Reconstruction of proton relative stopping power with a granular calorimeter detector model
Authors:
M. Aehle,
J. Alme,
G. G. Barnaföldi,
G. Bíró,
T. Bodova,
V. Borshchov,
A. van den Brink,
M. Chaar,
B. Dudás,
V. Eikeland,
G. Feofilov,
C. Garth,
N. R. Gauger,
O. Grøttvik,
H. Helstrup,
S. Igolkin,
Zs. Jólesz,
R. Keidel,
C. Kobdaj,
T. Kortus,
L. Kusch,
V. Leonhardt,
S. Mehendale,
R. Ningappa,
O. H. Odland
, et al. (27 additional authors not shown)
Abstract:
Proton computed tomography (pCT) aims to facilitate precise dose planning for hadron therapy, a promising and effective method for cancer treatment. Hadron therapy utilizes protons and heavy ions to deliver well focused doses of radiation, leveraging the Bragg peak phenomenon to target tumors while sparing healthy tissues. The Bergen pCT Collaboration aims to develop a novel pCT scanner, and accom…
▽ More
Proton computed tomography (pCT) aims to facilitate precise dose planning for hadron therapy, a promising and effective method for cancer treatment. Hadron therapy utilizes protons and heavy ions to deliver well focused doses of radiation, leveraging the Bragg peak phenomenon to target tumors while sparing healthy tissues. The Bergen pCT Collaboration aims to develop a novel pCT scanner, and accompanying reconstruction algorithms to overcome current limitations. This paper focuses on advancing the track- and image reconstruction algorithms, thereby enhancing the precision of the dose planning and reducing side effects of hadron therapy. A neural network aided track reconstruction method is presented.
△ Less
Submitted 4 March, 2025;
originally announced March 2025.
-
ALICE ITS3: a bent stitched MAPS-based vertex detector
Authors:
Ola Groettvik
Abstract:
The ALICE ITS3 is a novel vertex detector replacing the innermost layers of ITS2 during LS3. Composed of three truly cylindrical layers of wafer-sized 65 nm stitched Monolithic Active Pixel Sensors, ITS3 provides high-resolution tracking of charged particles generated in heavy-ion collisions. This contribution presents an overview of the ITS3 detector, highlighting its design features, integration…
▽ More
The ALICE ITS3 is a novel vertex detector replacing the innermost layers of ITS2 during LS3. Composed of three truly cylindrical layers of wafer-sized 65 nm stitched Monolithic Active Pixel Sensors, ITS3 provides high-resolution tracking of charged particles generated in heavy-ion collisions. This contribution presents an overview of the ITS3 detector, highlighting its design features, integration and cooling, and the ongoing development towards the final sensor. Furthermore, the paper introduces the off-detector service electronics, which play an essential role in the readout, control, and power supply of the detector.
△ Less
Submitted 10 January, 2024; v1 submitted 9 January, 2024;
originally announced January 2024.
-
Performance of the Electromagnetic Pixel Calorimeter Prototype EPICAL-2
Authors:
J. Alme,
R. Barthel,
A. van Bochove,
V. Borshchov,
R. Bosley,
A. van den Brink,
E. Broeils,
H. Büsching,
V. N. Eikeland,
O. S. Groettvik,
Y. H. Han,
N. van der Kolk,
J. H. Kim,
T. J. Kim,
Y. Kwon,
M. Mager,
Q. W. Malik,
E. Okkinga,
T. Y. Park,
T. Peitzmann,
F. Pliquett,
M. Protsenko,
F. Reidt,
S. van Rijk,
K. Røed
, et al. (9 additional authors not shown)
Abstract:
The first evaluation of an ultra-high granularity digital electromagnetic calorimeter prototype using 1.0-5.8 GeV/c electrons is presented. The $25\times10^6$ pixel detector consists of 24 layers of ALPIDE CMOS MAPS sensors, with a pitch of around 30~$μ$m, and has a depth of almost 20 radiation lengths of tungsten absorber. Ultra-thin cables allow for a very compact design. The properties that are…
▽ More
The first evaluation of an ultra-high granularity digital electromagnetic calorimeter prototype using 1.0-5.8 GeV/c electrons is presented. The $25\times10^6$ pixel detector consists of 24 layers of ALPIDE CMOS MAPS sensors, with a pitch of around 30~$μ$m, and has a depth of almost 20 radiation lengths of tungsten absorber. Ultra-thin cables allow for a very compact design. The properties that are critical for physics studies are measured: electromagnetic shower response, energy resolution and linearity. The stochastic energy resolution is comparable with the state-of-the art resolution for a Si-W calorimeter, with data described well by a simulation model using GEANT and Allpix$^2$. The performance achieved makes this technology a good candidate for use in the ALICE FoCal upgrade, and in general demonstrates the strong potential for future applications in high-energy physics.
△ Less
Submitted 28 December, 2022; v1 submitted 6 September, 2022;
originally announced September 2022.
-
Results from the EPICAL-2 Ultra-High Granularity Electromagnetic Calorimeter Prototype
Authors:
T. Peitzmann,
J. Alme,
R. Barthel,
A. van Bochove,
V. Borshchov,
R. Bosley,
A. van den Brink,
E. Broeils,
H. Büsching,
V. N. Eikeland,
O. S. Groettvik,
Y. H. Han,
N. van der Kolk,
J. H. Kim,
T. J. Kim,
Y. Kwon,
M. Mager,
Q. W. Malik,
E. Okkinga,
T. Y. Park,
F. Pliquett,
M. Protsenko,
F. Reidt,
S. van Rijk,
K. Røed
, et al. (9 additional authors not shown)
Abstract:
A prototype of a new type of calorimeter has been designed and constructed, based on a silicon-tungsten sampling design using pixel sensors with digital readout. It makes use of the Alpide MAPS sensor developed for the ALICE ITS upgrade. A binary readout is possible due to the pixel size of $\approx 30 \times 30 \, μ\mathrm{m}^2$. This prototype has been successfully tested with cosmic muons and w…
▽ More
A prototype of a new type of calorimeter has been designed and constructed, based on a silicon-tungsten sampling design using pixel sensors with digital readout. It makes use of the Alpide MAPS sensor developed for the ALICE ITS upgrade. A binary readout is possible due to the pixel size of $\approx 30 \times 30 \, μ\mathrm{m}^2$. This prototype has been successfully tested with cosmic muons and with test beams at DESY and the CERN SPS. We report on performance results obtained at DESY, showing good energy resolution and linearity, and compare to detailed MC simulations. Also shown are preliminary results of the high-energy performance as measured at the SPS. The two-shower separation capabilities are discussed.
△ Less
Submitted 27 September, 2022; v1 submitted 5 July, 2022;
originally announced July 2022.
-
Exploration of Differentiability in a Proton Computed Tomography Simulation Framework
Authors:
Max Aehle,
Johan Alme,
Gergely Gábor Barnaföldi,
Johannes Blühdorn,
Tea Bodova,
Vyacheslav Borshchov,
Anthony van den Brink,
Viljar Eikeland,
Gregory Feofilov,
Christoph Garth,
Nicolas R. Gauger,
Ola Grøttvik,
Håvard Helstrup,
Sergey Igolkin,
Ralf Keidel,
Chinorat Kobdaj,
Tobias Kortus,
Lisa Kusch,
Viktor Leonhardt,
Shruti Mehendale,
Raju Ningappa Mulawade,
Odd Harald Odland,
George O'Neill,
Gábor Papp,
Thomas Peitzmann
, et al. (25 additional authors not shown)
Abstract:
Objective. Algorithmic differentiation (AD) can be a useful technique to numerically optimize design and algorithmic parameters by, and quantify uncertainties in, computer simulations. However, the effectiveness of AD depends on how "well-linearizable" the software is. In this study, we assess how promising derivative information of a typical proton computed tomography (pCT) scan computer simulati…
▽ More
Objective. Algorithmic differentiation (AD) can be a useful technique to numerically optimize design and algorithmic parameters by, and quantify uncertainties in, computer simulations. However, the effectiveness of AD depends on how "well-linearizable" the software is. In this study, we assess how promising derivative information of a typical proton computed tomography (pCT) scan computer simulation is for the aforementioned applications.
Approach. This study is mainly based on numerical experiments, in which we repeatedly evaluate three representative computational steps with perturbed input values. We support our observations with a review of the algorithmic steps and arithmetic operations performed by the software, using debugging techniques.
Main results. The model-based iterative reconstruction (MBIR) subprocedure (at the end of the software pipeline) and the Monte Carlo (MC) simulation (at the beginning) were piecewise differentiable. Jumps in the MBIR function arose from the discrete computation of the set of voxels intersected by a proton path. Jumps in the MC function likely arose from changes in the control flow that affect the amount of consumed random numbers. The tracking algorithm solves an inherently non-differentiable problem.
Significance. The MC and MBIR codes are ready for the integration of AD, and further research on surrogate models for the tracking subprocedure is necessary.
△ Less
Submitted 12 May, 2023; v1 submitted 11 February, 2022;
originally announced February 2022.
-
First demonstration of in-beam performance of bent Monolithic Active Pixel Sensors
Authors:
ALICE ITS project,
:,
G. Aglieri Rinella,
M. Agnello,
B. Alessandro,
F. Agnese,
R. S. Akram,
J. Alme,
E. Anderssen,
D. Andreou,
F. Antinori,
N. Apadula,
P. Atkinson,
R. Baccomi,
A. Badalà,
A. Balbino,
C. Bartels,
R. Barthel,
F. Baruffaldi,
I. Belikov,
S. Beole,
P. Becht,
A. Bhatti,
M. Bhopal,
N. Bianchi
, et al. (230 additional authors not shown)
Abstract:
A novel approach for designing the next generation of vertex detectors foresees to employ wafer-scale sensors that can be bent to truly cylindrical geometries after thinning them to thicknesses of 20-40$μ$m. To solidify this concept, the feasibility of operating bent MAPS was demonstrated using 1.5$\times$3cm ALPIDE chips. Already with their thickness of 50$μ$m, they can be successfully bent to ra…
▽ More
A novel approach for designing the next generation of vertex detectors foresees to employ wafer-scale sensors that can be bent to truly cylindrical geometries after thinning them to thicknesses of 20-40$μ$m. To solidify this concept, the feasibility of operating bent MAPS was demonstrated using 1.5$\times$3cm ALPIDE chips. Already with their thickness of 50$μ$m, they can be successfully bent to radii of about 2cm without any signs of mechanical or electrical damage. During a subsequent characterisation using a 5.4GeV electron beam, it was further confirmed that they preserve their full electrical functionality as well as particle detection performance.
In this article, the bending procedure and the setup used for characterisation are detailed. Furthermore, the analysis of the beam test, including the measurement of the detection efficiency as a function of beam position and local inclination angle, is discussed. The results show that the sensors maintain their excellent performance after bending to radii of 2cm, with detection efficiencies above 99.9% at typical operating conditions, paving the way towards a new class of detectors with unprecedented low material budget and ideal geometrical properties.
△ Less
Submitted 17 August, 2021; v1 submitted 27 May, 2021;
originally announced May 2021.