The Poynting effect is a paragon of nonlinear soft matter mechanics. It is the tendency (found in all incompressible, isotropic, hyperelastic solids) exhibited by a soft block to expand vertically when sheared horizontally. It can be observed whenever the length of the cuboid is at least four times its thickness. Here we show that the Poynting effect can be easily reversed and the cuboid can shrin…
▽ More
The Poynting effect is a paragon of nonlinear soft matter mechanics. It is the tendency (found in all incompressible, isotropic, hyperelastic solids) exhibited by a soft block to expand vertically when sheared horizontally. It can be observed whenever the length of the cuboid is at least four times its thickness. Here we show that the Poynting effect can be easily reversed and the cuboid can shrink vertically, simply by reducing this aspect ratio. In principle, this discovery means that for a given solid, say one used as a seismic wave absorber under a building, an optimal ratio exists where vertical displacements and vibrations can be completely eliminated. Here we first recall the classical theoretical treatment of the positive Poynting effect, and then show experimentally how it can be reversed. Using Finite Element simulations, we then investigate how the effect can be suppressed. We find that cubes always provide a reverse Poynting effect, irrespective of their material properties (in the third-order theory of weakly nonlinear elasticity).
△ Less
Submitted 4 June, 2025;
originally announced June 2025.
Experimental assessment of clinical MRI-induced global SAR distributions in head phantoms
Authors:
J. Blackwell,
G. Oluniran,
B. Tuohy,
M. Destrade,
M. J. Kraśny,
N. Colgan
Abstract:
Objective: Accurate estimation of SAR is critical to safeguarding vulnerable patients who require an MRI procedure. The increased static field strength and RF duty cycle capabilities in modern MRI scanners mean that systems can easily exceed safe SAR levels for patients. Advisory protocols routinely used to establish quality assurance protocols are not required to advise on the testing of MRI SAR…
▽ More
Objective: Accurate estimation of SAR is critical to safeguarding vulnerable patients who require an MRI procedure. The increased static field strength and RF duty cycle capabilities in modern MRI scanners mean that systems can easily exceed safe SAR levels for patients. Advisory protocols routinely used to establish quality assurance protocols are not required to advise on the testing of MRI SAR levels and is not routinely measured in annual medical physics quality assurance checks. This study aims to develop a head phantom and protocol that can independently verify global SAR for MRI clinical scanners.
Methods: A four-channel birdcage head coil was used for RF transmission and signal reception. Proton resonance shift thermometry was used to estimate SAR. The SAR estimates were verified by comparing results against two other independent measures, then applied to a further four scanners at field strengths of 1.5 T and 3 T.
Results: Scanner output SAR values ranged from 0.42 to 1.52 W/kg. Percentage SAR differences between independently estimated values and those calculated by the scanners differed by 0-2.3%.
Conclusion: We have developed a quality assurance protocol to independently verify the SAR output of MRI scanners.
△ Less
Submitted 10 September, 2020;
originally announced September 2020.