Calibration and GEANT4 Simulations of the Phase II Proton Compute Tomography (pCT) Range Stack Detector
Authors:
S. A. Uzunyan,
G. Blazey,
S. Boi,
G. Coutrakon,
A. Dyshkant,
K. Francis,
D. Hedin,
E. Johnson,
J. Kalnins,
V. Zutshi,
R. Ford,
J. E. Rauch,
P. Rubinov,
G. Sellberg,
P. Wilson,
M. Naimuddin
Abstract:
Northern Illinois University in collaboration with Fermi National Accelerator Laboratory (FNAL) and Delhi University has been designing and building a proton CT scanner for applications in proton treatment planning. The Phase II proton CT scanner consists of eight planes of tracking detectors with two X and two Y coordinate measurements both before and after the patient. In addition, a range stack…
▽ More
Northern Illinois University in collaboration with Fermi National Accelerator Laboratory (FNAL) and Delhi University has been designing and building a proton CT scanner for applications in proton treatment planning. The Phase II proton CT scanner consists of eight planes of tracking detectors with two X and two Y coordinate measurements both before and after the patient. In addition, a range stack detector consisting of a stack of thin scintillator tiles, arranged in twelve eight-tile frames, is used to determine the water equivalent path length (WEPL) of each track through the patient. The X-Y coordinates and WEPL are required input for image reconstruction software to find the relative (proton) stopping powers (RSP) value of each voxel in the patient and generate a corresponding 3D image. In this Note we describe tests conducted in 2015 at the proton beam at the Central DuPage Hospital in Warrenville, IL, focusing on the range stack calibration procedure and comparisons with the GEANT~4 range stack simulation.
△ Less
Submitted 4 January, 2016; v1 submitted 3 January, 2016;
originally announced January 2016.
-
Development of a proton Computed Tomography Detector System
Authors:
Md. Naimuddin,
G. Coutrakon,
G. Blazey,
S. Boi,
A. Dyshkant,
B. Erdelyi,
D. Hedin,
E. Johnson,
J. Krider,
V. Rukalin,
S. A. Uzunyan,
V. Zutshi,
R. Fordt,
G. Sellberg,
J. E. Rauch,
M. Roman,
P. Rubinov,
P. Wilson
Abstract:
Computer tomography is one of the most promising new methods to image abnormal tissues inside the human body. Tomography is also used to position the patient accurately before radiation therapy. Hadron therapy for treating cancer has become one of the most advantageous and safe options. In order to fully utilize the advantages of hadron therapy, there is a necessity of performing radiography with…
▽ More
Computer tomography is one of the most promising new methods to image abnormal tissues inside the human body. Tomography is also used to position the patient accurately before radiation therapy. Hadron therapy for treating cancer has become one of the most advantageous and safe options. In order to fully utilize the advantages of hadron therapy, there is a necessity of performing radiography with hadrons as well. In this paper we present the development of a proton computed tomography system. Our second-generation proton tomography system consists of two upstream and two downstream trackers made up of fibers as active material and a range detector consisting of plastic scintillators. We present details of the detector system, readout electronics, and data acquisition system as well as the commissioning of the entire system. We also present preliminary results from the test beam of the range detector.
△ Less
Submitted 1 October, 2015;
originally announced October 2015.
-
A New Proton CT Scanner
Authors:
S. A. Uzunyan,
G. Blazey,
S. Boi,
G. Coutrakon,
A. Dyshkant,
B. Erdelyi,
D. Hedin,
E. Johnson,
J. Krider,
V. Rykalin,
V. Zutshi,
R. Ford,
G. Sellberg,
J . E. Rauch,
M. Roman,
P. Rubinov,
P. Wilson,
M. Naimuddin
Abstract:
The design, construction, and preliminary testing of a second generation proton CT scanner is presented. All current treatment planning systems at proton therapy centers use X-ray CT as the primary imaging modality for treatment planning to calculate doses to tumor and healthy tissues. One of the limitations of X-ray CT is in the conversion of X-ray attenuation coefficients to relative (proton) st…
▽ More
The design, construction, and preliminary testing of a second generation proton CT scanner is presented. All current treatment planning systems at proton therapy centers use X-ray CT as the primary imaging modality for treatment planning to calculate doses to tumor and healthy tissues. One of the limitations of X-ray CT is in the conversion of X-ray attenuation coefficients to relative (proton) stopping powers, or RSP. This results in more proton range uncertainty, larger target volumes and therefore, more dose to healthy tissues. To help improve this, we present a novel scanner capable of high dose rates, up to 2~MHz, and large area coverage, 20~x~24~cm$^2$, for imaging an adult head phantom and reconstructing more accurate RSP values.
△ Less
Submitted 4 January, 2016; v1 submitted 29 August, 2014;
originally announced September 2014.
-
Development of a proton Computed Tomography (pCT) scanner at NIU
Authors:
S. A. Uzunyan,
G. Blazey,
S. Boi,
G. Coutrakon,
A. Dyshkant,
B. Erdelyi,
A. Gearhart,
D. Hedin,
E. Johnson,
J. Krider,
V. Zutshi,
R. Ford,
T. Fitzpatrick,
G. Sellberg,
J. E. Rauch,
M. Roman,
P. Rubinov,
P. Wilson,
K. Lalwani,
M. Naimuddin
Abstract:
We describe the development of a proton Computed Tomography (pCT) scanner at Northern Illinois University (NIU) in collaboration with Fermilab and Delhi University. This paper provides an overview of major components of the scanner and a detailed description of the data acquisition system (DAQ).
We describe the development of a proton Computed Tomography (pCT) scanner at Northern Illinois University (NIU) in collaboration with Fermilab and Delhi University. This paper provides an overview of major components of the scanner and a detailed description of the data acquisition system (DAQ).
△ Less
Submitted 13 December, 2013;
originally announced December 2013.