Comparative studies of plastic scintillator strips with high technical attenuation length for the total-body J-PET scanner
Authors:
L. Kaplon,
J. Baran,
N. Chug,
A. Coussat,
C. Curceanu,
E. Czerwinski,
M. Dadgar,
K. Dulski,
J. Gajewski,
A. Gajos,
B. Hiesmayr,
E. Kavya Valsan,
K. Klimaszewski,
G. Korcyl,
T. Kozik,
W. Krzemien,
D. Kumar,
G. Moskal,
S. Niedzwiecki,
D. Panek,
S. Parzych,
E. Perez del Rio,
L. Raczynski,
A. Rucinski,
S. Sharma
, et al. (9 additional authors not shown)
Abstract:
Plastic scintillator strips are considered as one of the promising solutions for the cost-effective construction of total-body positron emission tomography, (PET) system. The purpose of the performed measurements is to compare the transparency of long plastic scintillators with dimensions 6 mm x 24 mm x 1000 mm and with all surfaces polished. Six different types of commercial, general purpose, blu…
▽ More
Plastic scintillator strips are considered as one of the promising solutions for the cost-effective construction of total-body positron emission tomography, (PET) system. The purpose of the performed measurements is to compare the transparency of long plastic scintillators with dimensions 6 mm x 24 mm x 1000 mm and with all surfaces polished. Six different types of commercial, general purpose, blue-emitting plastic scintillators with low attenuation of visible light were tested, namely: polyvinyl toluene-based BC-408, EJ-200, RP-408, and polystyrene-based Epic, SP32 and UPS-923A. For determination of the best type of plastic scintillator for totalbody Jagiellonian positron emission tomograph (TB-J-PET) construction, emission and transmission spectra, and technical attenuation length (TAL) of blue light-emitting by the scintillators were measured and compared. The TAL values were determined with the use of UV lamp as excitation source, and photodiode as light detector. Emission spectra of investigated scintillators have maxima in the range from 420 nm to 429 nm. The BC-408 and EJ-200 have the highest transmittance values of about 90% at the maximum emission wavelength measured through a 6 mm thick scintillator strip and the highest technical attenuation length reaching about 2000 mm, allowing assembly of long detection modules for time-of-flight (TOF) J-PET scanners. Influence of the 6 mm x 6 mm, 12 mm x 6 mm, 24 mm x 6 mm cross-sections of the 1000 mm long EJ-200 plastic scintillator on the TAL and signal intensity was measured. The highest TAL value was determined for samples with 24 mm x 6 mm cross-section.
△ Less
Submitted 3 August, 2024; v1 submitted 28 July, 2024;
originally announced July 2024.
-
Discrete symmetries tested at 10$^{-4}$ precision using linear polarization of photons from positronium annihilations
Authors:
Paweł Moskal,
Eryk Czerwiński,
Juhi Raj,
Steven D. Bass,
Ermias Y. Beyene,
Neha Chug,
Aurélien Coussat,
Catalina Curceanu,
Meysam Dadgar,
Manish Das,
Kamil Dulski,
Aleksander Gajos,
Marek Gorgol,
Beatrix C. Hiesmayr,
Bożena Jasińska,
Krzysztof Kacprzak,
Tevfik Kaplanoglu,
Łukasz Kapłon,
Konrad Klimaszewski,
Paweł Konieczka,
Grzegorz Korcyl,
Tomasz Kozik,
Wojciech Krzemień,
Deepak Kumar,
Simbarashe Moyo
, et al. (16 additional authors not shown)
Abstract:
Discrete symmetries play an important role in particle physics with violation of CP connected to the matter-antimatter imbalance in the Universe. We report the most precise test of P, T and CP invariance in decays of ortho-positronium, performed with methodology involving polarization of photons from these decays. Positronium, the simplest bound state of an electron and positron, is of recent inte…
▽ More
Discrete symmetries play an important role in particle physics with violation of CP connected to the matter-antimatter imbalance in the Universe. We report the most precise test of P, T and CP invariance in decays of ortho-positronium, performed with methodology involving polarization of photons from these decays. Positronium, the simplest bound state of an electron and positron, is of recent interest with discrepancies reported between measured hyperfine energy structure and theory at the level of $10^{-4}$ signaling a need for better understanding of the positronium system at this level. We test discrete symmetries using photon polarizations determined via Compton scattering in the dedicated J-PET tomograph on an event-by-event basis and without the need to control the spin of the positronium with an external magnetic field, in contrast to previous experiments. Our result is consistent with QED expectations at the level of 0.0007 and one standard deviation.
△ Less
Submitted 22 January, 2024;
originally announced January 2024.
-
Feasibility studies for imaging e$^{+}$e$^{-}$ annihilation with modular multi-strip detectors
Authors:
S. Sharma,
L. Povolo,
S. Mariazzi,
G. Korcyl,
K. Kacprzak,
D. Kumar,
S. Niedzwiecki,
J. Baran,
E. Beyene,
R. S. Brusa,
R. Caravita,
N. Chug,
A. Coussat,
C. Curceanu,
E. Czerwinski,
M. Dadgar,
M. Das,
K. Dulski,
K. Eliyan,
A. Gajos,
N. Gupta,
B. C. Hiesmayr,
L. Kaplon,
T. Kaplanoglu,
K. Klimaszewski
, et al. (19 additional authors not shown)
Abstract:
Studies based on imaging the annihilation of the electron (e$^{-}$) and its antiparticle positron (e$^{+}$) open up several interesting applications in nuclear medicine and fundamental research. The annihilation process involves both the direct conversion of e$^{+}$e$^{-}$ into photons and the formation of their atomically bound state, the positronium atom (Ps), which can be used as a probe for fu…
▽ More
Studies based on imaging the annihilation of the electron (e$^{-}$) and its antiparticle positron (e$^{+}$) open up several interesting applications in nuclear medicine and fundamental research. The annihilation process involves both the direct conversion of e$^{+}$e$^{-}$ into photons and the formation of their atomically bound state, the positronium atom (Ps), which can be used as a probe for fundamental studies. With the ability to produce large quantities of Ps, manipulate them in long-lived Ps states, and image their annihilations after a free fall or after passing through atomic interferometers, this purely leptonic antimatter system can be used to perform inertial sensing studies in view of a direct test of Einstein equivalence principle. It is envisioned that modular multistrip detectors can be exploited as potential detection units for this kind of studies. In this work, we report the results of the first feasibility study performed on a e$^{+}$ beamline using two detection modules to evaluate their reconstruction performance and spatial resolution for imaging e$^{+}$e$^{-}$ annihilations and thus their applicability for gravitational studies of Ps.
△ Less
Submitted 12 September, 2023;
originally announced September 2023.
-
Testing CPT symmetry in ortho-positronium decays with positronium annihilation tomography
Authors:
Paweł Moskal,
Aleksander Gajos,
Muhsin Mohammed,
Jyoti Chhokar,
Neha Chug,
Catalina Curceanu,
Eryk Czerwiński,
Meysam Dadgar,
Kamil Dulski,
Marek Gorgol,
Jacek Goworek,
Beatrix Hiesmayr,
Bożena Jasińska,
Krzysztof Kacprzak,
Łukasz Kapłon,
Hanieh Karimi,
Daria Kisielewska,
Konrad Klimaszewski,
Grzegorz Korcyl,
Paweł Kowalski,
Nikodem Krawczyk,
Wojciech Krzemień,
Tomasz Kozik,
Ewelina Kubicz,
Szymon Niedźwiecki
, et al. (13 additional authors not shown)
Abstract:
Charged lepton system symmetry under combined charge, parity, and time-reversal transformation (CPT) remains scarcely tested. Despite stringent quantum-electrodynamic limits, discrepancies in predictions for the electron-positron bound state (positronium atom) motivate further investigation, including fundamental symmetry tests. While CPT noninvariance effects could be manifested in non-vanishing…
▽ More
Charged lepton system symmetry under combined charge, parity, and time-reversal transformation (CPT) remains scarcely tested. Despite stringent quantum-electrodynamic limits, discrepancies in predictions for the electron-positron bound state (positronium atom) motivate further investigation, including fundamental symmetry tests. While CPT noninvariance effects could be manifested in non-vanishing angular correlations between final-state photons and spin of annihilating positronium, measurements were previously limited by knowledge of the latter. Here, we demonstrate tomographic reconstruction techniques applied to three-photon annihilations of ortho-positronium atoms to estimate their spin polarisation without magnetic field or polarised positronium source. We use a plastic-scintillator-based positron-emission-tomography scanner to record ortho-positronium (o-Ps) annihilations with single-event estimation of o-Ps spin and determine the complete spectrum of an angular correlation operator sensitive to CPT-violating effects. We find no violation at the precision level of 10^{-4}, with an over threefold improvement on the previous measurement.
△ Less
Submitted 8 December, 2021;
originally announced December 2021.
-
The J-PET detector -- a tool for precision studies of ortho-positronium decays
Authors:
K. Dulski,
S. D. Bass,
J. Chhokar,
N. Chug,
C. Curceanu,
E. Czerwiński,
M. Dadgar,
J. Gajewski,
A. Gajos,
M. Gorgol,
R. Del Grande,
B. C. Hiesmayr,
B. Jasińska,
K. Kacprzak,
Ł. Kapłon,
H. Karimi,
D. Kisielewska,
K. Klimaszewski,
P. Kopka,
G. Korcyl,
P. Kowalski,
T. Kozik,
N. Krawczyk,
W. Krzemień,
E. Kubicz
, et al. (19 additional authors not shown)
Abstract:
The J-PET tomograph is constructed from plastic scintillator strips arranged axially in concentric cylindrical layers. It enables investigations of positronium decays by measurement of the time, position, polarization and energy deposited by photons in the scintillators, in contrast to studies conducted so far with crystal and semiconductor based detection systems where the key selection of events…
▽ More
The J-PET tomograph is constructed from plastic scintillator strips arranged axially in concentric cylindrical layers. It enables investigations of positronium decays by measurement of the time, position, polarization and energy deposited by photons in the scintillators, in contrast to studies conducted so far with crystal and semiconductor based detection systems where the key selection of events is based on the measurement of the photons energies. In this article we show that the J-PET tomograph system is capable of exclusive measurements of the decays of ortho-positronium atoms. We present the first positronium production results, its lifetime distribution measurements and discuss estimation of the influence of various background sources. The tomograph s performance demonstrated here makes it suitable for precision studies of positronium decays including entanglement of the final state photons, positron annihilation lifetime spectroscopy plus molecular imaging diagnostics.
△ Less
Submitted 25 May, 2021; v1 submitted 9 June, 2020;
originally announced June 2020.
-
Commissioning of the J-PET detector in view of the positron annihilation lifetime spectroscopy
Authors:
K. Dulski,
C. Curceanu,
E. Czerwiński,
A. Gajos,
M. Gorgol,
N. Gupta-Sharma,
B. C. Hiesmayr,
B. Jasińska,
K. Kacprzak Ł. Kapłon,
D. Kisielewska,
K. Klimaszewski,
G. Korcyl,
P. Kowalski,
N. Krawczyk,
W. Krzemień,
T. Kozik,
E. Kubicz,
M. Mohammed,
Sz. Niedźwiecki,
M. Pałka,
M. Pawlik-Niedźwiecka,
L. Raczyński,
J. Raj,
K. Rakoczy,
Z. Rudy
, et al. (8 additional authors not shown)
Abstract:
The Jagiellonian Positron Emission Tomograph (J-PET) is the first PET device built from plastic scintillators. It is a multi-purpose detector designed for medical imaging and for studies of properties of positronium atoms in porous matter and in living organisms. In this article we report on the commissioning of the J-PET detector in view of studies of positronium decays. We present results of ana…
▽ More
The Jagiellonian Positron Emission Tomograph (J-PET) is the first PET device built from plastic scintillators. It is a multi-purpose detector designed for medical imaging and for studies of properties of positronium atoms in porous matter and in living organisms. In this article we report on the commissioning of the J-PET detector in view of studies of positronium decays. We present results of analysis of the positron lifetime measured in the porous polymer. The obtained results prove that J-PET is capable of performing simultaneous imaging of the density distribution of annihilation points as well as positron annihilation lifetime spectroscopy.
△ Less
Submitted 12 September, 2018;
originally announced September 2018.
-
Analysis procedure of the positronium lifetime spectra for the J-PET detector
Authors:
K. Dulski,
B. Zgardzinska,
P. Bialas,
C. Curceanu E. Czerwinski,
A. Gajos,
B. Głowacz,
M. Gorgol,
B. C. Hiesmayr,
B. Jasinska,
D. Kisielewska-Kaminska,
G. Korcyl,
P. Kowalski,
T. Kozik,
N. Krawczyk,
W. Krzemiez,
E. Kubicz,
M. Mohammed,
M. Pawlik-Niedzwiecka,
S. Niedzwiecki,
M. Palka,
L. Raczynski,
J. Raj,
Z. Rudy,
N. G. Sharma,
S. Sharma
, et al. (8 additional authors not shown)
Abstract:
Positron Annihilation Lifetime Spectroscopy (PALS) has shown to be a powerful tool to study the nanostructures of porous materials. Positron Emissions Tomography (PET) are devices allowing imaging of metabolic processes e.g. in human bodies. A newly developed device, the J-PET (Jagiellonian PET), will allow PALS in addition to imaging, thus combining both analyses providing new methods for physics…
▽ More
Positron Annihilation Lifetime Spectroscopy (PALS) has shown to be a powerful tool to study the nanostructures of porous materials. Positron Emissions Tomography (PET) are devices allowing imaging of metabolic processes e.g. in human bodies. A newly developed device, the J-PET (Jagiellonian PET), will allow PALS in addition to imaging, thus combining both analyses providing new methods for physics and medicine. In this contribution we present a computer program that is compatible with the J-PET software. We compare its performance with the standard program LT 9.0 by using PALS data from hexane measurements at different temperatures. Our program is based on an iterative procedure, and our fits prove that it performs as good as LT 9.0.
△ Less
Submitted 16 October, 2017;
originally announced October 2017.
-
Time calibration of the J-PET detector
Authors:
M. Skurzok,
M. Silarski,
D. Alfs,
P. Bialas,
Shivani,
C. Curceanu,
E. Czerwinski,
K. Dulski,
A. Gajos,
B. Glowacz,
M. Gorgol,
B. C. Hiesmayr,
B. Jasinska,
D. Kisielewska-Kaminska,
G. Korcyl,
P. Kowalski,
T. Kozik,
N. Krawczyk,
W. Krzemien,
E. Kubicz,
M. Mohammed,
M. Pawlik-Niedzwiecka,
S. Niedzwiecki,
M. Palka,
L. Raczynski
, et al. (10 additional authors not shown)
Abstract:
The Jagiellonian Positron Emission Tomograph (J-PET) project carried out in the Institute of Physics of the Jagiellonian University is focused on construction and tests of the first prototype of PET scanner for medical diagnostic which allows for the simultaneous 3D imaging of the whole human body using organic scintillators. The J-PET prototype consists of 192 scintillator strips forming three cy…
▽ More
The Jagiellonian Positron Emission Tomograph (J-PET) project carried out in the Institute of Physics of the Jagiellonian University is focused on construction and tests of the first prototype of PET scanner for medical diagnostic which allows for the simultaneous 3D imaging of the whole human body using organic scintillators. The J-PET prototype consists of 192 scintillator strips forming three cylindrical layers which are optimized for the detection of photons from the electron-positron annihilation with high time- and high angular-resolutions. In this article we present time calibration and synchronization of the whole J-PET detection system by irradiating each single detection module with a 22Na source and a small detector providing common reference time for synchronization of all the modules.
△ Less
Submitted 16 October, 2017;
originally announced October 2017.