System performance of a cryogenic test-bed for the time-division multiplexing readout for NewAthena X-IFU
Authors:
Davide Vaccaro,
Jan van der Kuur,
Paul van der Hulst,
Tobias Vos,
Martin de Wit,
Luciano Gottardi,
Kevin Ravensberg,
Emanuele Taralli,
Joseph Adams,
Simon Bandler,
Douglas Bennet,
James Chervenak,
Bertrand Doriese,
Malcolm Durkin,
Johnathon Gard,
Carl Reintsema,
Kazuhiro Sakai,
Steven Smith,
Joel Ullom,
Nicholas Wakeham,
Jan-Willem den Herder,
Brian jackson,
Pourya Khosropanah,
Jian-Rong Gao,
Peter Roelfsema
, et al. (1 additional authors not shown)
Abstract:
The X-ray Integral Field Unit (X-IFU) is an instrument of ESA's future NewAthena space observatory, with the goal to provide high-energy resolution ($<$ 4 eV at X-ray energies up to 7 keV) and high-spatial resolution (9") spectroscopic imaging over the X-ray energy range from 200 eV to 12 keV, by means of an array of about 1500 transition-edge sensors (TES) read out via SQUID time-division multipl…
▽ More
The X-ray Integral Field Unit (X-IFU) is an instrument of ESA's future NewAthena space observatory, with the goal to provide high-energy resolution ($<$ 4 eV at X-ray energies up to 7 keV) and high-spatial resolution (9") spectroscopic imaging over the X-ray energy range from 200 eV to 12 keV, by means of an array of about 1500 transition-edge sensors (TES) read out via SQUID time-division multiplexing (TDM). A TDM-based laboratory test-bed has been assembled at SRON, hosting an array of $75\times 75\ \upmu$m$^2$ TESs that are read out via 2-column $\times$ 32-row TDM. A system component that is critical to high-performance operation is the wiring harness that connects the room-temperature electronics to the cryogenic readout componentry. We report here on our characterization of such a test-bed, whose harness has a length close to what envisioned for X-IFU, which allowed to achieve a co-added energy resolution at a level of 2.7~eV FWHM at 6~keV via 32-row readout. In addition, we provide an outlook on the integration of TDM readout into the X-IFU Focal-Plane Assembly Development Model.
△ Less
Submitted 5 November, 2024; v1 submitted 9 September, 2024;
originally announced September 2024.
-
ASCENT - A balloon-borne hard X-ray imaging spectroscopy telescope using transition edge sensor microcalorimeter detectors
Authors:
Fabian Kislat,
Daniel Becker,
Douglas Bennett,
Adrika Dasgupta,
Joseph Fowler,
Christopher L. Fryer,
Johnathon Gard,
Ephraim Gau,
Danielle Gurgew,
Keon Harmon,
Takayuki Hayashi,
Scott Heatwole,
Md Arman Hossen,
Henric Krawczynski,
R. James Lanzi,
Jason Legere,
John A. B. Mates,
Mark McConnell,
Johanna Nagy,
Takashi Okajima,
Toshiki Sato,
Daniel Schmidt,
Sean Spooner,
Daniel Swetz,
Keisuke Tamura
, et al. (4 additional authors not shown)
Abstract:
Core collapse supernovae are thought to be one of the main sources in the galaxy of elements heavier than iron. Understanding the origin of the elements is thus tightly linked to our understanding of the explosion mechanism of supernovae and supernova nucleosynthesis. X-ray and gamma-ray observations of young supernova remnants, combined with improved theoretical modeling, have resulted in enormou…
▽ More
Core collapse supernovae are thought to be one of the main sources in the galaxy of elements heavier than iron. Understanding the origin of the elements is thus tightly linked to our understanding of the explosion mechanism of supernovae and supernova nucleosynthesis. X-ray and gamma-ray observations of young supernova remnants, combined with improved theoretical modeling, have resulted in enormous improvements in our knowledge of these events. The isotope ${}^{44}$Ti is one of the most sensitive probes of the innermost regions of the core collapse engine, and its spatial and velocity distribution are key observables. Hard X-ray imaging spectroscopy with the Nuclear Spectroscopic Telescope Array (NuSTAR) has provided new insights into the structure of the supernova remnant Cassiopeia A (Cas A), establishing the convective nature of the supernova engine. However, many questions about the details of this engine remain. We present here the concept for a balloon-borne follow-up mission called ASCENT (A SuperConducting ENergetic x-ray Telescope). ASCENT uses transition edge sensor gamma-ray microcalorimeter detectors with a demonstrated 55 eV Full Width Half Maximum (FWHM) energy resolution at 97 keV. This 8--16-fold improvement in energy resolution over NuSTAR will allow high resolution imaging and spectroscopy of the ${}^{44}$Ti emission. This will allow a detailed reconstruction of gamma-ray line redshifts, widths, and shapes, allowing us to address questions such as: What is the source of the neutron star "kicks"? What is the dominant production pathway for ${}^{44}$Ti? Is the engine of Cas A unique?
△ Less
Submitted 4 January, 2023;
originally announced January 2023.
-
The 511-CAM Mission: A Pointed 511 keV Gamma-Ray Telescope with a Focal Plane Detector Made of Stacked Transition Edge Sensor Microcalorimeter Arrays
Authors:
Farzane Shirazi,
Md. Arman Hossen,
Daniel Becker,
Daniel Schmidt,
Daniel Swetz,
Douglas Bennett,
Dana Braun,
Ephraim Gau,
Fabian Kislat,
Johnathon Gard,
John Mates,
Joel Weber,
Nicole Rodriguez Cavero,
Sohee Chun,
Lindsey Lisalda,
Andrew West,
Bhupal Dev,
Francesc Ferrer,
Richard Bose,
Joel Ullom,
Henric Krawczynski
Abstract:
The 511 keV gamma-ray emission from the galactic center region may fully or partially originate from the annihilation of positrons from dark matter particles with electrons from the interstellar medium. Alternatively, the positrons could be created by astrophysical sources, involving exclusively standard model physics. We describe here a new concept for a 511 keV mission called 511-CAM (511 keV ga…
▽ More
The 511 keV gamma-ray emission from the galactic center region may fully or partially originate from the annihilation of positrons from dark matter particles with electrons from the interstellar medium. Alternatively, the positrons could be created by astrophysical sources, involving exclusively standard model physics. We describe here a new concept for a 511 keV mission called 511-CAM (511 keV gamma-ray CAmera using Micro-calorimeters) that combines focusing gamma-ray optics with a stack of Transition Edge Sensor (TES) microcalorimeter arrays in the focal plane. The 511-CAM detector assembly has a projected 511 keV energy resolution of 390 eV Full Width Half Maximum (FWHM) or better, and improves by a factor of at least 11 on the performance of state-of-the-art Ge-based Compton telescopes. Combining this unprecedented energy resolution with sub-arcmin angular resolutions afforded by Laue lens or channeling optics could make substantial contributions to identifying the origin of the 511 keV emission by discovering and characterizing point sources and measuring line-of-sight velocities of the emitting plasmas.
△ Less
Submitted 12 September, 2023; v1 submitted 29 June, 2022;
originally announced June 2022.
-
A Microwave SQUID Multiplexer Optimized for Bolometric Applications
Authors:
B. Dober,
Z. Ahmed,
K. Arnold,
D. T. Becker,
D. A. Bennett,
J. A. Connors,
A. Cukierman,
J. M. D'Ewart,
S. M. Duff,
J. E. Dusatko,
J. C. Frisch,
J. D. Gard,
S. W. Henderson,
R. Herbst,
G. C. Hilton,
J. Hubmayr,
Y. Li,
J. A. B. Mates,
H. McCarrick,
C. D Reintsema,
M. Silva-Feaver,
L. Ruckman,
J. N. Ullom,
L. R. Vale,
D. D. Van Winkle
, et al. (5 additional authors not shown)
Abstract:
A microwave SQUID multiplexer ($μ$MUX) has been optimized for coupling to large arrays of superconducting transition-edge sensor (TES) bolometers. We present the scalable cryogenic multiplexer chip design in a 1820-channel multiplexer configuration for the 4-8 GHz rf band. The key metrics of yield, sensitivity, and crosstalk are determined through measurements of 455 readout channels, which span 4…
▽ More
A microwave SQUID multiplexer ($μ$MUX) has been optimized for coupling to large arrays of superconducting transition-edge sensor (TES) bolometers. We present the scalable cryogenic multiplexer chip design in a 1820-channel multiplexer configuration for the 4-8 GHz rf band. The key metrics of yield, sensitivity, and crosstalk are determined through measurements of 455 readout channels, which span 4-5 GHz. The median white-noise level is 45 pA/$\sqrt{\textrm{Hz}}$, evaluated at 2 Hz, with a 1/f knee $\leq$ 20 mHz after common-mode subtraction. The white-noise level decreases the sensitivity of a TES bolometer optimized for detection of the cosmic microwave background at 150 GHz by only 3%. The measured crosstalk between any channel pair is $\leq$ 0.3%.
△ Less
Submitted 19 January, 2021; v1 submitted 15 October, 2020;
originally announced October 2020.
-
Highly-multiplexed microwave SQUID readout using the SLAC Microresonator Radio Frequency (SMuRF) Electronics for Future CMB and Sub-millimeter Surveys
Authors:
Shawn W. Henderson,
Zeeshan Ahmed,
Jason Austermann,
Daniel Becker,
Douglas A. Bennett,
David Brown,
Saptarshi Chaudhuri,
Hsiao-Mei Sherry Cho,
John M. D'Ewart,
Bradley Dober,
Shannon M. Duff,
John E. Dusatko,
Sofia Fatigoni,
Josef C. Frisch,
Jonathon D. Gard,
Mark Halpern,
Gene C. Hilton,
Johannes Hubmayr,
Kent D. Irwin,
Ethan D. Karpel,
Sarah S. Kernasovskiy,
Stephen E. Kuenstner,
Chao-Lin Kuo,
Dale Li,
John A. B. Mates
, et al. (7 additional authors not shown)
Abstract:
The next generation of cryogenic CMB and submillimeter cameras under development require densely instrumented sensor arrays to meet their science goals. The readout of large numbers ($\sim$10,000--100,000 per camera) of sub-Kelvin sensors, for instance as proposed for the CMB-S4 experiment, will require substantial improvements in cold and warm readout techniques. To reduce the readout cost per se…
▽ More
The next generation of cryogenic CMB and submillimeter cameras under development require densely instrumented sensor arrays to meet their science goals. The readout of large numbers ($\sim$10,000--100,000 per camera) of sub-Kelvin sensors, for instance as proposed for the CMB-S4 experiment, will require substantial improvements in cold and warm readout techniques. To reduce the readout cost per sensor and integration complexity, efforts are presently focused on achieving higher multiplexing density while maintaining readout noise subdominant to intrinsic detector noise. Highly-multiplexed cold readout technologies in active development include Microwave Kinetic Inductance Sensors (MKIDs) and microwave rf-SQUIDs. Both exploit the high quality factors of superconducting microwave resonators to densely channelize sub-Kelvin sensors into the bandwidth of a microwave transmission line. We present advancements in the development of a new warm readout system for microwave SQUID multiplexing, the SLAC Superconducting Microresonator RF electronics, or SMuRF. The SMuRF system is unique in its ability to track each tone, minimizing the total RF power required to read out each resonator, thereby significantly reducing the linearity requirements on the cold and warm readout. Here, we present measurements of the readout noise and linearity of the first full SMuRF system, including a demonstration of closed-loop tone tracking on a 528 channel cryogenic microwave SQUID multiplexer. SMuRF is being explored as a potential readout solution for several future CMB projects including Simons Observatory, BICEP Array, CCAT-prime, Ali-CPT, and CMB-S4. Parallel development of the platform is underway to adapt SMuRF to read out both MKID and fast X-ray TES calorimeter arrays.
△ Less
Submitted 11 September, 2018;
originally announced September 2018.
-
Microwave SQUID Multiplexer demonstration for Cosmic Microwave Background Imagers
Authors:
B. Dober,
D. T. Becker,
D. A. Bennett,
S. A. Bryan,
S. M. Duff,
J. D. Gard,
J. P. Hays-Wehle,
G. C. Hilton,
J. Hubmayr,
J. A. B. Mates,
C. D. Reintsema,
L. R. Vale,
J. N. Ullom
Abstract:
Key performance characteristics are demonstrated for the microwave SQUID multiplexer ($μ$MUX) coupled to transition edge sensor (TES) bolometers that have been optimized for cosmic microwave background (CMB) observations. In a 64-channel demonstration, we show that the $μ$MUX produces a white, input referred current noise level of 29~pA$/\sqrt{\mathrm{Hz}}$ at -77~dB microwave probe tone power, wh…
▽ More
Key performance characteristics are demonstrated for the microwave SQUID multiplexer ($μ$MUX) coupled to transition edge sensor (TES) bolometers that have been optimized for cosmic microwave background (CMB) observations. In a 64-channel demonstration, we show that the $μ$MUX produces a white, input referred current noise level of 29~pA$/\sqrt{\mathrm{Hz}}$ at -77~dB microwave probe tone power, which is well below expected fundamental detector and photon noise sources for a ground-based CMB-optimized bolometer. Operated with negligible photon loading, we measure 98~pA$/\sqrt{\mathrm{Hz}}$ in the TES-coupled channels biased at 65% of the sensor normal resistance. This noise level is consistent with that predicted from bolometer thermal fluctuation (i.e., phonon) noise. Furthermore, the power spectral density exhibits a white spectrum at low frequencies ($\sim$~100~mHz), which enables CMB mapping on large angular scales that constrain the physics of inflation. Additionally, we report cross-talk measurements that indicate a level below 0.3%, which is less than the level of cross-talk from multiplexed readout systems in deployed CMB imagers. These measurements demonstrate the $μ$MUX as a viable readout technique for future CMB imaging instruments.
△ Less
Submitted 15 December, 2017; v1 submitted 11 October, 2017;
originally announced October 2017.