Response of a Li-glass/multi-anode photomultiplier detector to collimated thermal-neutron beams
Authors:
E. Rofors,
N. Mauritzson,
H. Perrey,
R. Al Jebali,
J. R. M. Annand,
L. Boyd,
M. J. Christensen,
U. Clemens,
S. Desert,
R. Engels,
K. G. Fissum,
H. Frielinghaus,
C. Gheorghe,
R. Hall-Wilton,
S. Jaksch,
K. Kanaki,
S. Kazi,
G. Kemmerling,
I. Llamas Jansa,
V. Maulerova,
R. Montgomery,
T. Richter,
J. Scherzinger,
B. Seitz,
M. Shetty
Abstract:
The response of a position-sensitive Li-glass scintillator detector being developed for thermal-neutron detection with 6 mm position resolution has been investigated using collimated beams of thermal neutrons. The detector was moved perpendicularly through the neutron beams in 0.5 to 1.0 mm horizontal and vertical steps. Scintillation was detected in an 8 X 8 pixel multi-anode photomultiplier tube…
▽ More
The response of a position-sensitive Li-glass scintillator detector being developed for thermal-neutron detection with 6 mm position resolution has been investigated using collimated beams of thermal neutrons. The detector was moved perpendicularly through the neutron beams in 0.5 to 1.0 mm horizontal and vertical steps. Scintillation was detected in an 8 X 8 pixel multi-anode photomultiplier tube on an event-by-event basis. In general, several pixels registered large signals at each neutron-beam location. The number of pixels registering signal above a set threshold was investigated, with the maximization of the single-hit efficiency over the largest possible area of the detector as the primary goal. At a threshold of ~50% of the mean of the full-deposition peak, ~80% of the events were registered in a single pixel, resulting in an effective position resolution of ~5 mm in X and Y. Lower thresholds generally resulted in events demonstrating higher pixel multiplicities, but these events could also be localized with ~5 mm position resolution.
△ Less
Submitted 9 December, 2020; v1 submitted 13 October, 2020;
originally announced October 2020.
-
Response of a Li-glass/multi-anode photomultiplier detector to focused proton and deuteron beams
Authors:
E. Rofors,
J. Pallon,
R. Al Jebali,
J. R. M. Annand,
L. Boyd,
M. J. Christensen,
U. Clemens,
S. Desert,
M. Elfman,
R. Engels,
K. G. Fissum,
H. Frielinghaus,
R. Frost,
S. Gardner,
C. Gheorghe,
R. Hall-Wilton,
S. Jaksch,
K. Kanaki,
G. Kemmerling,
P. Kristiansson,
K. Livingston,
V. Maulerova,
N. Mauritzson,
R. Montgomery,
H. Perrey
, et al. (4 additional authors not shown)
Abstract:
The response of a position-sensitive Li-glass based scintillation detector to focused beams of 2.5 MeV protons and deuterons has been investigated. The beams were scanned across the detector in 0.5 mm horizontal and vertical steps perpendicular to the beams. Scintillation light was registered using an 8 by 8 pixel multi-anode photomultiplier tube. The signal amplitudes were recorded for each pixel…
▽ More
The response of a position-sensitive Li-glass based scintillation detector to focused beams of 2.5 MeV protons and deuterons has been investigated. The beams were scanned across the detector in 0.5 mm horizontal and vertical steps perpendicular to the beams. Scintillation light was registered using an 8 by 8 pixel multi-anode photomultiplier tube. The signal amplitudes were recorded for each pixel on an event-by-event basis. Several pixels generally registered considerable signals at each beam location. The number of pixels above set thresholds were investigated, with the optimization of the single-hit efficiency over the largest possible area as the goal. For both beams, at a threshold of ~50% of the mean of the full-deposition peak, ~80% of the events were registered in a single pixel, resulting in an effective position resolution of ~5 mm in X and Y.
△ Less
Submitted 18 May, 2020;
originally announced May 2020.
-
Response of a Li-glass/multi-anode photomultiplier detector to $α$-particles from $^{241}$Am
Authors:
E. Rofors,
H. Perrey,
R. Al Jebali,
J. R. M. Annand,
L. Boyd,
U. Clemens,
S. Desert,
R. Engels,
K. G. Fissum,
H. Frielinghaus,
C. Gheorghe,
R. Hall-Wilton,
S. Jaksch,
A. Jalgén,
K. Kanaki,
G. Kemmerling,
V. Maulerova,
N. Mauritzson,
R. Montgomery,
J. Scherzinger,
B. Seitz
Abstract:
The response of a position-sensitive Li-glass scintillator detector to $α$-particles from a collimated $^{241}$Am source scanned across the face of the detector has been measured. Scintillation light was read out by an 8 X 8 pixel multi-anode photomultiplier and the signal amplitude for each pixel has been recorded for every position on a scan. The pixel signal is strongly dependent on position an…
▽ More
The response of a position-sensitive Li-glass scintillator detector to $α$-particles from a collimated $^{241}$Am source scanned across the face of the detector has been measured. Scintillation light was read out by an 8 X 8 pixel multi-anode photomultiplier and the signal amplitude for each pixel has been recorded for every position on a scan. The pixel signal is strongly dependent on position and in general several pixels will register a signal (a hit) above a given threshold. The effect of this threshold on hit multiplicity is studied, with a view to optimize the single-hit efficiency of the detector.
△ Less
Submitted 19 December, 2018;
originally announced December 2018.
-
Concept for a Time-of-Flight Small Angle Neutron Scattering Instrument at the European Spallation Source
Authors:
S. Jaksch,
D. Martin-Rodriguez,
A. Ostermann,
J. Jestin,
S. Duarte Pinto,
W. G. Bouwman,
J. Uher,
R. Engels,
G. Kemmerling,
R. Hanslik,
H. Frielinghaus
Abstract:
A new Small Angle Neutron Scattering instrument is proposed for the European Spallation Source. The pulsed source requires a time-of-flight analysis of the gathered neutrons at the detector. The optimal instrument length is found to be rather large, which allows for a polarizer and a versatile collimation. The polarizer allows for studying magnetic samples and incoherent background subtraction. Th…
▽ More
A new Small Angle Neutron Scattering instrument is proposed for the European Spallation Source. The pulsed source requires a time-of-flight analysis of the gathered neutrons at the detector. The optimal instrument length is found to be rather large, which allows for a polarizer and a versatile collimation. The polarizer allows for studying magnetic samples and incoherent background subtraction. The wide collimation will host VSANS and SESANS options that increase the resolution of the instrument towards um and tens of um, respectively. Two 1m2 area detectors will cover a large solid angle simultaneously. The expected gains for this new instrument will lie in the range between 20 and 36, depending on the assessment criteria, when compared to up-to-date reactor based instruments. This will open new perspectives for fast kinetics, weakly scattering samples, and multi-dimensional contrast variation studies.
△ Less
Submitted 11 March, 2014;
originally announced March 2014.