Multi-Grid Detector for Neutron Spectroscopy: Results Obtained on Time-of-Flight Spectrometer CNCS
Authors:
M. Anastasopoulos,
R. Bebb,
K. Berry,
J. Birch,
T. Bryś,
J. -C. Buffet,
J. -F. Clergeau,
P. P. Deen,
G. Ehlers,
P. van Esch,
S. M. Everett,
B. Guerard,
R. Hall-Wilton,
K. Herwig,
L. Hultman,
C. Höglund,
I. Iruretagoiena,
F. Issa,
J. Jensen,
A. Khaplanov,
O. Kirstein,
I. Lopez-Higuera,
F. Piscitelli,
L. Robinson,
S. Schmidt
, et al. (1 additional authors not shown)
Abstract:
The Multi-Grid detector technology has evolved from the proof-of-principle and characterisation stages. Here we report on the performance of the Multi-Grid detector, the MG.CNCS prototype, which has been installed and tested at the Cold Neutron Chopper Spectrometer, CNCS at SNS. This has allowed a side-by-side comparison to the performance of $^3$He detectors on an operational instrument. The demo…
▽ More
The Multi-Grid detector technology has evolved from the proof-of-principle and characterisation stages. Here we report on the performance of the Multi-Grid detector, the MG.CNCS prototype, which has been installed and tested at the Cold Neutron Chopper Spectrometer, CNCS at SNS. This has allowed a side-by-side comparison to the performance of $^3$He detectors on an operational instrument. The demonstrator has an active area of 0.2 m$^2$. It is specifically tailored to the specifications of CNCS. The detector was installed in June 2016 and has operated since then, collecting neutron scattering data in parallel to the He-3 detectors of CNCS. In this paper, we present a comprehensive analysis of this data, in particular on instrument energy resolution, rate capability, background and relative efficiency. Stability, gamma-ray and fast neutron sensitivity have also been investigated. The effect of scattering in the detector components has been measured and provides input to comparison for Monte Carlo simulations. All data is presented in comparison to that measured by the $^3$He detectors simultaneously, showing that all features recorded by one detector are also recorded by the other. The energy resolution matches closely. We find that the Multi-Grid is able to match the data collected by $^3$He, and see an indication of a considerable advantage in the count rate capability. Based on these results, we are confident that the Multi-Grid detector will be capable of producing high quality scientific data on chopper spectrometers utilising the unprecedented neutron flux of the ESS.
△ Less
Submitted 3 April, 2017; v1 submitted 10 March, 2017;
originally announced March 2017.
-
A 10B-based neutron detector with stacked Multiwire Proportional Counters and macrostructured cathodes
Authors:
I. Stefanescu,
Y. Abdullahi,
J. Birch,
I. Defendi,
R. Hall-Wilton,
C. Hoglund,
L. Hultman,
M. Zee,
K. Zeitelhack
Abstract:
We present the results of the measurements of the detection efficiency for a 4.7 Å neutron beam incident upon a detector incorporating a stack of up to five MultiWire Proportional Counters (MWPC) with Boron-coated cathodes. The cathodes were made of Aluminum and had a surface exhibiting millimeter-deep V-shaped grooves of 45°, upon which the thin Boron film was deposited by DC magnetron sputtering…
▽ More
We present the results of the measurements of the detection efficiency for a 4.7 Å neutron beam incident upon a detector incorporating a stack of up to five MultiWire Proportional Counters (MWPC) with Boron-coated cathodes. The cathodes were made of Aluminum and had a surface exhibiting millimeter-deep V-shaped grooves of 45°, upon which the thin Boron film was deposited by DC magnetron sputtering. The incident neutrons interacting with the converter layer deposited on the sidewalls of the grooves have a higher capture probability, owing to the larger effective absorption film thickness. This leads to a higher overall detection efficiency for the grooved cathode when compared to a cathode with a flat surface. Both the experimental results and the predictions of the GEANT4 model suggests that a 5-counter detector stack with coated grooved cathodes has the same efficiency as a 7-counter stack with flat cathodes. The reduction in the number of counters in the stack without altering the detection efficiency will prove highly beneficial for large-area position-sensitive detectors for neutron scattering applications, for which the cost-effective manufacturing of the detector and associated readout electronics is an important objective. The proposed detector concept could be a technological option for one of the new chopper spectrometers and other instruments planned to be built at the future European Spallation Source in Sweden. These results with macrostructured cathodes generally apply not just to MWPCs but to other gaseous detectors as well.
△ Less
Submitted 30 September, 2013;
originally announced September 2013.
-
Multi-Grid Boron-10 detector for large area applications in neutron scattering science
Authors:
Ken Andersen,
Thierry Bigault,
Jens Birch,
Jean-Claude Buffet,
Jonathan Correa,
Patrick van Esch,
Bruno Guerard,
Richard Hall-Wilton,
Lars Hultman,
Carina Höglund,
Jens Jensen,
Anton Khaplanov,
Oliver Kirstein,
Francesco Piscitelli,
Christian Vettier
Abstract:
The present supply of 3He can no longer meet the detector demands of the upcoming ESS facility and continued detector upgrades at current neutron sources. Therefore viable alternative technologies are required to support the development of cutting-edge instrumentation for neutron scattering science. In this context, 10B-based detectors are being developed by collaboration between the ESS, ILL, and…
▽ More
The present supply of 3He can no longer meet the detector demands of the upcoming ESS facility and continued detector upgrades at current neutron sources. Therefore viable alternative technologies are required to support the development of cutting-edge instrumentation for neutron scattering science. In this context, 10B-based detectors are being developed by collaboration between the ESS, ILL, and Linköping University. This paper reports on progress of this technology and the prospects applying it in modern neutron scattering experiments. The detector is made-up of multiple rectangular gas counter tubes coated with B4C, enriched in 10B. An anode wire reads out each tube, thereby giving position of conversion in one of the lateral co-ordinates as well as in depth of the detector. Position resolution in the remaining co-ordinate is obtained by segmenting the cathode tube itself. Boron carbide films have been produced at Linköping University and a detector built at ILL. The characterization study is presented in this paper, including measurement of efficiency, effects of the fill gas species and pressure, coating thickness variation on efficiency and sensitivity to gamma-rays.
△ Less
Submitted 4 September, 2012;
originally announced September 2012.