Calibration of photomultipliers gain used in the J-PET detector
Authors:
T. Bednarski,
E. Czerwiński,
P. Moskal,
P. Białas,
K. Giergiel,
Ł. Kapłon,
A. Kochanowski,
G. Korcyl,
J. Kowal,
P. Kowalski,
T. Kozik,
W. Krzemień,
M. Molenda,
I. Moskal,
Sz. Niedźwiecki,
M. Pałka,
M. Pawlik,
L. Raczyński,
Z. Rudy,
P. Salabura,
N. G. Sharma,
M. Silarski,
A. Słomski,
J. Smyrski,
A. Strzelecki
, et al. (5 additional authors not shown)
Abstract:
Photomultipliers are commonly used in commercial PET scanner as devices which convert light produced in scintillator by gamma quanta from positron-electron annihilation into electrical signal. For proper analysis of obtained electrical signal, a photomultiplier gain curve must be known, since gain can be significantly different even between photomultipliers of the same model. In this article we de…
▽ More
Photomultipliers are commonly used in commercial PET scanner as devices which convert light produced in scintillator by gamma quanta from positron-electron annihilation into electrical signal. For proper analysis of obtained electrical signal, a photomultiplier gain curve must be known, since gain can be significantly different even between photomultipliers of the same model. In this article we describe single photoelectron method used for photomultipliers calibration applied for J-PET scanner, a novel PET detector being developed at the Jagiellonian University. Description of calibration method, an example of calibration curve and gain of few R4998 Hamamatsu photomultipliers are presented.
△ Less
Submitted 16 December, 2013; v1 submitted 10 December, 2013;
originally announced December 2013.
-
Simulations of gamma quanta scattering in a single module of the J-PET detector
Authors:
K. Szymański,
P. Moskal,
T. Bednarski,
P. Białas,
E. Czerwiński,
K. Giergiel,
Ł. Kapłon,
A. Kochanowski,
G. Korcyl,
J. Kowal,
P. Kowalski,
T. Kozik,
W. Krzemień,
M. Molenda,
I. Moskal,
Sz. Niedźwiecki,
M. Pałka,
M. Pawlik,
L. Raczyński,
Z. Rudy,
P. Salabura,
N. G. Sharma,
M. Silarski,
A. Słomski,
J. Smyrski
, et al. (5 additional authors not shown)
Abstract:
This article describes simulations of scattering of annihilation gamma quanta in a strip of plastic scintillator. Such strips constitute basic detection modules in a newly proposed Positron Emission Tomography which utilizes plastic scintillators instead of inorganic crystals. An algorithm simulating chain of Compton scatterings was elaborated and series of simulations have been conducted for the…
▽ More
This article describes simulations of scattering of annihilation gamma quanta in a strip of plastic scintillator. Such strips constitute basic detection modules in a newly proposed Positron Emission Tomography which utilizes plastic scintillators instead of inorganic crystals. An algorithm simulating chain of Compton scatterings was elaborated and series of simulations have been conducted for the scintillator strip with the cross section of 5 mm x 19 mm. Obtained results indicate that secondary interactions occur only in the case of about 8% of events and out of them only 25$\%$ take place in the distance larger than 0.5 cm from the primary interaction. It was also established that light signals produced at primary and secondary interactions overlap with the delay which distribution is characterized by FWHM of about 40 ps.
△ Less
Submitted 1 December, 2013;
originally announced December 2013.
-
A novel method for calibration and monitoring of time synchronization of TOF-PET scanners by means of cosmic rays
Authors:
M. Silarski,
E. Czerwiński,
T. Bednarski,
P. Moskal,
P. Białas,
Ł. Kapłon,
A. Kochanowski,
G. Korcyl,
J. Kowal,
P. Kowalski,
T. Kozik,
W. Krzemień,
M. Molenda,
Sz. Niedźwiecki,
M. Pałka,
M. Pawlik,
L. Raczyński,
Z. Rudy,
P. Salabura,
N. G. Sharma,
A. Słomski,
J. Smyrski,
A. Strzelecki,
W. Wiślicki,
M. Zieliński
, et al. (1 additional authors not shown)
Abstract:
All of the present methods for calibration and monitoring of TOF-PET scanner detectors utilize radioactive isotopes such as e.g. $^{22}$Na or $^{68}$Ge, which are placed or rotate inside the scanner. In this article we describe a novel method based on the cosmic rays application to the PET calibration and monitoring methods. The concept allows to overcome many of the drawbacks of the present metho…
▽ More
All of the present methods for calibration and monitoring of TOF-PET scanner detectors utilize radioactive isotopes such as e.g. $^{22}$Na or $^{68}$Ge, which are placed or rotate inside the scanner. In this article we describe a novel method based on the cosmic rays application to the PET calibration and monitoring methods. The concept allows to overcome many of the drawbacks of the present methods and it is well suited for newly developed TOF-PET scanners with a large longitudinal field of view. The method enables also monitoring of the quality of the scintillator materials and in general allows for the continuous quality assurance of the PET detector performance.
△ Less
Submitted 7 March, 2014; v1 submitted 24 November, 2013;
originally announced November 2013.