Silicon $4π$ spectrometer for $β$-decay electrons with energies up to 3~MeV
Authors:
I. E. Alexeev,
S. V. Bakhlanov,
E. A. Chmel,
A. V. Derbin,
I. S. Drachnev,
I. M. Kotina,
M. S. Mikulich,
V. N. Muratova,
N. V. Nyazova,
D. A. Semenov,
M. V. Trushin,
E. V. Unzhakov
Abstract:
We present a description of the originally developed $β$-spectrometer consisting of two Si(Li)-detectors with sensitive area thickness above 8~mm and $4π$-geometry. The full absorption spectrometer allows for direct measurements of $β$-spectra, disregarding the corrections to the response function induced by the electron backscattering from the crystal surface. In case of $β$-spectra of transition…
▽ More
We present a description of the originally developed $β$-spectrometer consisting of two Si(Li)-detectors with sensitive area thickness above 8~mm and $4π$-geometry. The full absorption spectrometer allows for direct measurements of $β$-spectra, disregarding the corrections to the response function induced by the electron backscattering from the crystal surface. In case of $β$-spectra of transitions to the excited state of the daughter isotope additional 3" BGO-detector is used in order to detect the $γ$-quanta in coincidence with the pair of Si(Li)-spectrometers.
△ Less
Submitted 21 October, 2022;
originally announced October 2022.
-
Influence of α-particles irradiation on the performance and defect levels structure of Al/SiO2/p-type Si surface barrier detector
Authors:
S. V. Bakhlanov,
N. V. Bazlov,
D. V. Danilov,
A. V. Derbin,
I. S. Drachnev,
I. M. Kotina,
O. I. Konkov,
A. M. Kuzmichev,
M. S. Mikulich,
V. N. Muratova,
M. V. Trushin,
E. V. Unzhakov
Abstract:
Deterioration of the operation parameters of Al/SiO2/p-type Si surface barrier detector upon irradiation with alpha-particles at room temperature was investigated. As a result of 40-days irradiation with a total fluence of 8*10^9 α-particles, an increase of α-peak FWHM from 70 keV to 100 keV was observed and explained by increase of the detector reverse current due to formation of a high concentra…
▽ More
Deterioration of the operation parameters of Al/SiO2/p-type Si surface barrier detector upon irradiation with alpha-particles at room temperature was investigated. As a result of 40-days irradiation with a total fluence of 8*10^9 α-particles, an increase of α-peak FWHM from 70 keV to 100 keV was observed and explained by increase of the detector reverse current due to formation of a high concentration of near mid-gap defect levels. Performed CV measurements revealed the appearance of at least 6*10^12 cm-3 radiation-induced acceptors at the depths where according to the TRIM simulations the highest concentration of vacancy-interstitial pairs was created by the incoming α-particles. The studies carried out by current-DLTS technique allowed to associate the observed increase of the acceptor concentration with the near mid-gap acceptor level at EV+0.56 eV. This level can be apparently associated with V2O defects recognized previously to be responsible for the space charge sign inversion in the irradiated n-type Si detectors.
△ Less
Submitted 11 January, 2021;
originally announced January 2021.
-
Precision measurement of $^{\mathbf{210}}$Bi $β$-spectrum
Authors:
I. E. Alekseev,
S. V. Bakhlanov,
A. V. Derbin,
I. S. Drachnev,
I. M. Kotina,
I. S. Lomskaya,
V. N. Muratova,
N. V. Niyazova,
D. A. Semenov,
M. V. Trushin,
E. V. Unzhakov
Abstract:
The precision measurement of the $β-$spectrum shape for $^{210}$Bi (historically RaE) have been performed with a spectrometer based on semiconductor Si(Li) detector. This first forbidden non-unique transition has the transition form-factor strongly deviated from unity and knowledge of its spectrum would play an important role in low-background physics in presence of $^{210}$Pb background. The meas…
▽ More
The precision measurement of the $β-$spectrum shape for $^{210}$Bi (historically RaE) have been performed with a spectrometer based on semiconductor Si(Li) detector. This first forbidden non-unique transition has the transition form-factor strongly deviated from unity and knowledge of its spectrum would play an important role in low-background physics in presence of $^{210}$Pb background. The measured transition form-factor could be approximated as $S(W) = 1 + (-0.4363 \pm 0.0037) W + (0.0523 \pm 0.0010) W^2$, that is in good agreement with previous studies and has significantly increased parameter precision.
△ Less
Submitted 11 January, 2021; v1 submitted 18 May, 2020;
originally announced May 2020.
-
Measurement of $^{144}\rm{Pr}$ beta-spectrum with Si(Li) detectors for the purpose of determining the spectrum of electron antineutrinos
Authors:
A. V. Derbin,
S. V. Bakhlanov,
I. S. Drachnev,
I. P. Filipov,
L. A. Lukyanchenko,
I. N. Machulin,
V. N. Muratova,
N. V. Pilipenko,
D. A. Semenov,
E. V. Unzhakov
Abstract:
Here we present the specifications of the newly developed beta-spectrometer based on thick full absorption Si(Li) detector. The spectrometer can be used for precision measurements of various beta-spectra, namely for the beta-spectrum shape study of $^{144}$Pr, which is considered to be the most promising anti-neutrino source for sterile neutrino searches.
Here we present the specifications of the newly developed beta-spectrometer based on thick full absorption Si(Li) detector. The spectrometer can be used for precision measurements of various beta-spectra, namely for the beta-spectrum shape study of $^{144}$Pr, which is considered to be the most promising anti-neutrino source for sterile neutrino searches.
△ Less
Submitted 19 November, 2017;
originally announced November 2017.
-
A measurement method of a detector response function for monochromatic electrons based on the Compton scattering
Authors:
S. V. Bakhlanov,
N. V. Bazlov,
A. V. Derbin,
I. S. Drachnev,
A. S. Kayunov,
V. N. Muratova,
D. A. Semenov,
E. V. Unzhakov
Abstract:
In this paper we present a method of scintillation detector energy calibration using the gamma-rays. The technique is based on the Compton scattering of gamma-rays in a scintillation detector and subsequent photoelectric absorption of the scattered photon in the Ge-detector. The novelty of this method is that the source of gamma rays, the germanium and scintillation detectors are immediately arran…
▽ More
In this paper we present a method of scintillation detector energy calibration using the gamma-rays. The technique is based on the Compton scattering of gamma-rays in a scintillation detector and subsequent photoelectric absorption of the scattered photon in the Ge-detector. The novelty of this method is that the source of gamma rays, the germanium and scintillation detectors are immediately arranged adjacent to each other. The method presents an effective solution for the detectors consisting of a low atomic number materials, when the ratio between Compton effect and photoelectric absorption is large and the mean path of gamma-rays is comparable to the size of the detector. The technique can be used for the precision measurements of the scintillator light yield dependence on the electron energy.
△ Less
Submitted 13 April, 2016;
originally announced April 2016.