On low-energy tail distortions in the detector responsefunction of x-ray microcalorimeter spectrometers
Authors:
G. C. O'Neil,
P. Szypryt,
E. Takacs,
J. N. Tan,
S. W. Buechele,
A. S. Naing,
Y. I. Joe,
D. Swetz,
D. R. Schmidt,
W. B. Doriese,
J. D. Gard,
C. D. Reintsema,
J. N. Ullom,
J. S. Villarrubia,
Yu. Ralchenko
Abstract:
We use narrow spectral lines from the x-ray spectra of various highlycharged ions to measure low-energy tail-like deviations from a Gaussian responsefunction in a microcalorimter x-ray spectrometer with Au absorbers at energiesfrom 650 eV to 3320 eV. We review the literature on low energy tails in othermicrocalorimter x-ray spectrometers and present a model that explains all thereviewed tail fract…
▽ More
We use narrow spectral lines from the x-ray spectra of various highlycharged ions to measure low-energy tail-like deviations from a Gaussian responsefunction in a microcalorimter x-ray spectrometer with Au absorbers at energiesfrom 650 eV to 3320 eV. We review the literature on low energy tails in othermicrocalorimter x-ray spectrometers and present a model that explains all thereviewed tail fraction measurements. In this model a low energy tail arises fromthe combination of electron escape and energy trapping associated with Bi x-rayabsorbers.
△ Less