Hit time and hit position reconstruction in the J-PET detector based on a library of averaged model signals
Authors:
P. Moskal,
N. G. Sharma,
M. Silarski,
T. Bednarski,
P. Białas,
J. Bułka,
E. Czerwiński,
A. Gajos,
D. Kamińska,
L. Kapłon,
A. Kochanowski,
G. Korcyl,
J. Kowal,
P. Kowalski,
T. Kozik,
W. Krzemień,
E. Kubicz,
Sz. Niedźwiecki,
M. Pałka,
L. Raczyński,
Z. Rudy,
O. Rundel,
P. Salabura,
A. Słomski,
J. Smyrski
, et al. (6 additional authors not shown)
Abstract:
In this article we present a novel method of hit time and hit position reconstruction in long scintillator detectors. We take advantage of the fact that for this kind of detectors amplitude and shape of registered signals depends strongly on the position where particle hit the detector. The reconstruction is based on determination of the degree of similarity between measured and averaged signals s…
▽ More
In this article we present a novel method of hit time and hit position reconstruction in long scintillator detectors. We take advantage of the fact that for this kind of detectors amplitude and shape of registered signals depends strongly on the position where particle hit the detector. The reconstruction is based on determination of the degree of similarity between measured and averaged signals stored in a library for a set of well-defined positions along the scintillator. Preliminary results of validation of the introduced method with experimental data obtained by means of the double strip prototype of the J-PET detector are presented.
△ Less
Submitted 6 August, 2015; v1 submitted 27 February, 2015;
originally announced February 2015.
-
GPU accelerated image reconstruction in a two-strip J-PET tomograph
Authors:
P. Białas,
J. Kowal,
A. Strzelecki,
T. Bednarski,
E. Czerwiński,
A. Gajos,
D. Kamińska,
Ł. Kapłon,
A. Kochanowski,
G. Korcyl,
P. Kowalski,
T. Kozik,
W. Krzemień,
E. Kubicz,
P. Moskal,
Sz. Niedźwiecki,
M. Pałka,
L. Raczyński,
Z. Rudy,
O. Rundel,
P. Salabura,
N. G. Sharma,
M. Silarski,
A. Słomski,
J. Smyrski
, et al. (4 additional authors not shown)
Abstract:
We present a fast GPU implementation of the image reconstruction routine, for a novel two strip PET detector that relies solely on the time of flight measurements.
We present a fast GPU implementation of the image reconstruction routine, for a novel two strip PET detector that relies solely on the time of flight measurements.
△ Less
Submitted 6 August, 2015; v1 submitted 26 February, 2015;
originally announced February 2015.
-
A pilot study of the novel J-PET plastic scintillator with 2-(4-styrylphenyl)benzoxazole as a wavelength shifter
Authors:
A. Wieczorek,
P. Moskal,
Sz. Niedźwiecki,
T. Bednarski,
P. Białas,
E. Czerwiński,
A. Danel,
A. Gajos,
A. Gruntowski,
D. Kamińska,
Ł. Kapłon,
A. Kochanowski,
G. Korcyl,
J. Kowal,
P. Kowalski,
T. Kozik,
W. Krzemień,
E. Kubicz,
M. Molenda,
M. Pałka,
L. Raczyński,
Z. Rudy,
O. Rundel,
P. Salabura,
N. G. Sharma
, et al. (8 additional authors not shown)
Abstract:
For the first time a molecule of 2-(4-styrylphenyl)benzoxazole containing benzoxazole and stilbene groups is applied as a scintillator dopant acting as a wavelength shifter. In this article a light yield of the plastic scintillator, prepared from styrene doped with 2 wt% of 2,5-diphenylbenzoxazole and 0.03 wt% of 2-(4-styrylphenyl)benzoxazole, is determined to be as large as 60% $\pm$ 2% of the an…
▽ More
For the first time a molecule of 2-(4-styrylphenyl)benzoxazole containing benzoxazole and stilbene groups is applied as a scintillator dopant acting as a wavelength shifter. In this article a light yield of the plastic scintillator, prepared from styrene doped with 2 wt% of 2,5-diphenylbenzoxazole and 0.03 wt% of 2-(4-styrylphenyl)benzoxazole, is determined to be as large as 60% $\pm$ 2% of the anthracene light output. There is a potential to improve this value in the future by the optimization of the additives concentrations.
△ Less
Submitted 4 July, 2015; v1 submitted 10 February, 2015;
originally announced February 2015.
-
A novel method for the line-of-response and time-of-flight reconstruction in TOF-PET detectors based on a library of synchronized model signals
Authors:
P. Moskal,
N. Zoń,
T. Bednarski,
P. Białas,
E. Czerwiński,
A. Gajos,
D. Kamińska,
Ł. Kapłon,
A. Kochanowski,
G. Korcyl,
J. Kowal,
P. Kowalski,
T. Kozik,
W. Krzemień,
E. Kubicz,
Sz. Niedźwiecki,
M. Pałka,
L. Raczyński,
Z. Rudy,
O. Rundel,
P. Salabura,
N. G. Sharma,
M. Silarski,
A. Słomski,
J. Smyrski
, et al. (4 additional authors not shown)
Abstract:
A novel method of hit time and hit position reconstruction in scintillator detectors is described. The method is based on comparison of detector signals with results stored in a library of synchronized model signals registered for a set of well-defined positions of scintillation points. The hit position is reconstructed as the one corresponding to the signal from the library which is most similar…
▽ More
A novel method of hit time and hit position reconstruction in scintillator detectors is described. The method is based on comparison of detector signals with results stored in a library of synchronized model signals registered for a set of well-defined positions of scintillation points. The hit position is reconstructed as the one corresponding to the signal from the library which is most similar to the measurement signal. The time of the interaction is determined as a relative time between the measured signal and the most similar one in the library. A degree of similarity of measured and model signals is defined as the distance between points representing the measurement- and model-signal in the multi-dimensional measurement space. Novelty of the method lies also in the proposed way of synchronization of model signals enabling direct determination of the difference between time-of-flights (TOF) of annihilation quanta from the annihilation point to the detectors. The introduced method was validated using experimental data obtained by means of the double strip prototype of the J-PET detector and $^{22}$Na sodium isotope as a source of annihilation gamma quanta.The detector was built out from plastic scintillator strips with dimensions of 5 mm x 19 mm x 300 mm, optically connected at both sides to photomultipliers,from which signals were sampled by means of the Serial Data Analyzer.Using the introduced method, the spatial and TOF resolution of about 1.3 cm ($σ$) and 125 ps ($σ$) were established, respectively.
△ Less
Submitted 10 February, 2015; v1 submitted 22 December, 2014;
originally announced December 2014.
-
Test of a single module of the J-PET scanner based on plastic scintillators
Authors:
P. Moskal,
Sz. Niedźwiecki,
T. Bednarski,
E. Czerwiński,
Ł. Kapłon,
E. Kubicz,
I. Moskal,
M. Pawlik-Niedźwiecka,
N. G. Sharma,
M. Silarski,
M. Zieliński,
N. Zoń,
P. Białas,
A. Gajos,
A. Kochanowski,
G. Korcyl,
J. Kowal,
P. Kowalski,
T. Kozik,
W. Krzemień,
M. Molenda,
M. Pałka,
L. Raczyński,
Z. Rudy,
P. Salabura
, et al. (5 additional authors not shown)
Abstract:
Time of Flight Positron Emission Tomography scanner based on plastic scintillators is being developed at the Jagiellonian University by the J-PET collaboration. The main challenge of the conducted research lies in the elaboration of a method allowing application of plastic scintillators for the detection of low energy gamma quanta. In this article we report on tests of a single detection module bu…
▽ More
Time of Flight Positron Emission Tomography scanner based on plastic scintillators is being developed at the Jagiellonian University by the J-PET collaboration. The main challenge of the conducted research lies in the elaboration of a method allowing application of plastic scintillators for the detection of low energy gamma quanta. In this article we report on tests of a single detection module built out from BC-420 plastic scintillator strip (with dimensions of 5x19x300mm^3) read out at two ends by Hamamatsu R5320 photomultipliers. The measurements were performed using collimated beam of annihilation quanta from the 68Ge isotope and applying the Serial Data Analyzer (Lecroy SDA6000A) which enabled sampling of signals with 50ps intervals. The time resolution of the prototype module was established to be better than 80ps (sigma) for a single level discrimination. The spatial resolution of the determination of the hit position along the strip was determined to be about 0.93cm (sigma) for the annihilation quanta. The fractional energy resolution for the energy E deposited by the annihilation quanta via the Compton scattering amounts to sigma(E)/E = 0.044/sqrt(E[MeV]) and corresponds to the sigma(E)/E of 7.5% at the Compton edge.
△ Less
Submitted 29 August, 2014; v1 submitted 28 July, 2014;
originally announced July 2014.
-
Calibration of photomultipliers gain used in the J-PET detector
Authors:
T. Bednarski,
E. Czerwiński,
P. Moskal,
P. Białas,
K. Giergiel,
Ł. Kapłon,
A. Kochanowski,
G. Korcyl,
J. Kowal,
P. Kowalski,
T. Kozik,
W. Krzemień,
M. Molenda,
I. Moskal,
Sz. Niedźwiecki,
M. Pałka,
M. Pawlik,
L. Raczyński,
Z. Rudy,
P. Salabura,
N. G. Sharma,
M. Silarski,
A. Słomski,
J. Smyrski,
A. Strzelecki
, et al. (5 additional authors not shown)
Abstract:
Photomultipliers are commonly used in commercial PET scanner as devices which convert light produced in scintillator by gamma quanta from positron-electron annihilation into electrical signal. For proper analysis of obtained electrical signal, a photomultiplier gain curve must be known, since gain can be significantly different even between photomultipliers of the same model. In this article we de…
▽ More
Photomultipliers are commonly used in commercial PET scanner as devices which convert light produced in scintillator by gamma quanta from positron-electron annihilation into electrical signal. For proper analysis of obtained electrical signal, a photomultiplier gain curve must be known, since gain can be significantly different even between photomultipliers of the same model. In this article we describe single photoelectron method used for photomultipliers calibration applied for J-PET scanner, a novel PET detector being developed at the Jagiellonian University. Description of calibration method, an example of calibration curve and gain of few R4998 Hamamatsu photomultipliers are presented.
△ Less
Submitted 16 December, 2013; v1 submitted 10 December, 2013;
originally announced December 2013.
-
Simulations of gamma quanta scattering in a single module of the J-PET detector
Authors:
K. Szymański,
P. Moskal,
T. Bednarski,
P. Białas,
E. Czerwiński,
K. Giergiel,
Ł. Kapłon,
A. Kochanowski,
G. Korcyl,
J. Kowal,
P. Kowalski,
T. Kozik,
W. Krzemień,
M. Molenda,
I. Moskal,
Sz. Niedźwiecki,
M. Pałka,
M. Pawlik,
L. Raczyński,
Z. Rudy,
P. Salabura,
N. G. Sharma,
M. Silarski,
A. Słomski,
J. Smyrski
, et al. (5 additional authors not shown)
Abstract:
This article describes simulations of scattering of annihilation gamma quanta in a strip of plastic scintillator. Such strips constitute basic detection modules in a newly proposed Positron Emission Tomography which utilizes plastic scintillators instead of inorganic crystals. An algorithm simulating chain of Compton scatterings was elaborated and series of simulations have been conducted for the…
▽ More
This article describes simulations of scattering of annihilation gamma quanta in a strip of plastic scintillator. Such strips constitute basic detection modules in a newly proposed Positron Emission Tomography which utilizes plastic scintillators instead of inorganic crystals. An algorithm simulating chain of Compton scatterings was elaborated and series of simulations have been conducted for the scintillator strip with the cross section of 5 mm x 19 mm. Obtained results indicate that secondary interactions occur only in the case of about 8% of events and out of them only 25$\%$ take place in the distance larger than 0.5 cm from the primary interaction. It was also established that light signals produced at primary and secondary interactions overlap with the delay which distribution is characterized by FWHM of about 40 ps.
△ Less
Submitted 1 December, 2013;
originally announced December 2013.
-
A novel method for calibration and monitoring of time synchronization of TOF-PET scanners by means of cosmic rays
Authors:
M. Silarski,
E. Czerwiński,
T. Bednarski,
P. Moskal,
P. Białas,
Ł. Kapłon,
A. Kochanowski,
G. Korcyl,
J. Kowal,
P. Kowalski,
T. Kozik,
W. Krzemień,
M. Molenda,
Sz. Niedźwiecki,
M. Pałka,
M. Pawlik,
L. Raczyński,
Z. Rudy,
P. Salabura,
N. G. Sharma,
A. Słomski,
J. Smyrski,
A. Strzelecki,
W. Wiślicki,
M. Zieliński
, et al. (1 additional authors not shown)
Abstract:
All of the present methods for calibration and monitoring of TOF-PET scanner detectors utilize radioactive isotopes such as e.g. $^{22}$Na or $^{68}$Ge, which are placed or rotate inside the scanner. In this article we describe a novel method based on the cosmic rays application to the PET calibration and monitoring methods. The concept allows to overcome many of the drawbacks of the present metho…
▽ More
All of the present methods for calibration and monitoring of TOF-PET scanner detectors utilize radioactive isotopes such as e.g. $^{22}$Na or $^{68}$Ge, which are placed or rotate inside the scanner. In this article we describe a novel method based on the cosmic rays application to the PET calibration and monitoring methods. The concept allows to overcome many of the drawbacks of the present methods and it is well suited for newly developed TOF-PET scanners with a large longitudinal field of view. The method enables also monitoring of the quality of the scintillator materials and in general allows for the continuous quality assurance of the PET detector performance.
△ Less
Submitted 7 March, 2014; v1 submitted 24 November, 2013;
originally announced November 2013.