First Linearity and Stability Characterization for CZT Detection System in a e$^+$e$^-$ Collider Environment
Authors:
Leonardo Abbene,
Francesco Artibani,
Manuele Bettelli,
Antonino Buttacavoli,
Fabio Principato,
Andrea Zappettini,
Massimiliano Bazzi,
Giacomo Borghi,
Mario Bragadireanu,
Michael Cargnelli,
Marco Carminati,
Alberto Clozza,
Francesco Clozza,
Luca De Paolis,
Raffaele Del Grande,
Kamil Dulski,
Laura Fabbietti,
Carlo Fiorini,
Carlo Guaraldo,
Mihail Iliescu,
Masahiko Iwasaki,
Aleksander Khreptak,
Simone Manti,
Johann Marton,
Pawel Moskal
, et al. (18 additional authors not shown)
Abstract:
The SIDDHARTA-2 collaboration built a new cadmium-zinc-telluride (CZT, CdZnTe)-based X-ray detection system, used for the first time in the DA$Φ$NE electron-positron collider at INFN-LNF. The aim of this work is to show that these detectors present optimal long- and short-term linearity and stability to perform precise spectroscopic measurements in a collider environment. The spectra used as refer…
▽ More
The SIDDHARTA-2 collaboration built a new cadmium-zinc-telluride (CZT, CdZnTe)-based X-ray detection system, used for the first time in the DA$Φ$NE electron-positron collider at INFN-LNF. The aim of this work is to show that these detectors present optimal long- and short-term linearity and stability to perform precise spectroscopic measurements in a collider environment. The spectra used as references for calibration are reported, and the results about the linearity and stability studies are presented. It is also discussed and showed what is the proper function to describe all the effects that alter the Gaussian shape in semiconductors, particularly evident in the CZT case. Good residuals and resolutions were obtained for all the calibrations. In a test run with the source and the collider beam on, it was demonstrated that the calibrations made with beam off are optimal also when the beam is on, and the actual systematics in a physics run were estimated. These promising results show the potentialities of this detector in the high rate environment of a particle collider, and pave the way for the use of CZT detectors in kaonic atoms researches and in accelerators, with applications for particle and nuclear physics.
△ Less
Submitted 16 October, 2024; v1 submitted 15 October, 2024;
originally announced October 2024.
-
Rigorous derivation of damped-driven wave turbulence theory
Authors:
Ricardo Grande,
Zaher Hani
Abstract:
We provide a rigorous justification of various kinetic regimes exhibited by the nonlinear Schrödinger equation with an additive stochastic forcing and a viscous dissipation. The importance of such damped-driven models stems from their wide empirical use in studying turbulence for nonlinear wave systems. The force injects energy into the system at large scales, which is then transferred across scal…
▽ More
We provide a rigorous justification of various kinetic regimes exhibited by the nonlinear Schrödinger equation with an additive stochastic forcing and a viscous dissipation. The importance of such damped-driven models stems from their wide empirical use in studying turbulence for nonlinear wave systems. The force injects energy into the system at large scales, which is then transferred across scales, thanks to the nonlinear wave interactions, until it is eventually dissipated at smaller scales. The presence of such scale-separated forcing and dissipation allows for the constant flux of energy in the intermediate scales, known as the inertial range, which is the focus of the vast amount of numerical and physical literature on wave turbulence.
Roughly speaking, our results provide a rigorous kinetic framework for this turbulent behavior by proving that the stochastic dynamics can be effectively described by a deterministic damped-driven kinetic equation, which carries the full picture of the turbulent energy dynamic across scales (like cascade spectra or other flux solutions). The analysis extends previous works in the unperturbed setting arXiv:1912.09518-arXiv:2301.07063 to the above empirically motivated damped driven setting. Here, in addition to the size $L$ of the system, and the strength $λ$ of the nonlinearity, an extra thermodynamic parameter has to be included in the kinetic limit ($L\to \infty, λ\to 0$), namely the strength $ν$ of the forcing and dissipation. Various regimes emerge depending on the relative sizes of $L$, $λ$ and $ν$, which give rise to different kinetic equations. Two major novelties of this work is the extension of the Feynman diagram analysis to additive stochastic objects, and the sharp asymptotic development of the leading terms in that expansion.
△ Less
Submitted 15 July, 2024;
originally announced July 2024.
-
Kaonic lead feasibility measurement at DAΦNE to solve the charged kaon mass discrepancy
Authors:
D. Bosnar,
L. Abbene,
C. Amsler,
F. Artibani,
M. Bazzi,
M. Bragadireanu,
A. Buttacavoli,
M. Cargnelli,
M. Carminati,
A. Clozza,
F. Clozza,
G. Deda,
L. De Paolis,
R. Del Grande,
K. Dulski,
L. Fabbietti,
C. Fiorini,
I. Friščić,
C. Guaraldo,
M. Iliescu,
M. Iwasaki,
A. Khreptak,
M. Makek,
S. Manti,
J. Marton
, et al. (17 additional authors not shown)
Abstract:
An HPGe detector equipped with a transistor reset preamplifier and readout with a CAEN DT5781 fast pulse digitizer was employed in the measurement of X-rays from kaonic lead at the DA$Φ$NE $e^+e^-$ collider at the Laboratori Nazionali di Frascati of INFN. A thin scintillator in front of a lead target was used to select kaons impinging on it and to form the trigger for the HPGe detector. We present…
▽ More
An HPGe detector equipped with a transistor reset preamplifier and readout with a CAEN DT5781 fast pulse digitizer was employed in the measurement of X-rays from kaonic lead at the DA$Φ$NE $e^+e^-$ collider at the Laboratori Nazionali di Frascati of INFN. A thin scintillator in front of a lead target was used to select kaons impinging on it and to form the trigger for the HPGe detector. We present the results of the kaonic lead feasibility measurement, where we show that the resolution of the HPGe detector in regular beam conditions remains the same as that without the beam and that a satisfactory background reduction can be achieved. This measurement serves as a test bed for future dedicated kaonic X-rays measurements for the more precise determination of the charged kaon mass.
△ Less
Submitted 21 May, 2024;
originally announced May 2024.
-
Numerical simulations of a stochastic dynamics leading to cascades and loss of regularity: applications to fluid turbulence and generation of fractional Gaussian fields
Authors:
Geoffrey Beck,
Charles-Edouard Bréhier,
Laurent Chevillard,
Ricardo Grande,
Wandrille Ruffenach
Abstract:
Motivated by the modeling of the spatial structure of the velocity field of three-dimensional turbulent flows, and the phenomenology of cascade phenomena, a linear dynamics has been recently proposed able to generate high velocity gradients from a smooth-in-space forcing term. It is based on a linear Partial Differential Equation (PDE) stirred by an additive random forcing term which is delta-corr…
▽ More
Motivated by the modeling of the spatial structure of the velocity field of three-dimensional turbulent flows, and the phenomenology of cascade phenomena, a linear dynamics has been recently proposed able to generate high velocity gradients from a smooth-in-space forcing term. It is based on a linear Partial Differential Equation (PDE) stirred by an additive random forcing term which is delta-correlated in time. The underlying proposed deterministic mechanism corresponds to a transport in Fourier space which aims at transferring energy injected at large scales towards small scales. The key role of the random forcing is to realize these transfers in a statistically homogeneous way. Whereas at finite times and positive viscosity the solutions are smooth, a loss of regularity is observed for the statistically stationary state in the inviscid limit. We here present novel simulations, based on finite volume methods in the Fourier domain and a splitting method in time, which are more accurate than the pseudo-spectral simulations. We show that the novel algorithm is able to reproduce accurately the expected local and statistical structure of the predicted solutions. We conduct numerical simulations in one, two and three spatial dimensions, and we display the solutions both in physical and Fourier spaces. We additionally display key statistical quantities such as second-order structure functions and power spectral densities at various viscosities.
△ Less
Submitted 8 March, 2024;
originally announced March 2024.
-
SIDDHARTA-2 apparatus for kaonic atoms research on the DA$Φ$NE collider
Authors:
F. Sirghi,
F. Sgaramella,
L. Abbene,
C. Amsler,
M. Bazzi,
G. Borghi,
D. Bosnar,
M. Bragadireanu,
A. Buttacavoli,
M. Carminati,
M. Cargnelli,
A. Clozza,
G. Deda,
L. De Paolis,
R. Del Grande,
K. Dulski,
L. Fabbietti,
C. Fiorini,
I. Friščić,
C. Guaraldo,
M. Iliescu,
M. Iwasaki,
A. Khreptak,
S. Manti,
J. Marton
, et al. (19 additional authors not shown)
Abstract:
SIDDHARTA-2 represents a state-of-the-art experiment designed to perform dedicated measurements of kaonic atoms, which are particular exotic atom configurations composed of a negatively charged kaon and a nucleus. Investigating these atoms provides an exceptional tool to comprehend the strong interactions in the non-perturbative regime involving strangeness. The experiment is installed at the DA…
▽ More
SIDDHARTA-2 represents a state-of-the-art experiment designed to perform dedicated measurements of kaonic atoms, which are particular exotic atom configurations composed of a negatively charged kaon and a nucleus. Investigating these atoms provides an exceptional tool to comprehend the strong interactions in the non-perturbative regime involving strangeness. The experiment is installed at the DA$Φ$NE electron-positron collider, of the INFN National Laboratory of Frascati (INFN-LNF) in Italy, aiming to perform the first-ever measurement of the 2p$\rightarrow$1s X-ray transitions in kaonic deuterium, a crucial step towards determining the isospin-dependent antikaon-nucleon scattering lengths. Based on the experience gained with the previous SIDDHARTA experiment, which performed the most precise measurement of the kaonic hydrogen 2p$\rightarrow$1s X-ray transitions, the present apparatus has been upgraded with innovative Silicon Drift Detectors (SDDs), distributed around a cryogenic gaseous target placed in a vacuum chamber at a short distance above the interaction region of the collider. We present a comprehensive description of the SIDDHARTA-2 setup including the optimization of its various components during the commissioning phase of the collider.
△ Less
Submitted 6 November, 2023;
originally announced November 2023.
-
CdZnTe detectors tested at the DAΦNE collider for future kaonic atoms measurements
Authors:
A. Scordo,
L. Abbene,
F. Artibani,
M. Bazzi,
M. Bettelli,
D. Bosnar,
G. Borghi,
M. Bragadireanu,
A. Buttacavoli,
M. Cargnelli,
M. Carminati,
A. Clozza,
F. Clozza,
L. De Paolis,
G. Deda,
R. Del Grande,
L. Fabbietti,
C. Fiorini,
I. Friščić,
C. Guaraldo,
M. Iliescu,
M. Iwasaki,
A. Khreptak,
S. Manti,
J. Marton
, et al. (20 additional authors not shown)
Abstract:
The SIDDHARTA-2 collaboration at the INFN Laboratories of Frascati (LNF) aims to perform groundbreaking measurements on kaonic atoms. In parallel and beyond the ongoing kaonic deuterium, presently running on the DA$Φ$NE collider at LNF, we plan to install additional detectors to perform further kaonic atoms' studies, taking advantage of the unique low energy and low momentum spread $K^-$ beam deli…
▽ More
The SIDDHARTA-2 collaboration at the INFN Laboratories of Frascati (LNF) aims to perform groundbreaking measurements on kaonic atoms. In parallel and beyond the ongoing kaonic deuterium, presently running on the DA$Φ$NE collider at LNF, we plan to install additional detectors to perform further kaonic atoms' studies, taking advantage of the unique low energy and low momentum spread $K^-$ beam delivered by the at-rest decay of the $φ$ meson. CdZnTe devices are ideal for detecting transitions toward both the upper and lower levels of intermediate-mass kaonic atoms, like kaonic carbon and aluminium, which have an important impact on the strangeness sector of nuclear physics. We present the results obtained in a set of preliminary tests conducted on DA$Φ$NE, in view of measurements foreseen in 2024, with the twofold aim to tune the timing window required to reject the extremely high electromagnetic background, and to quantify the readout electronics saturation effect due to the high rate, when placed close to the Interaction Region (IR). In the first test we used commercial devices and electronics, while for the second one both were customized at the IMEM-CNR of Parma and the University of Palermo. The results confirmed the possibility of finding and matching a proper timing window where to identify the signal events and proved better performances, in terms of energy resolution, of the custom system. In both cases, strong saturation effects were confirmed, accounting for a loss of almost 90\% of the events, which will be overcome by a dedicated shielding structure foreseen for the final experimental setup.
△ Less
Submitted 23 October, 2023;
originally announced October 2023.
-
Novel Machine Learning and Differentiable Programming Techniques applied to the VIP-2 Underground Experiment
Authors:
F Napolitano,
M Bazzi,
M Bragadireanu,
M Cargnelli,
A Clozza,
L De Paolis,
R Del Grande,
C Fiorini,
C Guaraldo,
M Iliescu,
M Laubenstein,
S Manti,
J Marton,
M Miliucci,
K Piscicchia,
A Porcelli,
A Scordo,
F Sgaramella,
D Sirghi,
F Sirghi,
O Doce,
J Zmeskal,
C Curceanu
Abstract:
In this work, we present novel Machine Learning and Differentiable Programming enhanced calibration techniques used to improve the energy resolution of the Silicon Drift Detectors (SDDs) of the VIP-2 underground experiment at the Gran Sasso National Laboratory (LNGS). We achieve for the first time a Full Width at Half Maximum (FWHM) in VIP-2 below 180 eV at 8 keV, improving around 10 eV on the pre…
▽ More
In this work, we present novel Machine Learning and Differentiable Programming enhanced calibration techniques used to improve the energy resolution of the Silicon Drift Detectors (SDDs) of the VIP-2 underground experiment at the Gran Sasso National Laboratory (LNGS). We achieve for the first time a Full Width at Half Maximum (FWHM) in VIP-2 below 180 eV at 8 keV, improving around 10 eV on the previous state-of-the-art. SDDs energy resolution is a key parameter in the VIP-2 experiment, which is dedicated to searches for physics beyond the standard quantum theory, targeting Pauli Exclusion Principle (PEP) violating atomic transitions. Additionally, we show that this method can correct for potential miscalibrations, requiring less fine-tuning with respect to standard methods.
△ Less
Submitted 26 May, 2023;
originally announced May 2023.
-
Measurements of high-n transitions in intermediate mass kaonic atoms by SIDDHARTA-2 at DA$\mathrmΦ$NE
Authors:
F. Sgaramella,
M. Tüchler,
C. Amsler,
M. Bazzi,
D. Bosnar,
M. Bragadireanu,
M. Cargnelli,
M. Carminati,
A. Clozza,
G. Deda,
R. Del Grande,
L. De Paolis,
L. Fabbietti,
C. Fiorini,
I. Friščić,
C. Guaraldo,
M. Iliescu,
M. Iwasaki,
A. Khreptak,
S. Manti,
J. Marton,
M. Miliucci,
P. Moskal,
F. Napolitano,
S. Niedźwiecki
, et al. (16 additional authors not shown)
Abstract:
The SIDDHARTA-2 experiment installed at the DA$\mathrmΦ$NE collider of INFN-LNF performed, for the first time, measurements of high-n transitions in intermediate mass kaonic atoms during the data taking campaigns of 2021 and 2022. Kaonic carbon, oxygen, nitrogen and aluminium transitions, which occur in the setup materials, were measured by using the kaons stopped in the gaseous helium target cell…
▽ More
The SIDDHARTA-2 experiment installed at the DA$\mathrmΦ$NE collider of INFN-LNF performed, for the first time, measurements of high-n transitions in intermediate mass kaonic atoms during the data taking campaigns of 2021 and 2022. Kaonic carbon, oxygen, nitrogen and aluminium transitions, which occur in the setup materials, were measured by using the kaons stopped in the gaseous helium target cell with aluminium frames and Kapton walls, and are reported in this paper. These new kaonic atoms measurements add valuable input to the kaonic atoms transitions data base, which is used as a reference for theories and models of the low-energy strong interaction between antikaon and nuclei. Moreover, these results pave the way for future dedicated kaonic atoms measurements through the whole periodic table and to a new era for the antikaon-nuclei studies at low energy.
△ Less
Submitted 22 April, 2023;
originally announced April 2023.
-
New opportunities for kaonic atoms measurements from CdZnTe detectors
Authors:
L. Abbene,
M. Bettelli,
A. Buttacavoli,
F. Principato,
A. Zappettini,
C. Amsler,
M. Bazzi,
D. Bosnar,
M. Bragadireanu,
M. Cargnelli,
M. Carminati,
A. Clozza,
G. Deda,
L. De Paolis,
R. Del Grande,
L. Fabbietti,
C. Fiorini,
I. Friščić,
C. Guaraldo,
M. Iliescu,
M. Iwasaki,
A. Khreptak,
S. Manti,
J. Marton,
M. Miliucci
, et al. (20 additional authors not shown)
Abstract:
We present the tests performed by the SIDDHARTA-2 collaboration at the DAΦNE collider with a quasi-hemispherical CdZnTe detector. The very good room-temperature energy resolution and efficiency in a wide energy range show that this detector technology is ideal for studying radiative transitions in intermediate and heavy-mass kaonic atoms. The CdZnTe detector was installed for the first time in an…
▽ More
We present the tests performed by the SIDDHARTA-2 collaboration at the DAΦNE collider with a quasi-hemispherical CdZnTe detector. The very good room-temperature energy resolution and efficiency in a wide energy range show that this detector technology is ideal for studying radiative transitions in intermediate and heavy-mass kaonic atoms. The CdZnTe detector was installed for the first time in an accelerator environment to perform tests on the background rejection capabilities, which were achieved by exploiting the SIDDHARTA-2 Luminosity Monitor. A spectrum with an $^{241}Am$ source has been acquired, with beams circulating in the main rings, and peak resolutions of 6% at 60 keV and of 2.2% at 511 keV have been achieved. The background suppression factor, which turned out to be of the order of $\simeq10^{5-6}$, opens the possibility to plan for future kaonic atom measurements with CdZnTe detectors.
△ Less
Submitted 28 January, 2023;
originally announced January 2023.
-
Studies of the linearity and stability of Silicon Drift Detectors for kaonic atoms X-ray spectroscopy
Authors:
A. Khreptak,
C. Amsler,
M. Bazzi,
D. Bosnar,
M. Bragadireanu,
M. Carminati,
M. Cargnelli,
A. Clozza,
G. Deda,
L. De Paolis,
R. Del Grande,
L. Fabbietti,
C. Fiorini,
C. Guaraldo,
M. Iliescu,
M. Iwasaki,
S. Manti,
J. Marton,
M. Miliucci,
P. Moskal,
F. Napolitano,
S. Niedźwiecki,
H. Ohnishi,
K. Piscicchia,
Y. Sada
, et al. (14 additional authors not shown)
Abstract:
The SIDDHARTA-2 experiment at the DA$Φ$NE collider aims to perform precision measurements of kaonic atoms X-ray spectroscopy for the investigation of the antikaon-nucleon strong interaction. To achieve this goal, novel large-area Silicon Drift Detectors (SDDs) have been developed. These devices have special geometry, field configuration and readout electronics that ensure excellent performance in…
▽ More
The SIDDHARTA-2 experiment at the DA$Φ$NE collider aims to perform precision measurements of kaonic atoms X-ray spectroscopy for the investigation of the antikaon-nucleon strong interaction. To achieve this goal, novel large-area Silicon Drift Detectors (SDDs) have been developed. These devices have special geometry, field configuration and readout electronics that ensure excellent performance in terms of linearity and stability. The paper presents preliminary results for the linearity determination and stability monitoring of the SDDs system during the measurement of kaonic deuterium carried out in the summer of 2022.
△ Less
Submitted 31 August, 2022;
originally announced August 2022.
-
First tests of the full SIDDHARTA-2 experimental apparatus with a 4 He gaseous target
Authors:
A. Scordo,
C. Amsler,
M. Bazzi,
D. Bosnar,
M. Bragadireanu,
M. Cargnelli,
M. Carminati,
A. Clozza,
G. Deda,
L. De Paolis,
R. Del Grande,
L. Fabbietti,
C. Fiorini,
C. Guaraldo,
M. Iliescu,
M. Iwasaki,
A. Khreptak,
P. King,
P. Levi Sandri,
S. Manti,
J. Marton,
M. Miliucci,
P. Moskal,
F. Napolitano,
S. Niedźwiecki
, et al. (16 additional authors not shown)
Abstract:
In this paper, we present the first tests performed after the full installation of the SIDDHARTA-2 experimental apparatus on the Interaction Region of the DAΦNE collider at the INFN National Laboratories of Frascati. Before starting the first measurement of the kaonic deuterium 2p{\rightarrow}1s transition, accurate evaluation of the background rejection. mainly achieved with the Kaon Trigger syst…
▽ More
In this paper, we present the first tests performed after the full installation of the SIDDHARTA-2 experimental apparatus on the Interaction Region of the DAΦNE collider at the INFN National Laboratories of Frascati. Before starting the first measurement of the kaonic deuterium 2p{\rightarrow}1s transition, accurate evaluation of the background rejection. mainly achieved with the Kaon Trigger system, was required. This run, performed in the period 04-26/05/2022 with a 4 He gaseous target, confirmed the 10^5 rejection factor obtained with a reduced version of the setup and different machine conditions in 2021. This important outcome motivated the filling of the target cell with deuterium and the starting of the measurement campaign of the kaonic deuterium 2p{\rightarrow}1s transition.
△ Less
Submitted 4 August, 2022;
originally announced August 2022.
-
The SIDDHARTA-2 calibration method for high precision kaonic atoms X-ray spectroscopy measurements
Authors:
F Sgaramella,
M Miliucci,
M Bazzi,
D Bosnar,
M Bragadireanu,
M Carminati,
M Cargnelli,
A Clozza,
G Deda,
L De Paolis,
R Del Grande,
C Fiorini,
C Guaraldo,
M Iliescu,
M Iwasaki,
P King,
P Levi Sandri,
J Marton,
P Moskal,
F Napolitano,
S Niedźwiecki,
K Piscicchia,
A Scordo,
H Shi,
M Silarski
, et al. (7 additional authors not shown)
Abstract:
The SIDDHARTA-2 experiment at the DA$Φ$NE collider aims to perform the first kaonic deuterium X-ray transitions to the fundamental level measurement, with a systematic error at the level of a few eV. To achieve this challenging goal the experimental apparatus is equipped with 384 Silicon Drift Detectors (SDDs) distributed around its cryogenic gaseous target. The SDDs developed by the SIDDHARTA-2 c…
▽ More
The SIDDHARTA-2 experiment at the DA$Φ$NE collider aims to perform the first kaonic deuterium X-ray transitions to the fundamental level measurement, with a systematic error at the level of a few eV. To achieve this challenging goal the experimental apparatus is equipped with 384 Silicon Drift Detectors (SDDs) distributed around its cryogenic gaseous target. The SDDs developed by the SIDDHARTA-2 collaboration are suitable for high precision kaonic atoms spectroscopy, thanks to their high energy and time resolutions combined with their radiation hardness. The energy response of each detector must be calibrated and monitored to keep the systematic error, due to processes such as gain fluctuations, at the level of 2-3 eV. This paper presents the SIDDHARTA-2 calibration method which was optimized during the preliminary phase of the experiment in the real background conditions of the DA$Φ$NE collider, which is a fundamental tool to guarantee the high precision spectroscopic performances of the system over long periods of data taking, as that required for the kaonic deuterium measurement.
△ Less
Submitted 28 January, 2022;
originally announced January 2022.
-
Kaonic Atoms at the DA$Φ$NE Collider with the SIDDHARTA-2 Experiment
Authors:
F Napolitano,
F Sgaramella,
M Bazzi,
D Bosnar,
M Bragadireanu,
M Carminati,
M Cargnelli,
A Clozza,
G Deda,
L De Paolis,
R Del Grande,
L Fabbietti,
C Fiorini,
C Guaraldo,
M Iliescu,
M Iwasaki,
P Levi Sandri,
J Marton,
M Miliucci,
P Moskal,
S Niedźwiecki,
K Piscicchia,
A Scordo,
H Shi,
D Sirghi
, et al. (7 additional authors not shown)
Abstract:
Kaonic atoms are a unique tool to explore quantum chromodynamics in the strangeness sector at low energy, with implications reaching neutron stars and dark matter. Precision X-ray spectroscopy can fully unlock the at-threshold isospin dependent antikaon-nucleon scattering lengths, via the atomic transitions to the fundamental level. While the SIDDHARTA experiment at the INFN-LNF DA$Φ$NE collider s…
▽ More
Kaonic atoms are a unique tool to explore quantum chromodynamics in the strangeness sector at low energy, with implications reaching neutron stars and dark matter. Precision X-ray spectroscopy can fully unlock the at-threshold isospin dependent antikaon-nucleon scattering lengths, via the atomic transitions to the fundamental level. While the SIDDHARTA experiment at the INFN-LNF DA$Φ$NE collider successfully measured kaonic hydrogen, its successor SIDDHARTA-2 is starting now its data taking campaign aiming to finally fully disentangle the isoscalar and isovector scattering lengths via the measurement of kaonic deuterium. An overview of the first experimental results from a preparatory run for the SIDDAHARTA-2 experiment is presented.
△ Less
Submitted 27 January, 2022;
originally announced January 2022.
-
A new kaonic helium measurement in gas by SIDDHARTINO at the DAΦNE collider
Authors:
D Sirghi,
F Sirghi,
F Sgaramella,
M Bazzi,
D Bosnar,
M Bragadireanu,
M Carminati,
M Cargnelli,
A Clozza,
G Deda,
L De Paolis,
R Del Grande,
L Fabbietti,
C Fiorini,
C Guaraldo,
M Iliescu,
M Iwasaki,
P Levi Sandri,
J Marton,
M Miliucci,
P Moskal,
F Napolitano,
S Niedźwiecki,
K Piscicchia,
A Scordo
, et al. (8 additional authors not shown)
Abstract:
The SIDDHARTINO experiment at the DAΦNE Collider of INFN-LNF, the pilot run for the SIDDHARTA-2 experiment which aims to perform the measurement of kaonic deuterium transitions to the fundamental level, has successfully been concluded. The paper reports the main results of this run, including the optimization of various components of the apparatus, among which the degrader needed to maximize the f…
▽ More
The SIDDHARTINO experiment at the DAΦNE Collider of INFN-LNF, the pilot run for the SIDDHARTA-2 experiment which aims to perform the measurement of kaonic deuterium transitions to the fundamental level, has successfully been concluded. The paper reports the main results of this run, including the optimization of various components of the apparatus, among which the degrader needed to maximize the fraction of kaons stopped inside the target, through measurements of kaonic helium transitions to the 2p level. The obtained shift and width values are ε_2p = E_exp-E_e.m = 0.2 {\pm} 2.5(stat) {\pm} 2(syst) eV and Γ_2p = 8 {\pm} 10 eV (stat), respectively. This new measurement of the shift, in particular, represents the most precise one for a gaseous target and is expected to contribute to a better understanding of the kaon-nuclei interaction at low energy.
△ Less
Submitted 24 January, 2022;
originally announced January 2022.
-
Silicon Drift Detectors spectroscopic response during the SIDDHARTA-2 Kaonic Helium run at the DAΦNE collider
Authors:
Marco Miliucci,
Massimiliano Bazzi,
Damir Bosnar,
Mario Bragadireanu,
Marco Carminati,
Michael Cargnelli,
Alberto Clozza,
Catalina Curceanu,
Griseld Deda,
Luca De Paolis,
Raffaele Del Grande,
Carlo Fiorini,
Carlo Guaraldo,
Mihail Iliescu,
Masahiko Iwasaki,
Pietro King,
Paolo Levi Sandri,
Johann Marton,
Paweł Moskal,
Fabrizio Napolitano,
Szymon Niedźwiecki,
Kristian Piscicchia,
Alessandro Scordo,
Francesco Sgaramella,
Hexi Shi
, et al. (8 additional authors not shown)
Abstract:
A large-area Silicon Drift Detectors (SDDs) system has been developed by the SIDDHARTA2 collaboration for high precision light kaonic atoms X-ray spectroscopy at the DAΦNE collider of Istituto Nazionale di Fisica Nucleare - Laboratori Nazionali di Frascati. The SDDs geometry and electric field configuration, combined with their read-out electronics, make these devices suitable to perform high prec…
▽ More
A large-area Silicon Drift Detectors (SDDs) system has been developed by the SIDDHARTA2 collaboration for high precision light kaonic atoms X-ray spectroscopy at the DAΦNE collider of Istituto Nazionale di Fisica Nucleare - Laboratori Nazionali di Frascati. The SDDs geometry and electric field configuration, combined with their read-out electronics, make these devices suitable to perform high precision light kaonic atoms spectroscopy measurements in the high background of the DAΦNE collider. This work presents the spectroscopic response of the SDDs system during the kaonic helium first exotic atoms run of SIDDHARTA-2, preliminary to the kaonic deuterium data taking campaign
△ Less
Submitted 2 November, 2021;
originally announced November 2021.
-
Synchronisation and calibration of the 24-modules J-PET prototype with 300~mm axial field of view
Authors:
P. Moskal,
T. Bednarski,
Sz. Niedzwiecki,
M. Silarski,
E. Czerwinski,
T. Kozik,
J. Chhokar,
M. Bała,
C. Curceanu,
R. Del Grande,
M. Dadgar,
K. Dulski,
A. Gajos,
M. Gorgol,
N. Gupta-Sharma,
B. C. Hiesmayr,
B. Jasinska,
K. Kacprzak,
L. Kaplon,
H. Karimi,
D. Kisielewska,
K. Klimaszewski,
G. Korcyl,
P. Kowalski,
N. Krawczyk
, et al. (14 additional authors not shown)
Abstract:
Research conducted in the framework of the J-PET project aims to develop a cost-effective total-body positron emission tomography scanner. As a first step on the way to construct a full-scale J-PET tomograph from long strips of plastic scintillators, a 24-strip prototype was built and tested. The prototype consists of detection modules arranged axially forming a cylindrical diagnostic chamber with…
▽ More
Research conducted in the framework of the J-PET project aims to develop a cost-effective total-body positron emission tomography scanner. As a first step on the way to construct a full-scale J-PET tomograph from long strips of plastic scintillators, a 24-strip prototype was built and tested. The prototype consists of detection modules arranged axially forming a cylindrical diagnostic chamber with the inner diameter of 360 mm and the axial field-of-view of 300 mm. Promising perspectives for a low-cost construction of a total-body PET scanner are opened due to an axial arrangement of strips of plastic scintillators, wchich have a small light attenuation, superior timing properties, and the possibility of cost-effective increase of the axial field-of-view. The presented prototype comprises dedicated solely digital front-end electronic circuits and a triggerless data acquisition system which required development of new calibration methods including time, thresholds and gain synchronization. The system and elaborated calibration methods including first results of the 24-module J-PET prototype are presented and discussed. The achieved coincidence resolving time equals to CRT = 490 $\pm$ 9 ps. This value can be translated to the position reconstruction accuracy $σ(Δl) =$ 18 mm which is fairly position-independent.
△ Less
Submitted 25 August, 2020;
originally announced August 2020.
-
Characterization of the SIDDHARTA-2 luminosity monitor
Authors:
M. Skurzok,
A. Scordo,
S. Niedzwiecki,
A. Baniahmad,
M. Bazzi,
D. Bosnar,
M. Bragadireanu,
M. Carminati,
M. Cargnelli,
A. Clozza,
C. Curceanu,
L. De Paolis,
R. Del Grande,
L. Fabbietti,
C. Fiorini,
C. Guaraldo,
M. Iliescu,
M. Iwasaki,
P. Levi Sandri,
J. Marton,
M. Miliucci,
P. Moskal,
K. Piscicchia,
F. Sgaramella,
H. Shi
, et al. (7 additional authors not shown)
Abstract:
A luminosity monitor, based on plastic scintillator detectors, has been developed for the SIDDHARTA-2 experiment aiming to perform high precision measurements of kaonic atoms and was installed in 2020 on the DAFNE $e^+e^-$ collider at LNF (Laboratori Nazionali di Frascati, INFN). The main goal of this system is to provide the~instantaneous and integrated luminosity of the DAFNE facility by measuri…
▽ More
A luminosity monitor, based on plastic scintillator detectors, has been developed for the SIDDHARTA-2 experiment aiming to perform high precision measurements of kaonic atoms and was installed in 2020 on the DAFNE $e^+e^-$ collider at LNF (Laboratori Nazionali di Frascati, INFN). The main goal of this system is to provide the~instantaneous and integrated luminosity of the DAFNE facility by measuring the rate of $K^+K^-$ correlated pairs emitted by the phi meson decay. This task requires an accurate timing of the DAQ signals, as well as timing resolution below 1ns, in order to disentangle the $K^\pm$ signals from the background minimum ionizing particles (MIPs) produced during the $e^+e^-$ collisions at DAFNE. In this paper the luminosity monitor concept as well as its laboratory characterization and the first results inside DAFNE are presented.
△ Less
Submitted 15 October, 2020; v1 submitted 12 August, 2020;
originally announced August 2020.
-
The J-PET detector -- a tool for precision studies of ortho-positronium decays
Authors:
K. Dulski,
S. D. Bass,
J. Chhokar,
N. Chug,
C. Curceanu,
E. Czerwiński,
M. Dadgar,
J. Gajewski,
A. Gajos,
M. Gorgol,
R. Del Grande,
B. C. Hiesmayr,
B. Jasińska,
K. Kacprzak,
Ł. Kapłon,
H. Karimi,
D. Kisielewska,
K. Klimaszewski,
P. Kopka,
G. Korcyl,
P. Kowalski,
T. Kozik,
N. Krawczyk,
W. Krzemień,
E. Kubicz
, et al. (19 additional authors not shown)
Abstract:
The J-PET tomograph is constructed from plastic scintillator strips arranged axially in concentric cylindrical layers. It enables investigations of positronium decays by measurement of the time, position, polarization and energy deposited by photons in the scintillators, in contrast to studies conducted so far with crystal and semiconductor based detection systems where the key selection of events…
▽ More
The J-PET tomograph is constructed from plastic scintillator strips arranged axially in concentric cylindrical layers. It enables investigations of positronium decays by measurement of the time, position, polarization and energy deposited by photons in the scintillators, in contrast to studies conducted so far with crystal and semiconductor based detection systems where the key selection of events is based on the measurement of the photons energies. In this article we show that the J-PET tomograph system is capable of exclusive measurements of the decays of ortho-positronium atoms. We present the first positronium production results, its lifetime distribution measurements and discuss estimation of the influence of various background sources. The tomograph s performance demonstrated here makes it suitable for precision studies of positronium decays including entanglement of the final state photons, positron annihilation lifetime spectroscopy plus molecular imaging diagnostics.
△ Less
Submitted 25 May, 2021; v1 submitted 9 June, 2020;
originally announced June 2020.
-
Modelling aerosol transport and virus exposure with numerical simulations in relation to SARS-CoV-2 transmission by inhalation indoors
Authors:
Ville Vuorinen,
Mia Aarnio,
Mikko Alava,
Ville Alopaeus,
Nina Atanasova,
Mikko Auvinen,
Nallannan Balasubramanian,
Hadi Bordbar,
Panu Erästö,
Rafael Grande,
Nick Hayward,
Antti Hellsten,
Simo Hostikka,
Jyrki Hokkanen,
Ossi Kaario,
Aku Karvinen,
Ilkka Kivistö,
Marko Korhonen,
Risto Kosonen,
Janne Kuusela,
Sami Lestinen,
Erkki Laurila,
Heikki Nieminen,
Petteri Peltonen,
Juho Pokki
, et al. (5 additional authors not shown)
Abstract:
We provide research findings on the physics of aerosol dispersion relevant to the hypothesized aerosol transmission of SARS-CoV-2. We utilize physics-based modeling at different levels of complexity, and literature on coronaviruses, to investigate the possibility of airborne transmission. The previous literature, our 0D-3D simulations by various physics-based models, and theoretical calculations,…
▽ More
We provide research findings on the physics of aerosol dispersion relevant to the hypothesized aerosol transmission of SARS-CoV-2. We utilize physics-based modeling at different levels of complexity, and literature on coronaviruses, to investigate the possibility of airborne transmission. The previous literature, our 0D-3D simulations by various physics-based models, and theoretical calculations, indicate that the typical size range of speech and cough originated droplets (d < 20microns) allows lingering in the air for O(1h) so that they could be inhaled. Consistent with the previous literature, numerical evidence on the rapid drying process of even large droplets, up to sizes O(100microns), into droplet nuclei/aerosols is provided. Based on the literature and the public media sources, we provide evidence that the infected individuals could have been exposed to aerosols/droplet nuclei by inhaling them in significant numbers e.g. O(100). By 3D computational fluid dynamics (CFD) simulations, we give examples on the transport and dilution of aerosols (d<20microns) over distances O(10m) in generic environments. We study susceptible and infected individuals in generic public places by Monte-Carlo modeling. The model accounts for the locally varying aerosol concentration levels which the susceptible accumulate via inhalation. The introduced concept, 'exposure time' to virus containing aerosols is proposed to complement the traditional 'safety distance' thinking. We show that the exposure time to inhale O(100) aerosols could range from O(1s) to O(1min) or even to O(1h) depending on the situation. The Monte Carlo analysis provides clear quantitative insight to the exposure time in different public indoor environments.
△ Less
Submitted 26 May, 2020;
originally announced May 2020.
-
Hit-time and hit-position reconstruction in strips of plastic scintillators using multi-threshold readouts
Authors:
N. G. Sharma,
M. Silarski,
J. Chhokar,
E. Czerwinski,
C. Curceanu,
K. Dulski,
K. Farbaniec,
A. Gajos,
R. Del Grande,
M. Gorgol,
B. C. Hiesmayr,
B. Jasinska,
K. Kacprzak,
L. Kaplon,
D. Kisielewska,
K. Klimaszewski,
G. Korcyl,
P. Kowalski,
N. Krawczyk,
W. Krzemien,
T. Kozik,
E. Kubicz,
M. Mohammed,
Sz. Niedzwiecki,
M. Palka
, et al. (10 additional authors not shown)
Abstract:
In this article, a new method for the reconstruction of hit-position and hit-time of photons in long scintillator detectors is investigated. This research is motivated by the recent development of the positron emission tomography scanners based on plastic scintillators. The proposed method constitutes a new way of signal processing in Multi-Voltage-Technique. It is based on the determination of th…
▽ More
In this article, a new method for the reconstruction of hit-position and hit-time of photons in long scintillator detectors is investigated. This research is motivated by the recent development of the positron emission tomography scanners based on plastic scintillators. The proposed method constitutes a new way of signal processing in Multi-Voltage-Technique. It is based on the determination of the degree of similarity between the registered signals and the synchronized model signals stored in a library. The library was established for a set of well defined hit-positions along the length of the scintillator. The Mahalanobis distance was used as a measure of similarity between the two compared signals. The method was validated on the experimental data measured using two-strips J-PET prototype with dimensions of 5x9x300 mm$^3$. The obtained Time-of-Flight (TOF) and spatial resolutions amount to 325~ps (FWHM) and 25~mm (FWHM), respectively. The TOF resolution was also compared to the results of an analogous study done using Linear Fitting method. The best TOF resolution was obtained with this method at four pre-defined threshold levels which was comparable to the resolution achieved from the Mahalanobis distance at two pre-defined threshold levels. Although the algorithm of Linear Fitting method is much simpler to apply than the Mahalanobis method, the application of the Mahalanobis distance requires a lower number of applied threshold levels and, hence, decreases the costs of electronics used in PET scanner.
△ Less
Submitted 27 April, 2020;
originally announced April 2020.
-
Estimating relationship between the Time Over Threshold and energy loss by photons in plastic scintillators used in the J-PET scanner
Authors:
S. Sharma,
J. Chhokar,
C. Curceanu,
E. Czerwinski,
M. Dadgar,
K. Dulski,
J. Gajewski,
A. Gajos,
M. Gorgol,
N. Gupta-Sharma,
R. Del Grande,
B. C. Hiesmayr,
B. Jasinska,
K. Kacprzak,
L. Kaplon,
H. Karimi,
D. Kisielewska,
K. Klimaszewski,
G. Korcyl,
P. Kowalski,
T. Kozik,
N. Krawczyk,
W. Krzemien,
E. Kubicz,
M. Mohammed
, et al. (14 additional authors not shown)
Abstract:
Time-Over-Threshold (TOT) technique is being used widely due to its implications in developing the multi channel readouts mainly when fast signal processing is required. Using TOT technique as a measure of energy loss instead of charge integration methods significantly reduces the signals readout cost by combining the time and energy information. Therefore, this approach can potentially be used in…
▽ More
Time-Over-Threshold (TOT) technique is being used widely due to its implications in developing the multi channel readouts mainly when fast signal processing is required. Using TOT technique as a measure of energy loss instead of charge integration methods significantly reduces the signals readout cost by combining the time and energy information. Therefore, this approach can potentially be used in J-PET tomograph which is build from plastic scintillators characterized by fast light signals. The drawback in adopting this technique is lying in the non-linear correlation between input energy loss and TOT of the signal. The main motivation behind this work is to develop the relationship between TOT and energy loss and validate it with the J-PET tomograph.
The experiment was performed using the $^{22}$Na beta emitter source placed in the center of the J-PET tomograph. One can obtain primary photons of two different energies: 511 keV photon from the annihilation of positron (direct annihilation or through the formation of para-Positronim atom or pick-off process of ortho-Positronium atoms), and 1275 keV prompt photon. This allows to study the correlation between TOT values and energy loss for energy range up to 1000 keV. As the photon interacts dominantly via Compton scattering inside the plastic scintillator, there is no direct information of primary photon energy. However, using the J-PET geometry one can measure the scattering angle of the interacting photon. Since, $^{22}$Na source emits photons of two different energies, it is required to know unambiguously the energy of incident photons and its corresponding scattering angle for the estimation of energy deposition. In this work, the relationship between Time Over Threshold and energy loss by interacting photons inside the plastic scintillators used in J-PET scanner is established for a energy deposited range 100-1000 keV
△ Less
Submitted 27 November, 2019;
originally announced November 2019.
-
Performance assessment of the 2$γ$ positronium imaging with the total-body PET scanners
Authors:
P. Moskal,
D. Kisielewska,
Z. Bura,
C. Chhokar,
C. Curceanu,
E. Czerwiński,
M. Dadgar 1,
K. Dulski,
J. Gajewski,
A. Gajos,
M. Gorgol,
R. Del Grande,
B. C. Hiesmayr,
B. Jasińska,
K. Kacprzak,
A. Kamińska,
Ł. Kapłon,
H. Karimi,
G. Korcyl,
P. Kowalski,
N. Krawczyk,
W. Krzemień,
T. Kozik,
E. Kubicz,
P. Małczak
, et al. (17 additional authors not shown)
Abstract:
In living organisms the positron-electron annihilation (occurring during the PET imaging) proceeds in about 30% via creation of a metastable ortho-positronium atom. In the tissue, due to the pick-off and conversion processes, over 98% of ortho-positronia annihilate into two 511~keV photons. In this article we assess the feasibility for reconstruction of the mean ortho-positronium lifetime image ba…
▽ More
In living organisms the positron-electron annihilation (occurring during the PET imaging) proceeds in about 30% via creation of a metastable ortho-positronium atom. In the tissue, due to the pick-off and conversion processes, over 98% of ortho-positronia annihilate into two 511~keV photons. In this article we assess the feasibility for reconstruction of the mean ortho-positronium lifetime image based on annihilations into two photons. The main objectives of this work include: (i) estimation of the sensitivity of the total-body PET scanners for the ortho-positronium mean lifetime imaging using $2γ$ annihilations, and (ii) estimation of the spatial and time resolution of the ortho-positronium image as a function of the coincidence resolving time (CRT) of the scanner. Simulations are conducted assuming that radiopharmaceutical is labelled with $^{44}Sc$ isotope emitting one positron and one prompt gamma. The image is reconstructed on the basis of triple coincidence events. The ortho-positronium lifetime spectrum is determined for each voxel of the image. Calculations were performed for cases of total-body detectors build of (i) LYSO scintillators as used in the EXPLORER PET, and (ii) plastic scintillators as anticipated for the cost-effective total-body J-PET scanner. To assess the spatial and time resolution the three cases were considered assuming that CRT is equal to 140ps, 50ps and 10ps. The estimated total-body PET sensitivity for the registration and selection of image forming triple coincidences is larger by a factor of 12.2 (for LYSO PET) and by factor of 4.7 (for plastic PET) with respect to the sensitivity for the standard $2γ$ imaging by LYSO PET scanners with AFOV=20cm.
△ Less
Submitted 15 November, 2019;
originally announced November 2019.
-
VIP2 in LNGS - Testing the Pauli Exclusion Principle for electrons with high sensitivity
Authors:
J. Marton,
A. Pichler,
A. Amirkhani,
S. Bartalucci,
M. Bazzi,
S. Bertolucci,
M. Bragadireanu,
M. Cargnelli,
A. Clozza,
C. Curceanu,
R. Del Grande,
L. De Paolis,
J. -P. Egger,
C. Fiorini,
C. Guaraldo,
M. Iliescu,
M. Laubenstein,
E. Milotti,
M. Milucci,
D. Pietreanu,
K. Piscicchia,
A. Scordo,
H. Shi,
D. Sirghi,
F. Sirghi
, et al. (3 additional authors not shown)
Abstract:
The VIP2 (VIolation of the Pauli Exclusion Principle) experiment at the Gran Sasso underground laboratory (LNGS) is searching for possible violations of standard quantum mechanics predictions in atoms at very high sensitivity. We investigate atomic transitions with precision X-ray spectroscopy in order to test the Pauli Exclusion Principle (PEP) and therefore the related spin-statistics theorem. W…
▽ More
The VIP2 (VIolation of the Pauli Exclusion Principle) experiment at the Gran Sasso underground laboratory (LNGS) is searching for possible violations of standard quantum mechanics predictions in atoms at very high sensitivity. We investigate atomic transitions with precision X-ray spectroscopy in order to test the Pauli Exclusion Principle (PEP) and therefore the related spin-statistics theorem. We will present our experimental method for the search for "anomalous" (i.e. Pauli-forbidden) X-ray transitions in copper atoms, produced by "new" electrons, which could have tiny probability to undergo Pauli-forbidden transition to the ground state already occupied by two electrons. We will describe the VIP2 experimental setup, which is taking data at LNGS presently. The goal of VIP2 is to test the PEP for electrons with unprecedented accuracy, down to a limit in the probability that PEP is violated at the level of 10$^{-31}$. We will present current experimental results and discuss implications of a possible violation.
△ Less
Submitted 21 March, 2019;
originally announced March 2019.
-
Feasibility studies of the polarization of photons beyond the optical wavelength regime with the J-PET detector
Authors:
P. Moskal,
N. Krawczyk,
B. C. Hiesmayr,
M. Bała,
C. Curceanu,
E. Czerwiński,
K. Dulski,
A. Gajos,
M. Gorgol,
R. Del Grande,
B. Jasińska,
K. Kacprzak,
L. Kapłon,
D. Kisielewska,
K. Klimaszewski,
G. Korcyl,
P. Kowalski,
T. Kozik,
W. Krzemień,
E. Kubicz,
M. Mohammed,
Sz. Niedźwiecki,
M. Pałka,
J. Raj,
Z. Rudy
, et al. (9 additional authors not shown)
Abstract:
J-PET is a detector optimized for registration of photons from the electron-positron annihilation via plastic scintillators where photons interact predominantly via Compton scattering. Registration of both primary and scattered photons enables to determinate the linear polarization of the primary photon on the event by event basis with a certain probability. Here we present quantitative results on…
▽ More
J-PET is a detector optimized for registration of photons from the electron-positron annihilation via plastic scintillators where photons interact predominantly via Compton scattering. Registration of both primary and scattered photons enables to determinate the linear polarization of the primary photon on the event by event basis with a certain probability. Here we present quantitative results on the feasibility of such polarization measurements of photons from the decay of positronium with the J-PET and explore the physical limitations for the resolution of the polarization determination of 511 keV photons via Compton scattering. For scattering angles of about 82 deg (where the best contrast for polarization measurement is theoretically predicted) we find that the single event resolution for the determination of the polarization is about 40 deg (predominantly due to properties of the Compton effect). However, for samples larger than ten thousand events the J-PET is capable of determining relative average polarization of these photons with the precision of about few degrees. The obtained results open new perspectives for studies of various physics phenomena such as quantum entanglement and tests of discrete symmetries in decays of positronium and extend the energy range of polarization measurements by five orders of magnitude beyond the optical wavelength regime.
△ Less
Submitted 27 September, 2018;
originally announced September 2018.