Calculation of magnetic forces and torques on the Kibble coil
Authors:
Stephan Schlamminger,
Lorenz Keck,
Frank Seifert,
Leon S. Chao,
Darine Haddad,
Shisong Li
Abstract:
Analytically the forces and torques on a coil in a field of magnetic flux density can be calculated one of two ways. The line integral can be conducted along the coil's wire, summing up the differential force contribution. For each differential line segment, the force is obtained as a cross product with the magnetic flux density. Alternatively, the coil's energy in the field is the current times t…
▽ More
Analytically the forces and torques on a coil in a field of magnetic flux density can be calculated one of two ways. The line integral can be conducted along the coil's wire, summing up the differential force contribution. For each differential line segment, the force is obtained as a cross product with the magnetic flux density. Alternatively, the coil's energy in the field is the current times the flux threading the coil. Hence, the energy is obtained by executing a surface integral over the coil's open surface. Here, a dot product of the differential surface element with the magnetic flux density is executed under the integral sign. The forces and torques can then be obtained from the negative derivative of the energy with respect to the appropriate coordinate. For yoke-based Kibble balances, the latter method is much simpler since most of the flux is contained in the inner yoke of the magnet and can be written as a simple equation. Here, we use this method to provide simple equations and their results for finding the torques and forces that act on a coil in a yoke-based magnet system. We further introduce a simple method that allows the calculation of the position and orientation difference between the coil and the magnet from three measurements.
△ Less
Submitted 19 September, 2022; v1 submitted 25 August, 2022;
originally announced August 2022.
-
Magnet system for the Quantum Electro-Mechanical Metrology Suite
Authors:
Rafael R. Marangoni,
Darine Haddad,
Frank Seifert,
Leon S. Chao,
David B. Newell,
Stephan Schlamminger
Abstract:
The design of the permanent magnet system for the new Quantum Electro-Mechanical Metrology Suite (QEMMS) is described. The QEMMS, developed at the National Institute of Standards and Technology (NIST), consists of a Kibble balance, a programmable Josephson voltage standard, and a quantum Hall resistance standard. It will be used to measure masses up to $100\,\mathrm{g}$ with relative uncertainties…
▽ More
The design of the permanent magnet system for the new Quantum Electro-Mechanical Metrology Suite (QEMMS) is described. The QEMMS, developed at the National Institute of Standards and Technology (NIST), consists of a Kibble balance, a programmable Josephson voltage standard, and a quantum Hall resistance standard. It will be used to measure masses up to $100\,\mathrm{g}$ with relative uncertainties below $2\times 10^{-8}$. The magnet system is based on the design of the NIST-4 magnet system with significant changes to adopt to a smaller Kibble balance and to overcome known practical limitations. Analytical models are provided to describe the coil-current effect and model the forces required to split the magnet in two parts to install the coil. Both models are compared to simulation results obtained with finite element analysis and measurement results. Other aspects, such as the coil design and flatness of $Bl$ profile are considered.
△ Less
Submitted 26 December, 2019; v1 submitted 10 December, 2019;
originally announced December 2019.
-
Measurement of the Planck constant at the National Institute of Standards and Technology from 2015 to 2017
Authors:
D. Haddad,
F. Seifert,
L. S. Chao,
A. Possolo,
D. B. Newell,
J. R. Pratt,
C. J. Williams,
S. Schlamminger
Abstract:
Researchers at the National Institute of Standards and Technology(NIST) have measured the value of the Planck constant to be $h =6.626\,069\,934(89)\times 10^{-34}\,$J$\,$s (relative standard uncertainty $13\times 10^{-9}$). The result is based on over 10$\,$000 weighings of masses with nominal values ranging from 0.5$\,$kg to 2$\,$kg with the Kibble balance NIST-4. The uncertainty has been reduce…
▽ More
Researchers at the National Institute of Standards and Technology(NIST) have measured the value of the Planck constant to be $h =6.626\,069\,934(89)\times 10^{-34}\,$J$\,$s (relative standard uncertainty $13\times 10^{-9}$). The result is based on over 10$\,$000 weighings of masses with nominal values ranging from 0.5$\,$kg to 2$\,$kg with the Kibble balance NIST-4. The uncertainty has been reduced by more than twofold relative to a previous determination because of three factors: (1) a much larger data set than previously available, allowing a more realistic, and smaller, Type A evaluation; (2) a more comprehensive measurement of the back action of the weighing current on the magnet by weighing masses up to 2$\,$kg, decreasing the uncertainty associated with magnet non-linearity; (3) a rigorous investigation of the dependence of the geometric factor on the coil velocity reducing the uncertainty assigned to time-dependent leakage of current in the coil.
△ Less
Submitted 8 August, 2017;
originally announced August 2017.
-
First measurements of the flux integral with the NIST-4 watt balance
Authors:
D. Haddad,
F. Seifert,
L. S. Chao,
A. Cao,
G. Sineriz,
J. R. Pratt,
D. B. Newell,
S. Schlamminger
Abstract:
In early 2014, construction of a new watt balance, named NIST-4, has started at the National Institute of Standards and Technology (NIST). In a watt balance, the gravitational force of an unknown mass is compensated by an electromagnetic force produced by a coil in a magnet system. The electromagnetic force depends on the current in the coil and the magnetic flux integral. Most watt balances featu…
▽ More
In early 2014, construction of a new watt balance, named NIST-4, has started at the National Institute of Standards and Technology (NIST). In a watt balance, the gravitational force of an unknown mass is compensated by an electromagnetic force produced by a coil in a magnet system. The electromagnetic force depends on the current in the coil and the magnetic flux integral. Most watt balances feature an additional calibration mode, referred to as velocity mode, which allows one to measure the magnetic flux integral to high precision. In this article we describe first measurements of the flux integral in the new watt balance. We introduce measurement and data analysis techniques to assess the quality of the measurements and the adverse effects of vibrations on the instrument.
△ Less
Submitted 16 March, 2015;
originally announced March 2015.
-
A summary of the Planck constant measurements using a watt balance with a superconducting solenoid at NIST
Authors:
S. Schlamminger,
R. L. Steiner,
D. Haddad,
D. B. Newell,
F. Seifert,
L. S. Chao,
R. Liu,
E. R. Williams,
J. R. Pratt
Abstract:
Researchers at the National Institute of Standards and Technology have been using a watt balance, NIST-3, to measure the Planck constant $h$ for over ten years. Two recently published values disagree by more than one standard uncertainty. The motivation for the present manuscript is twofold. First, we correct the latest published number to take into account a recently discovered systematic error i…
▽ More
Researchers at the National Institute of Standards and Technology have been using a watt balance, NIST-3, to measure the Planck constant $h$ for over ten years. Two recently published values disagree by more than one standard uncertainty. The motivation for the present manuscript is twofold. First, we correct the latest published number to take into account a recently discovered systematic error in mass dissemination at the Bureau International des Poids et Mesures (BIPM). Second, we provide guidance on how to combine the two numbers into one final result. In order to adequately reflect the discrepancy, we added an additional systematic uncertainty to the published uncertainty budgets. The final value of $h$ measured with NIST-3 is $h = 6.626\,069\,36(37)\times 10^{-34}\,\mbox{J\,s}$. This result is $77(57) \times 10^{-9}$ fractionally higher than $h_{\mathrm{90}}$. Each number in parentheses gives the value of the standard uncertainty in the last two digits of the respective value and $h_{\mathrm{90}}$ is the conventional value of the Planck constant given by $h_{\mathrm{90}}\equiv 4 /( K_{\mathrm{J-90}}^2 R_{\mathrm{K-90}})$, where $K_{\mathrm{J-90}}$ and $R_{\mathrm{K-90}}$ denote the conventional values of the Josephson and von Klitzing constants, respectively.
△ Less
Submitted 27 January, 2015;
originally announced January 2015.
-
A determination of the local acceleration of gravity for the NIST-4 watt balance
Authors:
E. J. Leaman,
D. Haddad,
F. Seifert,
L. S. Chao,
A. Cao,
J. R. Pratt,
S. Schlamminger,
D. B. Newell
Abstract:
A new watt balance is being constructed at the National Institute of Standards and Technology (NIST) in preparation for the redefinition of the International System of Units and the realization of mass through an exact value of the Planck constant. The total relative uncertainty goal for this instrument of a few parts in $10^{8}$ requires that the local acceleration due to gravity be known at the…
▽ More
A new watt balance is being constructed at the National Institute of Standards and Technology (NIST) in preparation for the redefinition of the International System of Units and the realization of mass through an exact value of the Planck constant. The total relative uncertainty goal for this instrument of a few parts in $10^{8}$ requires that the local acceleration due to gravity be known at the location of a test mass with a relative uncertainty on the order of only a few parts in $10^{9}$. To make this determination, both the horizontal and vertical gradients of gravity must be known such that gravity may be tied from an absolute reference in the laboratory to the precise mass location. We describe the procedures used to model and measure gravity variations throughout the laboratory and give our results.
△ Less
Submitted 12 December, 2014;
originally announced December 2014.
-
A LEGO Watt Balance: An apparatus to determine a mass based on the new SI
Authors:
L. S. Chao,
S. Schlamminger,
D. B. Newell,
J. R. Pratt,
F. Seifert,
X. Zhang,
G. Sineriz,
M. Liu,
D. Haddad
Abstract:
A global effort to redefine our International System of Units (SI) is underway and the change to the new system is expected to occur in 2018. Within the newly redefined SI, the present base units will still exist but be derived from fixed numerical values of seven reference constants. More specifically, the unit of mass, the kilogram, will be realized through a fixed value of the Planck constant…
▽ More
A global effort to redefine our International System of Units (SI) is underway and the change to the new system is expected to occur in 2018. Within the newly redefined SI, the present base units will still exist but be derived from fixed numerical values of seven reference constants. More specifically, the unit of mass, the kilogram, will be realized through a fixed value of the Planck constant $h$. For instance, a watt balance can be used to realize the kilogram unit of mass within a few parts in $10^8$. Such a balance has been designed and constructed at the National Institute of Standards and Technology. For educational outreach and to demonstrate the principle, we have constructed a LEGO tabletop watt balance capable of measuring a gram size mass to 1 % relative uncertainty. This article presents the design, construction, and performance of the LEGO watt balance and its ability to determine $h$
△ Less
Submitted 7 August, 2015; v1 submitted 4 December, 2014;
originally announced December 2014.
-
Determination of the Planck constant using a watt balance with a superconducting magnet system at the National Institute of Standards and Technology
Authors:
Stephan Schlamminger,
Darine Haddad,
Frank Seifert,
Leon S Chao,
David B Newell,
Ruimin Liu,
Richard L Steiner,
Jon R Pratt
Abstract:
For the past two years, measurements have been performed with a watt balance at the National Institute of Standards and Technology (NIST) to determine the Planck constant. A detailed analysis of these measurements and their uncertainties has led to the value $h=6.626\,069\,79(30)\times 10^{-34}\,$J$\,$s. The relative standard uncertainty is $ 45\times 10^{-9}$. This result is $141\times 10^{-9}$ f…
▽ More
For the past two years, measurements have been performed with a watt balance at the National Institute of Standards and Technology (NIST) to determine the Planck constant. A detailed analysis of these measurements and their uncertainties has led to the value $h=6.626\,069\,79(30)\times 10^{-34}\,$J$\,$s. The relative standard uncertainty is $ 45\times 10^{-9}$. This result is $141\times 10^{-9}$ fractionally higher than $h_{90}$. Here $h_{90}$ is the conventional value of the Planck constant given by $h_{90}\equiv 4 /( K_{\mathrm{J-90}}^2R_{\mathrm{K-90}})$, where $K_{\mathrm{J-90}}$ and $R_{\mathrm{K-90}}$ denote the conventional values of the Josephson and von Klitzing constants, respectively.
△ Less
Submitted 24 April, 2014; v1 submitted 31 January, 2014;
originally announced January 2014.