phase grating (MPG) interferometer is a recently developed grating interferometry system capable of generating a directly resolvable interference pattern using a…
▽ More
X-ray grating interferometry allows for the simultaneous acquisition of attenuation, differential-phase contrast, and dark-field images, resulting from X-ray attenuation, refraction, and small-angle scattering, respectively. The modulated phase grating (MPG) interferometer is a recently developed grating interferometry system capable of generating a directly resolvable interference pattern using a relatively large period grating envelope function that is sampled at a pitch that allows for X-ray spatial coherence using a microfocus X-ray source or by use of a source G0 grating that follows the Lau condition. We present the theory of the MPG interferometry system for a 2-dimensional staggered grating, derived using Fourier optics, and we compare the theoretical predictions with experiments we have performed with a microfocus X-ray system at Pennington Biomedical Research Center, LSU. The theoretical and experimental fringe visibility is evaluated as a function of grating-to-detector distance. Quantitative experiments are performed with porous carbon and alumina samples, and qualitative analysis of attenuation and dark-field images of a dried anchovy are shown.
△ Less
Submitted 22 April, 2024;
originally announced April 2024.
-
Neutron Interferometry Using a Single Modulated Phase Grating
Authors:
I. J. Hidrovo,
J. Dey,
H. Meyer,
D. S. Hussey,
N. N. Klimov,
L. G. Butler,
K. Ham,
W. Newhauser
Abstract:
Neutron grating interferometry provides information on phase and small-angle scatter in addition to attenuation. Previously, phase grating moiré interferometers (PGMI) with two- or three-phase gratings have been developed. These phase-grating systems use the moiré far-field technique to avoid the need for high-aspect absorption gratings used in Talbot-Lau interferometers (TLI) which reduce the neu…
▽ More
Neutron grating interferometry provides information on phase and small-angle scatter in addition to attenuation. Previously, phase grating moiré interferometers (PGMI) with two- or three-phase gratings have been developed. These phase-grating systems use the moiré far-field technique to avoid the need for high-aspect absorption gratings used in Talbot-Lau interferometers (TLI) which reduce the neutron flux reaching the detector. We first demonstrate through theory and simulations a novel phase grating interferometer system for cold neutrons that requires a single modulated phase grating (MPG) for phase-contrast imaging, as opposed to the two- or three-phase gratings in previously employed PGMI systems. The MPG theory was compared to the full Sommerfeld-Rayleigh Diffraction integral simulator. Then we compared the MPG system to experiments in the literature that use a two-phase-grating-based PGMI with best-case visibility of around 39%. An MPG with a modulation period of 300 micron, pitch of 2 micron, and grating heights with a phase modulation of (pi,0), illuminated by a monochromatic beam, produces a visibility of 94.2% with comparable source-to-detector distance (SDD) as the two-phase-grating-based PGMI. Phase sensitivity, another important performance metric of the grating interferometer, was compared to values available in the literature, viz. the conventional TLI with phase sensitivity of 4.5 x 10E+3 for an SDD of 3.5 m and a beam wavelength of 0.44 nm. For a range of modulation periods, the MPG system provides comparable or greater theoretical maximum phase sensitivity of 4.1 x 10E+3 to 10.0 x 10E+3 for SDDs of up to 3.5 m. This proposed MPG system can provide high-performance PGMI that obviates the need to align two phase gratings.
△ Less
Submitted 18 April, 2023; v1 submitted 22 June, 2022;
originally announced June 2022.
-
Dose-response curve of EBT, EBT2, and EBT3 radiochromic films to synchrotron-produced monochromatic x-ray beams
Authors:
Thomas A. D. Brown,
Kenneth R. Hogstrom,
Diane Alvarez,
Kenneth L. Matthews II,
Kyungmin Ham,
Joseph P. Dugas
Abstract:
This work investigates the dose-response curves of GAFCHROMIC EBT, EBT2, and EBT3 radiochromic films using synchrotron-produced monochromatic x-ray beams. EBT2 film is being utilized for dose verification in photoactivated Auger electron therapy at the Louisiana State University CAMD synchrotron facility. Monochromatic beams of 25, 30, and 35 keV were generated on the tomography beamline at CAMD.…
▽ More
This work investigates the dose-response curves of GAFCHROMIC EBT, EBT2, and EBT3 radiochromic films using synchrotron-produced monochromatic x-ray beams. EBT2 film is being utilized for dose verification in photoactivated Auger electron therapy at the Louisiana State University CAMD synchrotron facility. Monochromatic beams of 25, 30, and 35 keV were generated on the tomography beamline at CAMD. Ion chamber depth-dose measurements were used to determine the dose delivered to films irradiated at depths from 0.7 to 8.5 cm in a 10x10x10-cm3 PMMA phantom. AAPM TG-61 protocol was applied to convert measured ionization into dose. Films were digitized using an Epson 1680 Professional flatbed scanner and analyzed using the net optical density (NOD) derived from the red channel. A dose-response curve was obtained at 35 keV for EBT film, and at 25, 30, and 35 keV for EBT2 and EBT3 films. Calibrations of films for 4 MV x-rays were obtained for comparison using a radiotherapy accelerator at Mary Bird Perkins Cancer Center. The sensitivity (NOD per unit dose) of EBT film at 35 keV relative to that for 4-MV x-rays was 0.73 and 0.76 for doses 50 and 100 cGy, respectively. The sensitivity of EBT2 film at 25, 30, and 35 keV relative to that for 4-MV x-rays varied from 1.09 - 1.07, 1.23 - 1.17, and 1.27 - 1.19 for doses 50 - 200 cGy, respectively. For EBT3 film the relative sensitivity was within 3% of unity for all three monochromatic x-ray beams. EBT and EBT2 film sensitivity showed strong energy dependence over an energy range of 25 keV - 4 MV, although this dependence becomes weaker for larger doses. EBT3 film shows weak energy dependence, indicating that it would be a better dosimeter for kV x-ray beams where beam hardening effects can result in large changes in the effective energy.
△ Less
Submitted 4 December, 2012;
originally announced December 2012.
-
Verification of TG-61 dose for synchrotron-produced monochromatic x-ray beams using fluence-normalized MCNP5 calculations
Authors:
Thomas A. D. Brown,
Kenneth R. Hogstrom,
Diane Alvarez,
Kenneth L. Matthews II,
Kyungmin Ham
Abstract:
Ion chamber dosimetry is being used to calibrate dose for cell irradiations designed to investigate photoactivated Auger electron therapy at the Louisiana State University CAMD synchrotron facility. This study performed a dosimetry intercomparison for synchrotron-produced monochromatic x-ray beams at 25 and 35 keV. Ion chamber depth-dose measurements in a PMMA phantom were compared with the produc…
▽ More
Ion chamber dosimetry is being used to calibrate dose for cell irradiations designed to investigate photoactivated Auger electron therapy at the Louisiana State University CAMD synchrotron facility. This study performed a dosimetry intercomparison for synchrotron-produced monochromatic x-ray beams at 25 and 35 keV. Ion chamber depth-dose measurements in a PMMA phantom were compared with the product of MCNP5 Monte Carlo calculations of dose per fluence and measured incident fluence. Monochromatic beams of 25 and 35 keV were generated on the tomography beamline at CAMD. A cylindrical, air-equivalent ion chamber was used to measure the ionization created in a 10x10x10-cm3 PMMA phantom for depths from 0.6 to 7.7 cm. The American Association of Physicists in Medicine TG-61 protocol was applied to convert measured ionization into dose. Photon fluence was determined using a NaI detector to make scattering measurements of the beam from a thin polyethylene target at angles 30 degrees to 60 degrees. Differential Compton and Rayleigh scattering cross sections obtained from xraylib, an ANSI C library for x-ray-matter interactions, were applied to derive the incident fluence. MCNP5 simulations of the irradiation geometry provided the dose deposition per photon fluence as a function of depth in the phantom. At 25 keV the fluence-normalized MCNP5 dose overestimated the ion-chamber measured dose by an average of 7.2+/-3.0% to 2.1+/-3.0% for PMMA depths from 0.6 to 7.7 cm, respectively. At 35 keV the fluence-normalized MCNP5 dose underestimated the ion-chamber measured dose by an average of 1.0+/-3.4% to 2.5+/-3.4%, respectively. These results showed that TG-61 ion chamber dosimetry, used to calibrate dose output for cell irradiations, agreed with fluence-normalized MCNP5 calculations to within approximately 7% and 3% at 25 and 35 keV, respectively.
△ Less
Submitted 4 December, 2012;
originally announced December 2012.