First application of superconducting transition-edge-sensor microcalorimeters to hadronic-atom x-ray spectroscopy
Authors:
S. Okada,
D. A. Bennett,
C. Curceanu,
W. B. Doriese,
J. W. Fowler,
J. Gard,
F. P. Gustafsson,
T. Hashimoto,
R. S. Hayano,
S. Hirenzaki,
J. P. Hays-Wehle,
G. C. Hilton,
N. Ikeno,
M. Iliescu,
S. Ishimoto,
K. Itahashi,
M. Iwasaki,
T. Koike,
K. Kuwabara,
Y. Ma,
J. Marton,
H. Noda,
G. C. O'Neil,
H. Outa,
C. D. Reintsema
, et al. (13 additional authors not shown)
Abstract:
High-resolution pionic-atom x-ray spectroscopy was performed with an x-ray spectrometer based on a 240-pixel array of superconducting transition-edge-sensor (TES) microcalorimeters at the piM1 beam line of the Paul Scherrer Institute. X-rays emitted by pionic carbon via the 4f->3d transition and the parallel 4d->3p transition were observed with a full-width-at-half-maximum energy resolution of 6.8…
▽ More
High-resolution pionic-atom x-ray spectroscopy was performed with an x-ray spectrometer based on a 240-pixel array of superconducting transition-edge-sensor (TES) microcalorimeters at the piM1 beam line of the Paul Scherrer Institute. X-rays emitted by pionic carbon via the 4f->3d transition and the parallel 4d->3p transition were observed with a full-width-at-half-maximum energy resolution of 6.8 eV at 6.4 keV. Measured x-ray energies are consistent with calculated electromagnetic values which considered the strong-interaction effect assessed via the Seki-Masutani potential for the 3p energy level, and favor the electronic population of two filled 1s electrons in the K-shell. Absolute energy calibration with an uncertainty of 0.1 eV was demonstrated under a high-rate hadron beam condition of 1.45 MHz. This is the first application of a TES spectrometer to hadronic-atom x-ray spectroscopy and is an important milestone towards next-generation high-resolution kaonic-atom x-ray spectroscopy.
△ Less
Submitted 18 August, 2016;
originally announced August 2016.