Characterization of a TES-based Anti-Coincidence Detector for Future Large Field-of-View X-ray Calorimetry Missions
Authors:
Samuel V. Hull,
Joseph S. Adams,
Simon R. Bandler,
Matthew Cherry,
James A. Chervenak,
Renata Cumbee,
Xavier Defay,
Enectali Figueroa-Feliciano,
Fred M. Finkbeiner,
Joshua Fuhrman,
Richard L. Kelley,
Christopher Kenney,
Caroline A. Kilbourne,
Noah Kurinsky,
Jennette Mateo,
Haruka Muramatsu,
Frederick S. Porter,
Kazuhiro Sakai,
Aviv Simchony,
Stephen J. Smith,
Zoe Smith,
Nicholas A. Wakeham,
Edward J. Wassell,
Sang H. Yoon,
Betty A. Young
Abstract:
Microcalorimeter instruments aboard future X-ray observatories will require an anti-coincidence (anti-co) detector to veto charged particle events and reduce the non-X-ray background. We have developed a large-format, TES-based prototype anti-coincidence detector that is particularly suitable for use with spatially-extended (~ 10 cm^2}) TES microcalorimeter arrays, as would be used for a future la…
▽ More
Microcalorimeter instruments aboard future X-ray observatories will require an anti-coincidence (anti-co) detector to veto charged particle events and reduce the non-X-ray background. We have developed a large-format, TES-based prototype anti-coincidence detector that is particularly suitable for use with spatially-extended (~ 10 cm^2}) TES microcalorimeter arrays, as would be used for a future large field-of-view X-ray missions. This prototype was developed in the context of the Line Emission Mapper (LEM) probe concept, which required a ~ 14 cm^2 anti-co detector with > 95% live time and a low-energy threshold below 20 keV. Our anti-co design employs parallel networks of quasiparticle-trap-assisted electrothermal feedback TESs (QETs) to detect the athermal phonon signal produced in the detector substrate by incident charged particles. We developed multiple prototype anti-co designs featuring 12 channels and up to 6300 QETs. Here we focus on a design utilizing tungsten TESs and present characterization results. Broad energy range measurements have been performed (4.1 keV -- 5.5 MeV). Based on noise and responsivity measurements, the implied low-energy threshold is < 1 keV and a live time fraction of > 96% can be achieved up to 5.5 MeV. We also find evidence of mm-scale-or-better spatial resolution and discuss the potential utility of this for future missions. Finally, we discuss the early development of a soild-state physics model of the anti-co towards understanding phonon propagation and quasiparticle production in the detector.
△ Less
Submitted 19 February, 2025;
originally announced February 2025.
-
Micro-X Sounding Rocket Payload Re-flight Progress
Authors:
J. S. Adams,
S. R. Bandler,
N. Bastidon,
M. E. Eckart,
E. Figueroa-Feliciano,
J. Fuhrman,
D. C. Goldfinger,
A. J. F. Hubbard,
D. Jardin,
R. L. Kelley,
C. A. Kilbourne,
R. E. Manzagol-Harwood,
D. McCammon,
T. Okajima,
F. S. Porter,
C. D. Reintsema,
S. J. Smith
Abstract:
Micro-X is an X-ray sounding rocket payload that had its first flight on July 22, 2018. The goals of the first flight were to operate a transition edge sensor (TES) X-ray microcalorimeter array in space and take a high-resolution spectrum of the Cassiopeia A supernova remnant. The first flight was considered a partial success. The array and its time-division multiplexing readout system were succes…
▽ More
Micro-X is an X-ray sounding rocket payload that had its first flight on July 22, 2018. The goals of the first flight were to operate a transition edge sensor (TES) X-ray microcalorimeter array in space and take a high-resolution spectrum of the Cassiopeia A supernova remnant. The first flight was considered a partial success. The array and its time-division multiplexing readout system were successfully operated in space, but due to a failure in the attitude control system, no time on-target was acquired. A re-flight has been scheduled for summer 2022. Since the first flight, modifications have been made to the detector systems to improve noise and reduce the susceptibility to magnetic fields. The three-stage SQUID circuit, NIST MUX06a, has been replaced by a two-stage SQUID circuit, NIST MUX18b. The initial laboratory results for the new detector system will be presented in this paper.
△ Less
Submitted 12 November, 2021;
originally announced November 2021.
-
First operation of Transition-Edge Sensors in space with the Micro-X sounding rocket
Authors:
J. S. Adams,
R. Baker,
S. R. Bandler,
N. Bastidon,
M. E. Danowski,
W. B. Doriese,
M. E. Eckart,
E. Figueroa-Feliciano,
J. Fuhrman,
D. C. Goldfinger,
S. N. T. Heine,
G. C. Hilton,
A. J. F. Hubbard,
D. Jardin,
R. L. Kelley,
C. A. Kilbourne,
R. E. Manzagol-Harwood,
D. McCammon,
T. Okajima,
F. S. Porter,
C. D. Reintsema,
P. Serlemitsos,
S. J. Smith,
P. Wikus
Abstract:
With its first flight in 2018, Micro-X became the first program to fly Transition-Edge Sensors and their SQUID readouts in space. The science goal was a high-resolution, spatially resolved X-ray spectrum of the Cassiopeia A Supernova Remnant. While a rocket pointing error led to no time on target, the data was used to demonstrate the flight performance of the instrument. The detectors observed X-r…
▽ More
With its first flight in 2018, Micro-X became the first program to fly Transition-Edge Sensors and their SQUID readouts in space. The science goal was a high-resolution, spatially resolved X-ray spectrum of the Cassiopeia A Supernova Remnant. While a rocket pointing error led to no time on target, the data was used to demonstrate the flight performance of the instrument. The detectors observed X-rays from the on-board calibration source, but a susceptibility to external magnetic fields limited their livetime. Accounting for this, no change was observed in detector response between ground operation and flight operation. This paper provides an overview of the first flight performance and focuses on the upgrades made in preparation for reflight. The largest changes have been upgrading the SQUIDs to mitigate magnetic susceptibility, synchronizing the clocks on the digital electronics to minimize beat frequencies, and replacing the mounts between the cryostat and the rocket skin to improve mechanical integrity. As the first flight performance was consistent with performance on the ground, reaching the instrument goals in the laboratory is considered a strong predictor of future flight performance.
△ Less
Submitted 3 March, 2021;
originally announced March 2021.