Achieving ultra-low and -uniform residual magnetic fields in a very large magnetically shielded room for fundamental physics experiments
Authors:
N. J. Ayres,
G. Ban,
G. Bison,
K. Bodek,
V. Bondar,
T. Bouillaud,
D. Bowles,
E. Chanel,
W. Chen,
P. -J. Chiu,
C. B. Crawford,
O. Naviliat-Cuncic,
C. B. Doorenbos,
S. Emmenegger,
M. Fertl,
A. Fratangelo,
W. C. Griffith,
Z. D. Grujic,
P. G. Harris,
K. Kirch,
V. Kletzl,
J. Krempel,
B. Lauss,
T. Lefort,
A. Lejuez
, et al. (25 additional authors not shown)
Abstract:
High-precision searches for an electric dipole moment of the neutron (nEDM) require stable and uniform magnetic field environments. We present the recent achievements of degaussing and equilibrating the magnetically shielded room (MSR) for the n2EDM experiment at the Paul Scherrer Institute. We present the final degaussing configuration that will be used for n2EDM after numerous studies. The optim…
▽ More
High-precision searches for an electric dipole moment of the neutron (nEDM) require stable and uniform magnetic field environments. We present the recent achievements of degaussing and equilibrating the magnetically shielded room (MSR) for the n2EDM experiment at the Paul Scherrer Institute. We present the final degaussing configuration that will be used for n2EDM after numerous studies. The optimized procedure results in a residual magnetic field that has been reduced by a factor of two. The ultra-low field is achieved with the full magnetic-field-coil system, and a large vacuum vessel installed, both in the MSR. In the inner volume of ~1.4 m^3, the field is now more uniform and below 300 pT. In addition, the procedure is faster and dissipates less heat into the magnetic environment, which in turn, reduces its thermal relaxation time from 12 h down to ~1.5 h.
△ Less
Submitted 28 September, 2023;
originally announced September 2023.
-
A large 'Active Magnetic Shield' for a high-precision experiment
Authors:
C. Abel,
N. J. Ayres,
G. Ban,
G. Bison,
K. Bodek,
V. Bondar,
T. Bouillaud,
E. Chanel,
J. Chen,
W. Chen,
P. -J. Chiu,
C. B. Crawford,
M. Daum,
C. B. Doorenbos,
S. Emmenegger,
L. Ferraris-Bouchez,
M. Fertl,
A. Fratangelo,
W. C. Griffith,
Z. D. Grujic,
P. Harris,
K. Kirch,
V. Kletzl,
P. A. Koss,
J. Krempel
, et al. (26 additional authors not shown)
Abstract:
We present a novel Active Magnetic Shield (AMS), designed and implemented for the n2EDM experiment at the Paul Scherrer Institute. The experiment will perform a high-sensitivity search for the electric dipole moment of the neutron. Magnetic-field stability and control is of key importance for n2EDM. A large, cubic, 5m side length, magnetically shielded room (MSR) provides a passive, quasi-static s…
▽ More
We present a novel Active Magnetic Shield (AMS), designed and implemented for the n2EDM experiment at the Paul Scherrer Institute. The experiment will perform a high-sensitivity search for the electric dipole moment of the neutron. Magnetic-field stability and control is of key importance for n2EDM. A large, cubic, 5m side length, magnetically shielded room (MSR) provides a passive, quasi-static shielding-factor of about 10^5 for its inner sensitive volume. The AMS consists of a system of eight complex, feedback-controlled compensation coils constructed on an irregular grid spanned on a volume of less than 1000m^3 around the MSR. The AMS is designed to provide a stable and uniform magnetic-field environment around the MSR, while being reasonably compact. The system can compensate static and variable magnetic fields up to +-50muT (homogeneous components) and +-5muT (first-order gradients), suppressing them to a few muT in the sub-Hertz frequency range. The presented design concept and implementation of the AMS fulfills the requirements of the n2EDM experiment and can be useful for other applications, where magnetically silent environments are important and spatial constraints inhibit simpler geometrical solutions.
△ Less
Submitted 14 July, 2023;
originally announced July 2023.
-
Search for an interaction mediated by axion-like particles with ultracold neutrons at the PSI
Authors:
N. J. Ayres,
G. Bison,
K. Bodek,
V. Bondar,
T. Bouillaud,
E. Chanel,
P. -J. Chiu,
B. Clement,
C. B. Crawford,
M. Daum,
C. B. Doorenbos,
S. Emmenegger,
M. Fertl,
P. Flaux,
W. C. Griffith,
P. G. Harris,
N. Hild,
M. Kasprzak,
K. Kirch,
V. Kletzl,
P. A. Koss,
J. Krempel,
B. Lauss,
T. Lefort,
P. Mohanmurthy
, et al. (22 additional authors not shown)
Abstract:
We report on a search for a new, short-range, spin-dependent interaction using a modified version of the experimental apparatus used to measure the permanent neutron electric dipole moment at the Paul Scherrer Institute. This interaction, which could be mediated by axion-like particles, concerned the unpolarized nucleons (protons and neutrons) near the material surfaces of the apparatus and polari…
▽ More
We report on a search for a new, short-range, spin-dependent interaction using a modified version of the experimental apparatus used to measure the permanent neutron electric dipole moment at the Paul Scherrer Institute. This interaction, which could be mediated by axion-like particles, concerned the unpolarized nucleons (protons and neutrons) near the material surfaces of the apparatus and polarized ultracold neutrons stored in vacuum. The dominant systematic uncertainty resulting from magnetic-field gradients was controlled to an unprecedented level of approximately 4 pT/cm using an array of optically-pumped cesium vapor magnetometers and magnetic-field maps independently recorded using a dedicated measurement device. No signature of a theoretically predicted new interaction was found, and we set a new limit on the product of the scalar and the pseudoscalar couplings $g_sg_pλ^2 < 8.3 \times 10^{-28}\,\text{m}^2$ (95% C.L.) in a range of $5\,μ\text{m} < λ< 25\,\text{mm}$ for the monopole-dipole interaction. This new result confirms and improves our previous limit by a factor of 2.7 and provides the current tightest limit obtained with free neutrons.
△ Less
Submitted 31 March, 2023;
originally announced March 2023.
-
The `n2EDM MSR' -- a very large magnetically shielded room with an exceptional performance for fundamental physics measurements
Authors:
N. J. Ayres,
G. Ban,
G. Bison,
K. Bodek,
V. Bondar,
T. Bouillaud,
B. Clement,
E. Chanel,
P. -J. Chiu,
C. B. Crawford,
M. Daum,
C. B. Doorenbos,
S. Emmenegger,
A. Fratangelo,
M. Fertl,
W. C. Griffith,
Z. D. Grujic,
P. G. Harris,
K. Kirch,
J. Krempel,
B. Lauss,
T. Lefort,
O. Naviliat-Cuncic,
D. Pais,
F. M. Piegsa
, et al. (19 additional authors not shown)
Abstract:
We present the magnetically shielded room (MSR) for the n2EDM experiment at the Paul Scherrer Institute which features an interior cubic volume with each side of length 2.92m, thus providing an accessible space of 25m3. The MSR has 87 openings up to 220mm diameter to operate the experimental apparatus inside, and an intermediate space between the layers for sensitive signal processing electronics.…
▽ More
We present the magnetically shielded room (MSR) for the n2EDM experiment at the Paul Scherrer Institute which features an interior cubic volume with each side of length 2.92m, thus providing an accessible space of 25m3. The MSR has 87 openings up to 220mm diameter to operate the experimental apparatus inside, and an intermediate space between the layers for sensitive signal processing electronics. The characterization measurements show a remanent magnetic field in the central 1m3 below 100pT, and a field below 600pT in the entire inner volume, up to 4\,cm to the walls. The quasi-static shielding factor at 0.01\,Hz measured with a sinusoidal 2muT peak-to-peak signal is about 100,000 in all three spatial directions and rises fast with frequency to reach 10^8 above 1Hz.
△ Less
Submitted 21 June, 2022;
originally announced June 2022.
-
Improved search for neutron to mirror-neutron oscillations in the presence of mirror magnetic fields with a dedicated apparatus at the PSI UCN source
Authors:
N. J. Ayres,
Z. Berezhiani,
R. Biondi,
G. Bison,
K. Bodek,
V. Bondar,
P. -J. Chiu,
M. Daum,
R. T. Dinani,
C. B. Doorenbos,
S. Emmenegger,
K. Kirch,
V. Kletzl,
J. Krempel,
B. Lauss,
D. Pais,
I. Rienaecker,
D. Ries,
N. Rossi,
D. Rozpedzik,
P. Schmidt-Wellenburg,
K. S. Tanaka,
J. Zejma,
N. Ziehl,
G. Zsigmond
Abstract:
While the international nEDM collaboration at the Paul Scherrer Institut (PSI) took data in 2017 that covered a considerable fraction of the parameter space of claimed potential signals of hypothetical neutron ($n$) to mirror-neutron ($n'$) transitions, it could not test all claimed signal regions at various mirror magnetic fields. Therefore, a new study of $n-n'$ oscillations using stored ultraco…
▽ More
While the international nEDM collaboration at the Paul Scherrer Institut (PSI) took data in 2017 that covered a considerable fraction of the parameter space of claimed potential signals of hypothetical neutron ($n$) to mirror-neutron ($n'$) transitions, it could not test all claimed signal regions at various mirror magnetic fields. Therefore, a new study of $n-n'$ oscillations using stored ultracold neutrons (UCNs)is underway at PSI, considerably expanding the reach in parameter space of mirror magnetic fields ($B'$) and oscillation time constants ($τ_{nn'}$). The new apparatus is designed to test for the anomalous loss of stored ultracold neutrons as a function of an applied magnetic field. The experiment is distinguished from its predecessors by its very large storage vessel (1.47\,m$^3$), enhancing its statistical sensitivity. In a test experiment in 2020 we have demonstrated the capabilities of our apparatus. However, the full analysis of our recent data is still pending. Based on already demonstrated performance, we will reach a sensitivity to oscillation times $τ_{nn'}/\sqrt{\cos(β)}$ well above hundred seconds, with $β$ being the angle between $B'$ and the applied magnetic field $B$. The scan of $B$ will allow the finding or the comprehensive exclusion of potential signals reported in the analysis of previous experiments and suggested to be consistent with neutron to mirror-neutron oscillations.
△ Less
Submitted 31 October, 2021;
originally announced November 2021.
-
Johnson-Nyquist Noise Effects in Neutron Electric-Dipole-Moment Experiments
Authors:
N. J. Ayres,
G. Ban,
G. Bison,
K. Bodek,
V. Bondar,
P. -J. Chiu,
B. Clement,
C. B. Crawford,
M. Daum,
S. Emmenegger,
M. Fertl,
A. Fratangelo,
W. C. Griffith,
Z. D. Grujić,
P. G. Harris,
K. Kirch,
P. A. Koss,
B. Lauss,
T. Lefort,
P. Mohanmurthy,
O. Naviliat-Cuncic,
D. Pais,
F. M. Piegsa,
G. Pignol,
D. Rebreyend
, et al. (15 additional authors not shown)
Abstract:
Magnetic Johnson-Nyquist noise (JNN) originating from metal electrodes, used to create a static electric field in neutron electric-dipole-moment (nEDM) experiments, may limit the sensitivity of measurements. We present here the first dedicated study on JNN applied to a large-scale long-measurement-time experiment with the implementation of a co-magnetometry. In this study, we derive surface- and v…
▽ More
Magnetic Johnson-Nyquist noise (JNN) originating from metal electrodes, used to create a static electric field in neutron electric-dipole-moment (nEDM) experiments, may limit the sensitivity of measurements. We present here the first dedicated study on JNN applied to a large-scale long-measurement-time experiment with the implementation of a co-magnetometry. In this study, we derive surface- and volume-averaged root-mean-square normal noise amplitudes at a certain frequency bandwidth for a cylindrical geometry. In addition, we model the source of noise as a finite number of current dipoles and demonstrate a method to simulate temporal and three-dimensional spatial dependencies of JNN. The calculations are applied to estimate the impact of JNN on measurements with the new apparatus, n2EDM, at the Paul Scherrer Institute. We demonstrate that the performances of the optically pumped $^{133}$Cs magnetometers and $^{199}$Hg co-magnetometers, which will be used in the apparatus, are not limited by JNN. Further, we find that in measurements deploying a co-magnetometer system, the impact of JNN is negligible for nEDM searches down to a sensitivity of $4\,\times\,10^{-28}\,e\cdot{\rm cm}$ in a single measurement; therefore, the use of economically and mechanically favored solid aluminum electrodes is possible.
△ Less
Submitted 9 July, 2021; v1 submitted 2 February, 2021;
originally announced February 2021.
-
The design of the n2EDM experiment
Authors:
N. J. Ayres,
G. Ban,
L. Bienstman,
G. Bison,
K. Bodek,
V. Bondar,
T. Bouillaud,
E. Chanel,
J. Chen,
P. -J. Chiu,
B. Clément,
C. Crawford,
M. Daum,
B. Dechenaux,
C. B. Doorenbos,
S. Emmenegger,
L. Ferraris-Bouchez,
M. Fertl,
A. Fratangelo,
P. Flaux,
D. Goupillière,
W. C. Griffith,
Z. D. Grujic,
P. G. Harris,
K. Kirch
, et al. (36 additional authors not shown)
Abstract:
We present the design of a next-generation experiment, n2EDM, currently under construction at the ultracold neutron source at the Paul Scherrer Institute (PSI) with the aim of carrying out a high-precision search for an electric dipole moment of the neutron. The project builds on experience gained with the previous apparatus operated at PSI until 2017, and is expected to deliver an order of magnit…
▽ More
We present the design of a next-generation experiment, n2EDM, currently under construction at the ultracold neutron source at the Paul Scherrer Institute (PSI) with the aim of carrying out a high-precision search for an electric dipole moment of the neutron. The project builds on experience gained with the previous apparatus operated at PSI until 2017, and is expected to deliver an order of magnitude better sensitivity with provision for further substantial improvements. An overview is given of the experimental method and setup, the sensitivity requirements for the apparatus are derived, and its technical design is described.
△ Less
Submitted 22 January, 2021; v1 submitted 21 January, 2021;
originally announced January 2021.
-
A search for neutron to mirror-neutron oscillations
Authors:
C. Abel,
N. J. Ayres,
G. Ban,
G. Bison,
K. Bodek,
V. Bondar,
E. Chanel,
P. -J. Chiu,
C. Crawford,
M. Daum,
R. T. Dinani,
S. Emmenegger,
P. Flaux,
L. Ferraris-Bouchez,
W. C. Griffith,
Z. D. Grujic,
N. Hild,
K. Kirch,
H. -C. Koch,
P. A. Koss,
A. Kozela,
J. Krempel,
B. Lauss,
T. Lefort,
A. Leredde
, et al. (18 additional authors not shown)
Abstract:
It has been proposed that there could be a mirror copy of the standard model particles, restoring the parity symmetry in the weak interaction on the global level. Oscillations between a neutral standard model particle, such as the neutron, and its mirror counterpart could potentially answer various standing issues in physics today. Astrophysical studies and terrestrial experiments led by ultracold…
▽ More
It has been proposed that there could be a mirror copy of the standard model particles, restoring the parity symmetry in the weak interaction on the global level. Oscillations between a neutral standard model particle, such as the neutron, and its mirror counterpart could potentially answer various standing issues in physics today. Astrophysical studies and terrestrial experiments led by ultracold neutron storage measurements have investigated neutron to mirror-neutron oscillations and imposed constraints on the theoretical parameters. Recently, further analysis of these ultracold neutron storage experiments has yielded statistically significant anomalous signals that may be interpreted as neutron to mirror-neutron oscillations, assuming nonzero mirror magnetic fields. The neutron electric dipole moment collaboration performed a dedicated search at the Paul Scherrer Institute and found no evidence of neutron to mirror-neutron oscillations. Thereby, the following new lower limits on the oscillation time were obtained: $τ_{nn'} > 352~$s at $B'=0$ (95% C.L.), $τ_{nn'} > 6~\text{s}$ for all $0.4~μ\text{T}<B'<25.7~μ\text{T}$ (95% C.L.), and $τ_{nn'}/\sqrt{\cosβ}>9~\text{s}$ for all $5.0~μ\text{T}<B'<25.4~μ\text{T}$ (95% C.L.), where $β$ is the fixed angle between the applied magnetic field and the local mirror magnetic field which is assumed to be bound to the Earth. These new constraints are the best measured so far around $B'\sim10~μ$T, and $B'\sim20~μ$T.
△ Less
Submitted 19 November, 2020; v1 submitted 23 September, 2020;
originally announced September 2020.
-
Measurement of the permanent electric dipole moment of the neutron
Authors:
C. Abel,
S. Afach,
N. J. Ayres,
C. A. Baker,
G. Ban,
G. Bison,
K. Bodek,
V. Bondar,
M. Burghoff,
E. Chanel,
Z. Chowdhuri,
P. -J. Chiu,
B. Clement,
C. B. Crawford,
M. Daum,
S. Emmenegger,
L. Ferraris-Bouchez,
M. Fertl,
P. Flaux,
B. Franke,
A. Fratangelo,
P. Geltenbort,
K. Green,
W. C. Griffith,
M. van der Grinten
, et al. (59 additional authors not shown)
Abstract:
We present the result of an experiment to measure the electric dipole moment (EDM) of the neutron at the Paul Scherrer Institute using Ramsey's method of separated oscillating magnetic fields with ultracold neutrons (UCN). Our measurement stands in the long history of EDM experiments probing physics violating time reversal invariance. The salient features of this experiment were the use of a Hg-19…
▽ More
We present the result of an experiment to measure the electric dipole moment (EDM) of the neutron at the Paul Scherrer Institute using Ramsey's method of separated oscillating magnetic fields with ultracold neutrons (UCN). Our measurement stands in the long history of EDM experiments probing physics violating time reversal invariance. The salient features of this experiment were the use of a Hg-199 co-magnetometer and an array of optically pumped cesium vapor magnetometers to cancel and correct for magnetic field changes. The statistical analysis was performed on blinded datasets by two separate groups while the estimation of systematic effects profited from an unprecedented knowledge of the magnetic field. The measured value of the neutron EDM is $d_{\rm n} = (0.0\pm1.1_{\rm stat}\pm0.2_{\rm sys})\times10^{-26}e\,{\rm cm}$.
△ Less
Submitted 31 January, 2020;
originally announced January 2020.
-
Optically Pumped Cs Magnetometers Enabling a High-Sensitivity Search for the Neutron Electric Dipole Moment
Authors:
C. Abel,
S. Afach,
N. J. Ayres,
G. Ban,
G. Bison,
K. Bodek,
V. Bondar,
E. Chanel,
P. -J. Chiu,
C. B. Crawford,
Z. Chowdhuri,
M. Daum,
S. Emmenegger,
L. Ferraris-Bouchez,
M. Fertl,
B. Franke,
W. C. Griffith,
Z. D. Grujić,
L. Hayen,
V. Hélaine,
N. Hild,
M. Kasprzak,
Y. Kermaidic,
K. Kirch,
P. Knowles
, et al. (35 additional authors not shown)
Abstract:
An array of sixteen laser-pumped scalar Cs magnetometers was part of the neutron electric dipole moment (nEDM) experiment taking data at the Paul Scherrer Institute in 2015 and 2016. It was deployed to measure the gradients of the experiment's magnetic field and to monitor their temporal evolution. The originality of the array lies in its compact design, in which a single near-infrared diode laser…
▽ More
An array of sixteen laser-pumped scalar Cs magnetometers was part of the neutron electric dipole moment (nEDM) experiment taking data at the Paul Scherrer Institute in 2015 and 2016. It was deployed to measure the gradients of the experiment's magnetic field and to monitor their temporal evolution. The originality of the array lies in its compact design, in which a single near-infrared diode laser drives all magnetometers that are located in a high-vacuum chamber, with a selection of the sensors mounted on a high-voltage electrode. We describe details of the Cs sensors' construction and modes of operation, emphasizing the accuracy and sensitivity of the magnetic field readout. We present two applications of the magnetometer array directly beneficial to the nEDM experiment: (i) the implementation of a strategy to correct for the drift of the vertical magnetic field gradient and (ii) a procedure to homogenize the magnetic field. The first reduces the uncertainty of the new nEDM result. The second enables transverse neutron spin relaxation times exceeding 1500 s, improving the statistical sensitivity of the nEDM experiment by about 35% and effectively increasing the rate of nEDM data taking by a factor of 1.8.
△ Less
Submitted 28 April, 2020; v1 submitted 10 December, 2019;
originally announced December 2019.
-
A New Cryogenic Apparatus to Search for the Neutron Electric Dipole Moment
Authors:
M. W. Ahmed,
R. Alarcon,
A. Aleksandrova,
S. Baessler,
L. Barron-Palos,
L. M. Bartoszek,
D. H. Beck,
M. Behzadipour,
I. Berkutov,
J. Bessuille,
M. Blatnik,
M. Broering,
L. J. Broussard,
M. Busch,
R. Carr,
V. Cianciolo,
S. M. Clayton,
M. D. Cooper,
C. Crawford,
S. A. Currie,
C. Daurer,
R. Dipert,
K. Dow,
D. Dutta,
Y. Efremenko
, et al. (69 additional authors not shown)
Abstract:
A cryogenic apparatus is described that enables a new experiment, nEDM@SNS, with a major improvement in sensitivity compared to the existing limit in the search for a neutron Electric Dipole Moment (EDM). It uses superfluid $^4$He to produce a high density of Ultra-Cold Neutrons (UCN) which are contained in a suitably coated pair of measurement cells. The experiment, to be operated at the Spallati…
▽ More
A cryogenic apparatus is described that enables a new experiment, nEDM@SNS, with a major improvement in sensitivity compared to the existing limit in the search for a neutron Electric Dipole Moment (EDM). It uses superfluid $^4$He to produce a high density of Ultra-Cold Neutrons (UCN) which are contained in a suitably coated pair of measurement cells. The experiment, to be operated at the Spallation Neutron Source at Oak Ridge National Laboratory, uses polarized $^3$He from an Atomic Beam Source injected into the superfluid $^4$He and transported to the measurement cells as a co-magnetometer. The superfluid $^4$He is also used as an insulating medium allowing significantly higher electric fields, compared to previous experiments, to be maintained across the measurement cells. These features provide an ultimate statistical uncertainty for the EDM of $2-3\times 10^{-28}$ e-cm, with anticipated systematic uncertainties below this level.
△ Less
Submitted 20 November, 2019; v1 submitted 26 August, 2019;
originally announced August 2019.
-
The neutron electric dipole moment experiment at the Spallation Neutron Source
Authors:
K. K. H. Leung,
M. Ahmed,
R. Alarcon,
A. Aleksandrova,
S. Baeßler,
L. Barrón-Palos,
L. Bartoszek,
D. H. Beck,
M. Behzadipour,
J. Bessuille,
M. A. Blatnik,
M. Broering,
L. J. Broussard,
M. Busch,
R. Carr,
P. -H. Chu,
V. Cianciolo,
S. M. Clayton,
M. D. Cooper,
C. Crawford,
S. A. Currie,
C. Daurer,
R. Dipert,
K. Dow,
D. Dutta
, et al. (68 additional authors not shown)
Abstract:
Novel experimental techniques are required to make the next big leap in neutron electric dipole moment experimental sensitivity, both in terms of statistics and systematic error control. The nEDM experiment at the Spallation Neutron Source (nEDM@SNS) will implement the scheme of Golub & Lamoreaux [Phys. Rep., 237, 1 (1994)]. The unique properties of combining polarized ultracold neutrons, polarize…
▽ More
Novel experimental techniques are required to make the next big leap in neutron electric dipole moment experimental sensitivity, both in terms of statistics and systematic error control. The nEDM experiment at the Spallation Neutron Source (nEDM@SNS) will implement the scheme of Golub & Lamoreaux [Phys. Rep., 237, 1 (1994)]. The unique properties of combining polarized ultracold neutrons, polarized $^3$He, and superfluid $^4$He will be exploited to provide a sensitivity to $\sim 10^{-28}\,e{\rm \,\cdot\, cm}$. Our cryogenic apparatus will deploy two small ($3\,{\rm L}$) measurement cells with a high density of ultracold neutrons produced and spin analyzed in situ. The electric field strength, precession time, magnetic shielding, and detected UCN number will all be enhanced compared to previous room temperature Ramsey measurements. Our $^3$He co-magnetometer offers unique control of systematic effects, in particular the Bloch-Siegert induced false EDM. Furthermore, there will be two distinct measurement modes: free precession and dressed spin. This will provide an important self-check of our results. Following five years of "critical component demonstration," our collaboration transitioned to a "large scale integration" phase in 2018. An overview of our measurement techniques, experimental design, and brief updates are described in these proceedings.
△ Less
Submitted 4 October, 2019; v1 submitted 6 March, 2019;
originally announced March 2019.