Correction of high-order phase variation effects in dynamic field monitoring
Authors:
Paul I. Dubovan,
Kyle M. Gilbert,
Corey A. Baron
Abstract:
Purpose: Field monitoring measures field perturbations, which can be accounted for during image reconstructions. In certain field monitoring environments, significant phase deviations can arise far from isocenter due to the finite extent of the gradient and/or main magnet. This can degrade the accuracy of field dynamics when field probes are placed near or outside the diameter spherical volume of…
▽ More
Purpose: Field monitoring measures field perturbations, which can be accounted for during image reconstructions. In certain field monitoring environments, significant phase deviations can arise far from isocenter due to the finite extent of the gradient and/or main magnet. This can degrade the accuracy of field dynamics when field probes are placed near or outside the diameter spherical volume of the gradient coils and/or main magnet, leading to corrupted image quality. The objective of this work was to develop a correction algorithm that reduces errors from highly nonlinear phase variations at distant field probes in field dynamic fits. Methods: The algorithm is split into three components. Component one fits phase coefficients one spatial order at a time, while the second implements a weighted least squares solution based on probe distance. After initial fitting, component three calculates phase residuals and removes the phase for distant probes before re-fitting. Two healthy volunteers were scanned on a head-only 7T MRI using diffusion-weighted single-shot spiral and EPI sequences and field monitoring was performed. Images were reconstructed with and without phase coefficient correction and compared qualitatively. Results: The algorithm was able to correct corrupted field dynamics, resulting in image quality improvements. Significant artefact reduction was observed when correcting higher order fits, especially for diffusion weighted images. Stepwise fitting provided the most correction benefit, which was marginally improved when adding weighted least squares and phase residual corrections. Conclusion: The proposed algorithm can mitigate effects of phase errors in field monitoring, providing improved reliability of field dynamic characterization.
△ Less
Submitted 23 January, 2023;
originally announced January 2023.
High-resolution single-shot spiral diffusion-weighted imaging at 7T using expanded encoding with compressed sensing
Authors:
Gabriel Varela-Mattatall,
Paul I. Dubovan,
Tales Santini,
Kyle M. Gilbert,
Ravi S. Menon,
Corey A. Baron
Abstract:
Purpose: The expanded encoding model incorporates spatially- and time-varying field perturbations for correction during reconstruction. So far, these reconstructions have used the conjugate gradient method with early stopping used as implicit regularization. However, this approach is likely suboptimal for low-SNR cases like diffusion or high-resolution MRI. Here, we investigate the extent that l1-…
▽ More
Purpose: The expanded encoding model incorporates spatially- and time-varying field perturbations for correction during reconstruction. So far, these reconstructions have used the conjugate gradient method with early stopping used as implicit regularization. However, this approach is likely suboptimal for low-SNR cases like diffusion or high-resolution MRI. Here, we investigate the extent that l1-wavelet regularization, or equivalently compressed sensing (CS), combined with expanded encoding improves trade-offs between spatial resolution, readout time and SNR for single-shot spiral diffusion-weighted imaging at 7T. The reconstructions were performed using our open-source GPU-enabled reconstruction toolbox, MatMRI, that allows inclusion of the different components of the expanded encoding model, with or without CS. Methods: In vivo accelerated single-shot spirals were acquired with five acceleration factors (2-6) and three in-plane spatial resolutions (1.5, 1.3, and 1.1 mm). From the in vivo reconstructions, we estimated diffusion tensors and computed fractional anisotropy maps. Then, simulations were used to quantitatively investigate and validate the impact of CS-based regularization on image quality when compared to a known ground truth. Results: In vivo reconstructions revealed improved image quality with retainment of small features when CS was used. Simulations showed that the joint use of the expanded encoding model and CS improves accuracy of image reconstructions (reduced mean-squared error) over the range of acceleration factors investigated. Conclusion: The expanded encoding model and CS regularization are complementary tools for single-shot spiral diffusion MRI, which enables both higher spatial resolutions and higher acceleration factors.
△ Less
Submitted 14 November, 2022;
originally announced November 2022.