Comparative studies of plastic scintillator strips with high technical attenuation length for the total-body J-PET scanner
Authors:
L. Kaplon,
J. Baran,
N. Chug,
A. Coussat,
C. Curceanu,
E. Czerwinski,
M. Dadgar,
K. Dulski,
J. Gajewski,
A. Gajos,
B. Hiesmayr,
E. Kavya Valsan,
K. Klimaszewski,
G. Korcyl,
T. Kozik,
W. Krzemien,
D. Kumar,
G. Moskal,
S. Niedzwiecki,
D. Panek,
S. Parzych,
E. Perez del Rio,
L. Raczynski,
A. Rucinski,
S. Sharma
, et al. (9 additional authors not shown)
Abstract:
Plastic scintillator strips are considered as one of the promising solutions for the cost-effective construction of total-body positron emission tomography, (PET) system. The purpose of the performed measurements is to compare the transparency of long plastic scintillators with dimensions 6 mm x 24 mm x 1000 mm and with all surfaces polished. Six different types of commercial, general purpose, blu…
▽ More
Plastic scintillator strips are considered as one of the promising solutions for the cost-effective construction of total-body positron emission tomography, (PET) system. The purpose of the performed measurements is to compare the transparency of long plastic scintillators with dimensions 6 mm x 24 mm x 1000 mm and with all surfaces polished. Six different types of commercial, general purpose, blue-emitting plastic scintillators with low attenuation of visible light were tested, namely: polyvinyl toluene-based BC-408, EJ-200, RP-408, and polystyrene-based Epic, SP32 and UPS-923A. For determination of the best type of plastic scintillator for totalbody Jagiellonian positron emission tomograph (TB-J-PET) construction, emission and transmission spectra, and technical attenuation length (TAL) of blue light-emitting by the scintillators were measured and compared. The TAL values were determined with the use of UV lamp as excitation source, and photodiode as light detector. Emission spectra of investigated scintillators have maxima in the range from 420 nm to 429 nm. The BC-408 and EJ-200 have the highest transmittance values of about 90% at the maximum emission wavelength measured through a 6 mm thick scintillator strip and the highest technical attenuation length reaching about 2000 mm, allowing assembly of long detection modules for time-of-flight (TOF) J-PET scanners. Influence of the 6 mm x 6 mm, 12 mm x 6 mm, 24 mm x 6 mm cross-sections of the 1000 mm long EJ-200 plastic scintillator on the TAL and signal intensity was measured. The highest TAL value was determined for samples with 24 mm x 6 mm cross-section.
△ Less
Submitted 3 August, 2024; v1 submitted 28 July, 2024;
originally announced July 2024.
Feasibility studies for imaging e$^{+}$e$^{-}$ annihilation with modular multi-strip detectors
Authors:
S. Sharma,
L. Povolo,
S. Mariazzi,
G. Korcyl,
K. Kacprzak,
D. Kumar,
S. Niedzwiecki,
J. Baran,
E. Beyene,
R. S. Brusa,
R. Caravita,
N. Chug,
A. Coussat,
C. Curceanu,
E. Czerwinski,
M. Dadgar,
M. Das,
K. Dulski,
K. Eliyan,
A. Gajos,
N. Gupta,
B. C. Hiesmayr,
L. Kaplon,
T. Kaplanoglu,
K. Klimaszewski
, et al. (19 additional authors not shown)
Abstract:
Studies based on imaging the annihilation of the electron (e$^{-}$) and its antiparticle positron (e$^{+}$) open up several interesting applications in nuclear medicine and fundamental research. The annihilation process involves both the direct conversion of e$^{+}$e$^{-}$ into photons and the formation of their atomically bound state, the positronium atom (Ps), which can be used as a probe for fu…
▽ More
Studies based on imaging the annihilation of the electron (e$^{-}$) and its antiparticle positron (e$^{+}$) open up several interesting applications in nuclear medicine and fundamental research. The annihilation process involves both the direct conversion of e$^{+}$e$^{-}$ into photons and the formation of their atomically bound state, the positronium atom (Ps), which can be used as a probe for fundamental studies. With the ability to produce large quantities of Ps, manipulate them in long-lived Ps states, and image their annihilations after a free fall or after passing through atomic interferometers, this purely leptonic antimatter system can be used to perform inertial sensing studies in view of a direct test of Einstein equivalence principle. It is envisioned that modular multistrip detectors can be exploited as potential detection units for this kind of studies. In this work, we report the results of the first feasibility study performed on a e$^{+}$ beamline using two detection modules to evaluate their reconstruction performance and spatial resolution for imaging e$^{+}$e$^{-}$ annihilations and thus their applicability for gravitational studies of Ps.
△ Less
Submitted 12 September, 2023;
originally announced September 2023.