Analysis procedure of the positronium lifetime spectra for the J-PET detector
Authors:
K. Dulski,
B. Zgardzinska,
P. Bialas,
C. Curceanu E. Czerwinski,
A. Gajos,
B. Głowacz,
M. Gorgol,
B. C. Hiesmayr,
B. Jasinska,
D. Kisielewska-Kaminska,
G. Korcyl,
P. Kowalski,
T. Kozik,
N. Krawczyk,
W. Krzemiez,
E. Kubicz,
M. Mohammed,
M. Pawlik-Niedzwiecka,
S. Niedzwiecki,
M. Palka,
L. Raczynski,
J. Raj,
Z. Rudy,
N. G. Sharma,
S. Sharma
, et al. (8 additional authors not shown)
Abstract:
Positron Annihilation Lifetime Spectroscopy (PALS) has shown to be a powerful tool to study the nanostructures of porous materials. Positron Emissions Tomography (PET) are devices allowing imaging of metabolic processes e.g. in human bodies. A newly developed device, the J-PET (Jagiellonian PET), will allow PALS in addition to imaging, thus combining both analyses providing new methods for physics…
▽ More
Positron Annihilation Lifetime Spectroscopy (PALS) has shown to be a powerful tool to study the nanostructures of porous materials. Positron Emissions Tomography (PET) are devices allowing imaging of metabolic processes e.g. in human bodies. A newly developed device, the J-PET (Jagiellonian PET), will allow PALS in addition to imaging, thus combining both analyses providing new methods for physics and medicine. In this contribution we present a computer program that is compatible with the J-PET software. We compare its performance with the standard program LT 9.0 by using PALS data from hexane measurements at different temperatures. Our program is based on an iterative procedure, and our fits prove that it performs as good as LT 9.0.
△ Less
Submitted 16 October, 2017;
originally announced October 2017.
-
Time calibration of the J-PET detector
Authors:
M. Skurzok,
M. Silarski,
D. Alfs,
P. Bialas,
Shivani,
C. Curceanu,
E. Czerwinski,
K. Dulski,
A. Gajos,
B. Glowacz,
M. Gorgol,
B. C. Hiesmayr,
B. Jasinska,
D. Kisielewska-Kaminska,
G. Korcyl,
P. Kowalski,
T. Kozik,
N. Krawczyk,
W. Krzemien,
E. Kubicz,
M. Mohammed,
M. Pawlik-Niedzwiecka,
S. Niedzwiecki,
M. Palka,
L. Raczynski
, et al. (10 additional authors not shown)
Abstract:
The Jagiellonian Positron Emission Tomograph (J-PET) project carried out in the Institute of Physics of the Jagiellonian University is focused on construction and tests of the first prototype of PET scanner for medical diagnostic which allows for the simultaneous 3D imaging of the whole human body using organic scintillators. The J-PET prototype consists of 192 scintillator strips forming three cy…
▽ More
The Jagiellonian Positron Emission Tomograph (J-PET) project carried out in the Institute of Physics of the Jagiellonian University is focused on construction and tests of the first prototype of PET scanner for medical diagnostic which allows for the simultaneous 3D imaging of the whole human body using organic scintillators. The J-PET prototype consists of 192 scintillator strips forming three cylindrical layers which are optimized for the detection of photons from the electron-positron annihilation with high time- and high angular-resolutions. In this article we present time calibration and synchronization of the whole J-PET detection system by irradiating each single detection module with a 22Na source and a small detector providing common reference time for synchronization of all the modules.
△ Less
Submitted 16 October, 2017;
originally announced October 2017.
-
Trilateration-based reconstruction of ortho-positronium decays into three photons with the J-PET detector
Authors:
A. Gajos,
D. Kamińska,
E. Czerwiński,
D. Alfs,
T. Bednarski,
P. Białas,
B. Głowacz,
M. Gorgol,
B. Jasińska,
Ł. Kapłon,
G. Korcyl,
P. Kowalski,
T. Kozik,
W. Krzemień,
E. Kubicz,
M. Mohammed,
Sz. Niedźwiecki,
M. Pałka,
M. Pawlik-Niedźwiecka,
L. Raczyński,
Z. Rudy,
O. Rundel,
N. G. Sharma,
M. Silarski,
A. Słomski
, et al. (6 additional authors not shown)
Abstract:
This work reports on a new reconstruction algorithm allowing to reconstruct the decays of ortho-positronium atoms into three photons using the places and times of photons recorded in the detector. The method is based on trilateration and allows for a simultaneous reconstruction of both location and time of the decay. Results of resolution tests of the new reconstruction in the J-PET detector based…
▽ More
This work reports on a new reconstruction algorithm allowing to reconstruct the decays of ortho-positronium atoms into three photons using the places and times of photons recorded in the detector. The method is based on trilateration and allows for a simultaneous reconstruction of both location and time of the decay. Results of resolution tests of the new reconstruction in the J-PET detector based on Monte Carlo simulations are presented, which yield a spatial resolution at the level of 2 cm (FWHM) for X and Y and at the level of 1 cm (FWHM) for Z available with the present resolution of J-PET after application of a kinematic fit. Prospects of employment of this method for studying angular correlations of photons in decays of polarized ortho-positronia for the needs of tests of CP and CPT discrete symmetries are also discussed. The new reconstruction method allows for discrimination of background from random three-photon coincidences as well as for application of a novel method for determination of the linear polarization of ortho-positronium atoms, which is also introduced in this work.
△ Less
Submitted 24 February, 2016;
originally announced February 2016.
-
Determination of the 3γfraction from positron annihilation in mesoporous materials for symmetry violation experiment with J-PET scanner
Authors:
B. Jasińska,
M. Gorgol,
M. Wiertel,
R. Zaleski,
D. Alfs,
T. Bednarski,
P. Białas,
E. Czerwiński,
K. Dulski,
A. Gajos,
B. Głowacz,
D. Kamińska,
Ł. Kapłon,
G. Korcyl,
P. Kowalski,
T. Kozik,
W. Krzemień,
E. Kubicz,
M. Mohammed,
Sz. Niedźwiecki,
M. Pałka,
L. Raczyński,
Z. Rudy,
O. Rundel,
N. G. Sharma
, et al. (8 additional authors not shown)
Abstract:
Various mesoporous materials were investigated to choose the best material for experiments requiring high yield of long-lived positronium. We found that the fraction of 3γannihilation determined using γ-ray energy spectra and positron annihilation lifetime spectra (PAL) changed from 20% to 25%. The 3gamma fraction and o-Ps formation probability in the polymer XAD-4 is found to be the largest. Elem…
▽ More
Various mesoporous materials were investigated to choose the best material for experiments requiring high yield of long-lived positronium. We found that the fraction of 3γannihilation determined using γ-ray energy spectra and positron annihilation lifetime spectra (PAL) changed from 20% to 25%. The 3gamma fraction and o-Ps formation probability in the polymer XAD-4 is found to be the largest. Elemental analysis performed using scanning electron microscop (SEM) equipped with energy-dispersive X-ray spectroscop EDS show high purity of the investigated materials.
△ Less
Submitted 17 February, 2016;
originally announced February 2016.
-
Potential of the J-PET detector for studies of discrete symmetries in decays of positronium atom - a purely leptonic system
Authors:
P. Moskal,
D. Alfs,
T. Bednarski,
P. Białas,
E. Czerwiński,
C. Curceanu,
A. Gajos,
B. Głowacz,
M. Gorgol,
B. C. Hiesmayr,
B. Jasińska,
D. Kamińska,
G. Korcyl,
P. Kowalski,
T. Kozik,
W. Krzemień,
N. Krawczyk,
E. Kubicz,
M. Mohammed,
Sz. Niedźwieckia,
M. Pawlik-Niedźwiecka,
L. Raczyński,
Z. Rudy,
M. Silarski,
A. Wieczorek
, et al. (2 additional authors not shown)
Abstract:
The Jagiellonian Positron Emission Tomograph (J-PET) was constructed as a prototype of the cost-effective scanner for the simultaneous metabolic imaging of the whole human body. Being optimized for the detection of photons from the electron-positron annihilation with high time- and high angular-resolution, it constitutes a multi-purpose detector providing new opportunities for studying the decays…
▽ More
The Jagiellonian Positron Emission Tomograph (J-PET) was constructed as a prototype of the cost-effective scanner for the simultaneous metabolic imaging of the whole human body. Being optimized for the detection of photons from the electron-positron annihilation with high time- and high angular-resolution, it constitutes a multi-purpose detector providing new opportunities for studying the decays of positronium atoms. Positronium is the lightest purely leptonic object decaying into photons. As an atom bound by a central potential it is a parity eigenstate, and as an atom built out of an electron and an anti-electron it is an eigenstate of the charge conjugation operator. Therefore, the positronium is a unique laboratory to study discrete symmetries whose precision is limited in principle by the effects due to the weak interactions expected at the level of (~10$^{-14}$) and photon-photon interactions expected at the level of (~10$^{-9}$). The J-PET detector enables to perform tests of discrete symmetries in the leptonic sector via the determination of the expectation values of the discrete-symmetries-odd operators, which may be constructed from the spin of ortho-positronium atom and the momenta and polarization vectors of photons originating from its annihilation. In this article we present the potential of the J-PET detector to test the C, CP, T and CPT symmetries in the decays of positronium atoms.
△ Less
Submitted 16 February, 2016;
originally announced February 2016.
-
Search for polarization effects in the antiproton production process
Authors:
D. Grzonka,
K. Kilian,
J. Ritman,
T. Sefzick,
W. Oelert,
M. Diermaier,
E. Widmann,
J. Zmeskal,
B. Glowacz,
P. Moskal,
M. Zielinski,
M. Wolke,
P. Nadel-Turonski,
M. Carmignotto,
T. Horn,
H. Mkrtchyan,
A. Asaturyan,
A. Mkrtchyan,
V. Tadevosyan,
S. Zhamkochyan,
S. Malbrunot-Ettenauer,
W. Eyrich,
F. Hauenstein,
A. Zink
Abstract:
For the production of a polarized antiproton beam various methods have been suggested including the possibility that antiprotons may be produced polarized which will be checked experimentally. The polarization of antiprotons produced under typical conditions for antiproton beam preparation will be measured at the CERN/PS. If the production process creates some polarization a polarized antiproton b…
▽ More
For the production of a polarized antiproton beam various methods have been suggested including the possibility that antiprotons may be produced polarized which will be checked experimentally. The polarization of antiprotons produced under typical conditions for antiproton beam preparation will be measured at the CERN/PS. If the production process creates some polarization a polarized antiproton beam could be prepared by a rather simple modification of the antiproton beam facility. The detection setup and the expected experimental conditions are described.
△ Less
Submitted 23 January, 2015;
originally announced January 2015.