On the Electrode Configurations in a Large Single Phase Liquid Xenon Detector for Dark Matter Searches
Authors:
P. Juyal,
K. L. Giboni,
X. Ji,
J. Liu
Abstract:
In the near future there will be the request for very large liquid Xenon (LXe) detectors for Dark Matter (DM) searches in the 50-ton range. To avoid an impractically long, single drift space of a dual-phase detector, it seems beneficial to use the single-phase technique. Since electrons then can drift in any direction, we can segment the homogeneous medium and thus avoid an excessive maximum drift…
▽ More
In the near future there will be the request for very large liquid Xenon (LXe) detectors for Dark Matter (DM) searches in the 50-ton range. To avoid an impractically long, single drift space of a dual-phase detector, it seems beneficial to use the single-phase technique. Since electrons then can drift in any direction, we can segment the homogeneous medium and thus avoid an excessive maximum drift path of order 4 m. The shorter detector length has several benefits, e.g. requiring a lower cathode voltage for the same drift field. We can easily split the TPC into two regions with the cathode in the center and two anodes at the top and bottom. One also can use multiple TPCs stacked on top of each other in the same liquid volume to reduce the maximum drift length even further.
A further division of the drift space by installing an additional anode in the center would require S2 photons to traverse the liquid for several times the Rayleigh scattering length in LXe, which is only 30 - 40 cm. This seems to be excessive for good x - y localization. We therefore suggest a geometry of two independent TPCs with two drift spaces each.
Despite earlier publications concerns persisted about the effect of shadowing. A detailed FEM model of the anode regions shows that with an aligned wire arrangement the drifting electrons impinge sideways on the anode in a narrow angular range of width 15$^{\circ}$ - 20$^{\circ}$. Most S2 photons are emitted in full view of the close-by PMT array. About 37% of the S2 photons are shadowed by the anode wire out of which 30\% will be reflected back again on the gold plating of the wires. Thus we can observe 74% of the total S2 light. Compared to a dual-phase detector, however, we do not suffer from the extraction efficiency, sometimes reported as low as 50%.
△ Less
Submitted 16 July, 2021;
originally announced July 2021.
-
On Proportional Scintillation in Very Large LXe Detectors
Authors:
P. Juyal,
K. L. Giboni,
X. Ji,
J. Liu
Abstract:
The charge read out of a LXe detector via Proportional Scintillation in the liquid phase was first realized by the Waseda group 40 years ago, but at that time the technical challenges were overwhelming. Although the tests were successful, this method was finally discarded and eventually nearly forgotten. For present day large LXe Dark Matter detectors, this approach was not even considered. Instea…
▽ More
The charge read out of a LXe detector via Proportional Scintillation in the liquid phase was first realized by the Waseda group 40 years ago, but at that time the technical challenges were overwhelming. Although the tests were successful, this method was finally discarded and eventually nearly forgotten. For present day large LXe Dark Matter detectors, this approach was not even considered. Instead the Dual Phase technology was selected despite many limitations and challenges. In two independent studies the groups from Columbia University and Shanghai Jiao Tong University reevaluated Proportional Scintillation in the liquid phase. Both established the merits for very large LXe detectors, but the Columbia group also encountered apparent limitations, namely the shadowing of the light by the anode wires and a dependence of the pulse shape on the drift path of the electrons in the anode region. The discrepancies between the two studies, however, are not intrinsic to the technique, but a direct consequence of the chosen geometry. Taking the geometrical differences into account the results match without ambiguity. They also agree with the original results from the Waseda group. With the technical problems solved, the path is now open to use this method in future large LXe TPCs.
△ Less
Submitted 29 October, 2019;
originally announced October 2019.
-
A LN$_2$ Based Cooling System for a Next Generation Liquid Xenon Dark Matter Detector
Authors:
K. L. Giboni,
P. Juyal,
E. Aprile,
Y. Zhang,
J. Naganoma
Abstract:
In recent years cooling technology for Liquid Xenon (LXe) detectors has advanced driven by the development of Dark Matter (DM) detectors with target mass in the 100 - 1,000 kg range. The next generation of DM detectors based on LXe will be in the 50,000 kg (50 t) range requiring more than 1 kW of cooling power. Most of the prior cooling methods become impractical at this level. For cooling a 50 t…
▽ More
In recent years cooling technology for Liquid Xenon (LXe) detectors has advanced driven by the development of Dark Matter (DM) detectors with target mass in the 100 - 1,000 kg range. The next generation of DM detectors based on LXe will be in the 50,000 kg (50 t) range requiring more than 1 kW of cooling power. Most of the prior cooling methods become impractical at this level. For cooling a 50 t scale LXe detector, a method is proposed in which Liquid Nitrogen (LN$_2$) in a small local reservoir cools the xenon gas via a cold finger. The cold finger incorporates a heating unit to provide temperature regulation.
The proposed cooling method is simple, reliable, and suitable for the required long-term operation for a rare event search. The device can be easily integrated into present cooling systems, e.g. the 'Cooling Bus' employed for the PandaX I and II experiments. It is still possible to cool indirectly with no part of the cooling or temperature control system getting in direct contact with the clean xenon in the detector. Also the cooling device can be mounted at a large distance, i.e. the detector is cooled remotely from a distance of 5 - 10 m. The method was tested in a laboratory setup at Columbia University to carry out different measurements with a small LXe detector and behaved exactly as predicted.
△ Less
Submitted 20 September, 2019;
originally announced September 2019.
-
Performance of Photosensors in the PandaX-I Experiment
Authors:
Shaoli Li,
Xun Chen,
Karl L. Giboni,
Guodong Guo,
Xiangdong Ji,
Qing Lin,
Jianglai Liu,
Yajun Mao,
Kaixuan Ni,
Xiangxiang Ren,
Andi Tan,
Mengjiao Xiao,
Xiaopeng Zhou
Abstract:
We report the long term performance of the photosensors, 143 one-inch R8520-406 and 37 three-inch R11410-MOD photomultipliers from Hamamatsu, in the first phase of the PandaX dual-phase xenon dark matter experiment. This is the first time that a significant number of R11410 photomultiplier tubes were operated in liquid xenon for an extended period, providing important guidance to the future large…
▽ More
We report the long term performance of the photosensors, 143 one-inch R8520-406 and 37 three-inch R11410-MOD photomultipliers from Hamamatsu, in the first phase of the PandaX dual-phase xenon dark matter experiment. This is the first time that a significant number of R11410 photomultiplier tubes were operated in liquid xenon for an extended period, providing important guidance to the future large xenon-based dark matter experiments.
△ Less
Submitted 29 January, 2016; v1 submitted 18 November, 2015;
originally announced November 2015.
-
Low-mass dark matter search results from full exposure of PandaX-I experiment
Authors:
PandaX Collaboration,
Xiang Xiao,
Xun Chen,
Andi Tan,
Yunhua Chen,
Xiangyi Cui,
Deqing Fang,
Changbo Fu,
Karl L. Giboni,
Haowei Gong,
Guodong Guo,
Ming He,
Xiangdong Ji,
Yonglin Ju,
Siao Lei,
Shaoli Li,
Qing Lin,
Huaxuan Liu,
Jianglai Liu,
Xiang Liu,
Wolfgang Lorenzon,
Yugang Ma,
Yajun Mao,
Kaixuan Ni,
Kirill Pushkin
, et al. (21 additional authors not shown)
Abstract:
We report the results of a weakly-interacting massive particle (WIMP) dark matter search using the full 80.1\;live-day exposure of the first stage of the PandaX experiment (PandaX-I) located in the China Jin-Ping Underground Laboratory. The PandaX-I detector has been optimized for detecting low-mass WIMPs, achieving a photon detection efficiency of 9.6\%. With a fiducial liquid xenon target mass o…
▽ More
We report the results of a weakly-interacting massive particle (WIMP) dark matter search using the full 80.1\;live-day exposure of the first stage of the PandaX experiment (PandaX-I) located in the China Jin-Ping Underground Laboratory. The PandaX-I detector has been optimized for detecting low-mass WIMPs, achieving a photon detection efficiency of 9.6\%. With a fiducial liquid xenon target mass of 54.0\,kg, no significant excess event were found above the expected background. A profile likelihood analysis confirms our earlier finding that the PandaX-I data disfavor all positive low-mass WIMP signals reported in the literature under standard assumptions. A stringent bound on the low mass WIMP is set at WIMP mass below 10\,GeV/c$^2$, demonstrating that liquid xenon detectors can be competitive for low-mass WIMP searches.
△ Less
Submitted 20 August, 2015; v1 submitted 4 May, 2015;
originally announced May 2015.
-
First dark matter search results from the PandaX-I experiment
Authors:
PandaX Collaboration,
Mengjiao Xiao,
Xiang Xiao,
Li Zhao,
Xiguang Cao,
Xun Chen,
Yunhua Chen,
Xiangyi Cui,
Deqing Fang,
Changbo Fu,
Karl L. Giboni,
Haowei Gong,
Guodong Guo,
Jie Hu,
Xingtao Huang,
Xiangdong Ji,
Yonglin Ju,
Siao Lei,
Shaoli Li,
Qing Lin,
Huaxuan Liu,
Jianglai Liu,
Xiang Liu,
Wolfgang Lorenzon,
Yugang Ma
, et al. (22 additional authors not shown)
Abstract:
We report on the first dark-matter (DM) search results from PandaX-I, a low threshold dual-phase xenon experiment operating at the China Jinping Underground Laboratory. In the 37-kg liquid xenon target with 17.4 live-days of exposure, no DM particle candidate event was found. This result sets a stringent limit for low-mass DM particles and disfavors the interpretation of previously-reported positi…
▽ More
We report on the first dark-matter (DM) search results from PandaX-I, a low threshold dual-phase xenon experiment operating at the China Jinping Underground Laboratory. In the 37-kg liquid xenon target with 17.4 live-days of exposure, no DM particle candidate event was found. This result sets a stringent limit for low-mass DM particles and disfavors the interpretation of previously-reported positive experimental results. The minimum upper limit, $3.7\times10^{-44}$\,cm$^2$, for the spin-independent isoscalar DM-particle-nucleon scattering cross section is obtained at a DM-particle mass of 49\,GeV/c$^2$ at 90\% confidence level.
△ Less
Submitted 4 September, 2014; v1 submitted 21 August, 2014;
originally announced August 2014.
-
PandaX: A Liquid Xenon Dark Matter Experiment at CJPL
Authors:
X. G. Cao,
X. Chen,
Y. H. Chen,
X. Y. Cui,
D. Q. Fang,
C. B. Fu,
K. L. Giboni,
H. W. Gong,
G. D. Guo,
M. He,
J. Hu,
X. T. Huang,
X. D. Ji,
Y. L. Ju,
S. L. Li,
Q. Lin,
H. X. Liu,
J. L. Liu,
X. Liu,
W. Lorenzon,
Y. G. Ma,
Y. J. Mao,
K. X. Ni,
K. Pushkin,
X. X. Ren
, et al. (25 additional authors not shown)
Abstract:
PandaX is a large upgradable liquid-xenon detector system that can be used for both direct dark-matter detection and $^{136}$Xe double-beta decay search. It is located in the Jinping Deep-Underground Laboratory in Sichuan, China. The detector operates in dual-phase mode, allowing detection of both prompt scintillation, and ionization charge through proportional scintillation. The central time proj…
▽ More
PandaX is a large upgradable liquid-xenon detector system that can be used for both direct dark-matter detection and $^{136}$Xe double-beta decay search. It is located in the Jinping Deep-Underground Laboratory in Sichuan, China. The detector operates in dual-phase mode, allowing detection of both prompt scintillation, and ionization charge through proportional scintillation. The central time projection chamber will be staged, with the first stage accommodating a target mass of about 120\,kg. In stage II, the target mass will be increased to about 0.5\,ton. In the final stage, the detector can be upgraded to a multi-ton target mass. In this paper a detailed description of the stage-I detector design and performance results established during the commissioning phase is presented.
△ Less
Submitted 12 May, 2014;
originally announced May 2014.
-
The Cryogenic System for the Panda-X Dark Matter Search Experiment
Authors:
Zhao Li,
Karl Ludwig Giboni,
Haowei Gong,
Xiangdong Ji,
Andy Tan
Abstract:
Panda-X is a liquid xenon dual-phase detector for the Dark Matter Search. The first modestly-sized module will soon be installed in the China JinPing Deep Underground Laboratory in Sichuan province, P.R. China. The cryogenics system is designed to handle much larger detectors, even the final version in the ton scale. Special attention has been paid to the reliability, serviceability, and adaptabil…
▽ More
Panda-X is a liquid xenon dual-phase detector for the Dark Matter Search. The first modestly-sized module will soon be installed in the China JinPing Deep Underground Laboratory in Sichuan province, P.R. China. The cryogenics system is designed to handle much larger detectors, even the final version in the ton scale. Special attention has been paid to the reliability, serviceability, and adaptability to the requirements of a growing experiment. The system is cooled by a single Iwatani PC150 Pulse Tube Refrigerator. After subtracting all thermal losses, the remaining cooling power is still 82W. The fill speed was 9 SLPM, but could be boosted by LN2 assisted cooling to 40 SLPM. For the continuous recirculation and purification through a hot getter, a heat exchanger was employed to reduce the required cooling power. The recirculation speed is limited to 35 SLPM by the gas pump. At this speed, recirculation only adds 18.5 W to the heat load of the system, corresponding to a 95.2 % efficiency of the heat exchanger.
△ Less
Submitted 21 July, 2012;
originally announced July 2012.
-
The XENON100 Dark Matter Experiment
Authors:
XENON100 Collaboration,
E. Aprile,
K. Arisaka,
F. Arneodo,
A. Askin,
L. Baudis,
A. Behrens,
E. Brown,
J. M. R. Cardoso,
B. Choi,
D. Cline,
S. Fattori,
A. D. Ferella,
K. L. Giboni,
A. Kish,
C. W. Lam,
R. F. Lang,
K. E. Lim,
J. A. M. Lopes,
T. Marrodan Undagoitia,
Y. Mei,
A. J. Melgarejo Fernandez,
K. Ni,
U. Oberlack,
S. E. A. Orrigo
, et al. (11 additional authors not shown)
Abstract:
The XENON100 dark matter experiment uses liquid xenon (LXe) in a time projection chamber (TPC) to search for Xe nuclear recoils resulting from the scattering of dark matter Weakly Interacting Massive Particles (WIMPs). In this paper we present a detailed description of the detector design and present performance results, as established during the commissioning phase and during the first science ru…
▽ More
The XENON100 dark matter experiment uses liquid xenon (LXe) in a time projection chamber (TPC) to search for Xe nuclear recoils resulting from the scattering of dark matter Weakly Interacting Massive Particles (WIMPs). In this paper we present a detailed description of the detector design and present performance results, as established during the commissioning phase and during the first science runs.
The active target of XENON100 contains 62 kg of LXe, surrounded by an LXe veto of 99 kg, both instrumented with photomultiplier tubes (PMTs) operating inside the liquid or in Xe gas. The LXe target and veto are contained in a low-radioactivity stainless steel vessel, embedded in a passive radiation shield. The experiment is installed underground at the Laboratori Nazionali del Gran Sasso (LNGS), Italy and has recently published results from a 100 live-days dark matter search. The ultimate design goal of XENON100 is to achieve a spin-independent WIMP-nucleon scattering cross section sensitivity of σ= 2x10^-45 cm^2 for a 100 GeV/c^2 WIMP.
△ Less
Submitted 8 February, 2012; v1 submitted 11 July, 2011;
originally announced July 2011.
-
A search for light dark matter in XENON10 data
Authors:
J. Angle,
E. Aprile,
F. Arneodo,
L. Baudis,
A. Bernstein,
A. Bolozdynya,
P. Brusov,
L. C. C. Coelho,
C. E. Dahl,
L. DeViveiros,
A. D. Ferella,
L. M. P. Fernandes,
S. Fiorucci,
R. J. Gaitskell,
K. L. Giboni,
R. Gomez,
R. Hasty,
L. Kastens,
J. Kwong,
J. A. M. Lopes,
N. Madden,
A. Manalaysay,
A. Manzur,
D. N. McKinsey,
M. E. Monzani
, et al. (12 additional authors not shown)
Abstract:
We report results of a search for light (<10 GeV) particle dark matter with the XENON10 detector. The event trigger was sensitive to a single electron, with the analysis threshold of 5 electrons corresponding to 1.4 keV nuclear recoil energy. Considering spin-independent dark matter-nucleon scattering, we exclude cross sections σ_n>3.5x10^{-42} cm^2, for a dark matter particle mass m_χ=8 GeV. We f…
▽ More
We report results of a search for light (<10 GeV) particle dark matter with the XENON10 detector. The event trigger was sensitive to a single electron, with the analysis threshold of 5 electrons corresponding to 1.4 keV nuclear recoil energy. Considering spin-independent dark matter-nucleon scattering, we exclude cross sections σ_n>3.5x10^{-42} cm^2, for a dark matter particle mass m_χ=8 GeV. We find that our data strongly constrain recent elastic dark matter interpretations of excess low-energy events observed by CoGeNT and CRESST-II, as well as the DAMA annual modulation signal.
△ Less
Submitted 7 May, 2013; v1 submitted 15 April, 2011;
originally announced April 2011.
-
Material screening and selection for XENON100
Authors:
XENON100 Collaboration,
E. Aprile,
K. Arisaka,
F. Arneodo,
A. Askin,
L. Baudis,
A. Behrens,
K. Bokeloh,
E. Brown,
J. M. R. Cardoso,
B. Choi,
D. Cline,
S. Fattori,
A. D. Ferella,
K. L. Giboni,
A. Kish,
C. W. Lam,
J. Lamblin,
R. F. Lang,
K. E. Lim,
J. A. M. Lopes,
T. Marrodan Undagoitia,
Y. Mei,
A. J. Melgarejo Fernandez,
K. Ni
, et al. (16 additional authors not shown)
Abstract:
Results of the extensive radioactivity screening campaign to identify materials for the construction of XENON100 are reported. This Dark Matter search experiment is operated underground at Laboratori Nazionali del Gran Sasso (LNGS), Italy. Several ultra sensitive High Purity Germanium detectors (HPGe) have been used for gamma ray spectrometry. Mass spectrometry has been applied for a few low mass…
▽ More
Results of the extensive radioactivity screening campaign to identify materials for the construction of XENON100 are reported. This Dark Matter search experiment is operated underground at Laboratori Nazionali del Gran Sasso (LNGS), Italy. Several ultra sensitive High Purity Germanium detectors (HPGe) have been used for gamma ray spectrometry. Mass spectrometry has been applied for a few low mass plastic samples. Detailed tables with the radioactive contaminations of all screened samples are presented, together with the implications for XENON100.
△ Less
Submitted 22 June, 2011; v1 submitted 30 March, 2011;
originally announced March 2011.
-
Xenon Recirculation-Purification with a Heat Exchanger
Authors:
K. L. Giboni,
E. Aprile,
B. Choi,
T. Haruyama,
R. F. Lang,
K. E. Lim,
A. J. Melgarejo,
G. Plante
Abstract:
Liquid-xenon based particle detectors have been dramatically growing in size during the last years, and are now exceeding the one-ton scale. The required high xenon purity is usually achieved by continuous recirculation of xenon gas through a high-temperature getter. This challenges the traditional way of cooling these large detectors, since in a thermally well insulated detector, most of the cool…
▽ More
Liquid-xenon based particle detectors have been dramatically growing in size during the last years, and are now exceeding the one-ton scale. The required high xenon purity is usually achieved by continuous recirculation of xenon gas through a high-temperature getter. This challenges the traditional way of cooling these large detectors, since in a thermally well insulated detector, most of the cooling power is spent to compensate losses from recirculation. The phase change during recondensing requires five times more cooling power than cooling the gas from ambient temperature to -100C (173 K). Thus, to reduce the cooling power requirements for large detectors, we propose to use the heat from the purified incoming gas to evaporate the outgoing xenon gas, by means of a heat exchanger. Generally, a heat exchanger would appear to be only of very limited use, since evaporation and liquefaction occur at zero temperature difference. However, the use of a recirculation pump reduces the pressure of the extracted liquid, forces it to evaporate, and thus cools it down. We show that this temperature difference can be used for an efficient heat exchange process. We investigate the use of a commercial parallel plate heat exchanger with a small liquid xenon detector. Although we expected to be limited by the available cooling power to flow rates of about 2 SLPM, rates in excess of 12 SLPM can easily be sustained, limited only by the pump speed and the impedance of the flow loop. The heat exchanger operates with an efficiency of (96.8 +/- 0.5)%. This opens the possibility for fast xenon gas recirculation in large-scale experiments, while minimizing thermal losses.
△ Less
Submitted 4 March, 2011;
originally announced March 2011.
-
Lowering the low-energy threshold of xenon detectors
Authors:
P. Sorensen,
J. Angle,
E. Aprile,
F. Arneodo,
L. Baudis,
A. Bernstein,
A. Bolozdynya,
P. Brusov,
L. C. C. Coelho,
C. E. Dahl,
L. DeViveiros,
A. D. Ferella,
L. M. P. Fernandes,
S. Fiorucci,
R. J. Gaitskell,
K. L. Giboni,
R. Gomez,
R. Hasty,
L. Kastens,
J. Kwong,
J. A. M. Lopes,
N. Madden,
A. Manalaysay,
A. Manzur,
D. N. McKinsey
, et al. (12 additional authors not shown)
Abstract:
We show that the energy threshold for nuclear recoils in the XENON10 dark matter search data can be lowered to ~1 keV, by using only the ionization signal. In other words, we make no requirement that a valid event contain a primary scintillation signal. We therefore relinquish incident particle type discrimination, which is based on the ratio of ionization to scintillation in liquid xenon. This me…
▽ More
We show that the energy threshold for nuclear recoils in the XENON10 dark matter search data can be lowered to ~1 keV, by using only the ionization signal. In other words, we make no requirement that a valid event contain a primary scintillation signal. We therefore relinquish incident particle type discrimination, which is based on the ratio of ionization to scintillation in liquid xenon. This method compromises the detector's ability to precisely determine the z coordinate of a particle interaction. However, we show for the first time that it is possible to discriminate bulk events from surface events based solely on the ionization signal.
△ Less
Submitted 29 November, 2010;
originally announced November 2010.
-
Design and Performance of the XENON10 Dark Matter Experiment
Authors:
E. Aprile,
J. Angle,
F. Arneodo,
L. Baudis,
A. Bernstein,
A. Bolozdynya,
P. Brusov,
L. C. C. Coelho,
C. E. Dahl,
L. DeViveiros,
A. D. Ferella,
L. M. P. Fernandes,
S. Fiorucci,
R. J. Gaitskell,
K. L. Giboni,
R. Gomez,
R. Hasty,
L. Kastens,
J. Kwong,
J. A. M. Lopes,
N. Madden,
A. Manalaysay,
A. Manzur,
D. N. McKinsey,
M. E. Monzani
, et al. (14 additional authors not shown)
Abstract:
XENON10 is the first two-phase xenon time projection chamber (TPC) developed within the XENON dark matter search program. The TPC, with an active liquid xenon (LXe) mass of about 14 kg, was installed at the Gran Sasso underground laboratory (LNGS) in Italy, and operated for more than one year, with excellent stability and performance. Results from a dark matter search with XENON10 have been publ…
▽ More
XENON10 is the first two-phase xenon time projection chamber (TPC) developed within the XENON dark matter search program. The TPC, with an active liquid xenon (LXe) mass of about 14 kg, was installed at the Gran Sasso underground laboratory (LNGS) in Italy, and operated for more than one year, with excellent stability and performance. Results from a dark matter search with XENON10 have been published elsewhere. In this paper, we summarize the design and performance of the detector and its subsystems, based on calibration data using sources of gamma-rays and neutrons as well as background and Monte Carlo simulations data. The results on the detector's energy threshold, energy and position resolution, and overall efficiency show a performance that exceeds design specifications, in view of the very low energy threshold achieved (<10 keVr) and the excellent energy resolution achieved by combining the ionization and scintillation signals, detected simultaneously.
△ Less
Submitted 16 January, 2010;
originally announced January 2010.
-
Constraints on inelastic dark matter from XENON10
Authors:
J. Angle,
E. Aprile,
F. Arneodo,
L. Baudis,
A. Bernstein,
A. Bolozdynya,
P. Brusov,
L. C. C. Coelho,
C. E. Dahl,
L. DeViveiros,
A. D. Ferella,
L. M. P. Fernandes,
S. Fiorucci,
R. J. Gaitskell,
K. L. Giboni,
R. Gomez,
R. Hasty,
L. Kastens,
J. Kwong,
J. A. M. Lopes,
N. Madden,
A. Manalaysay,
A. Manzur,
D. N. McKinsey,
M. E. Monzani
, et al. (12 additional authors not shown)
Abstract:
It has been suggested that dark matter particles which scatter inelastically from detector target nuclei could explain the apparent incompatibility of the DAMA modulation signal (interpreted as evidence for particle dark matter) with the null results from CDMS-II and XENON10. Among the predictions of inelastically interacting dark matter are a suppression of low-energy events, and a population of…
▽ More
It has been suggested that dark matter particles which scatter inelastically from detector target nuclei could explain the apparent incompatibility of the DAMA modulation signal (interpreted as evidence for particle dark matter) with the null results from CDMS-II and XENON10. Among the predictions of inelastically interacting dark matter are a suppression of low-energy events, and a population of nuclear recoil events at higher nuclear recoil equivalent energies. This is in stark contrast to the well-known expectation of a falling exponential spectrum for the case of elastic interactions. We present a new analysis of XENON10 dark matter search data extending to E$_{nr}=75$ keV nuclear recoil equivalent energy. Our results exclude a significant region of previously allowed parameter space in the model of inelastically interacting dark matter. In particular, it is found that dark matter particle masses $m_χ\gtrsim150$ GeV are disfavored.
△ Less
Submitted 30 November, 2010; v1 submitted 19 October, 2009;
originally announced October 2009.
-
New Measurement of the Relative Scintillation Efficiency of Xenon Nuclear Recoils Below 10 keV
Authors:
E. Aprile,
L. Baudis,
B. Choi,
K. L. Giboni,
K. E. Lim,
A. Manalaysay,
M. E. Monzani,
G. Plante,
R. Santorelli,
M. Yamashita
Abstract:
Liquid xenon is an important detection medium in direct dark matter experiments, which search for low-energy nuclear recoils produced by the elastic scattering of WIMPs with quarks. The two existing measurements of the relative scintillation efficiency of nuclear recoils below 20 keV lead to inconsistent extrapolations at lower energies. This results in a different energy scale and thus sensitiv…
▽ More
Liquid xenon is an important detection medium in direct dark matter experiments, which search for low-energy nuclear recoils produced by the elastic scattering of WIMPs with quarks. The two existing measurements of the relative scintillation efficiency of nuclear recoils below 20 keV lead to inconsistent extrapolations at lower energies. This results in a different energy scale and thus sensitivity reach of liquid xenon dark matter detectors. We report a new measurement of the relative scintillation efficiency below 10 keV performed with a liquid xenon scintillation detector, optimized for maximum light collection. Greater than 95% of the interior surface of this detector was instrumented with photomultiplier tubes, giving a scintillation yield of 19.6 photoelectrons/keV electron equivalent for 122 keV gamma rays. We find that the relative scintillation efficiency for nuclear recoils of 5 keV is 0.14, staying constant around this value up to 10 keV. For higher energy recoils we measure a value around 20%, consistent with previously reported data. In light of this new measurement, the XENON10 experiment's results on spin-independent WIMP-nucleon cross section, which were calculated assuming a constant 0.19 relative scintillation efficiency, change from $8.8\times10^{-44}$ cm$^2$ to $9.9\times10^{-44}$ cm$^2$ for WIMPs of mass 100 GeV/c$^2$, and from $4.4\times10^{-44}$ cm$^2$ to $5.6\times10^{-44}$ cm$^2$ for WIMPs of mass 30 GeV/c$^2$.
△ Less
Submitted 16 May, 2010; v1 submitted 1 October, 2008;
originally announced October 2008.
-
The scintillation and ionization yield of liquid xenon for nuclear recoils
Authors:
P. Sorensen,
A. Manzur,
C. E. Dahl,
J. Angle,
E. Aprile,
F. Arneodo,
L. Baudis,
A. Bernstein,
A. Bolozdynya,
P. Brusov,
L. C. C. Coelho,
L. DeViveiros,
A. D. Ferella,
L. M. P. Fernandes,
S. Fiorucci,
R. J. Gaitskell,
K. L. Giboni,
R. Gomez,
R. Hasty,
L. Kastens,
J. Kwong,
J. A. M. Lopes,
N. Madden,
A. Manalaysay,
D. N. McKinsey
, et al. (12 additional authors not shown)
Abstract:
XENON10 is an experiment designed to directly detect particle dark matter. It is a dual phase (liquid/gas) xenon time-projection chamber with 3D position imaging. Particle interactions generate a primary scintillation signal (S1) and ionization signal (S2), which are both functions of the deposited recoil energy and the incident particle type. We present a new precision measurement of the relati…
▽ More
XENON10 is an experiment designed to directly detect particle dark matter. It is a dual phase (liquid/gas) xenon time-projection chamber with 3D position imaging. Particle interactions generate a primary scintillation signal (S1) and ionization signal (S2), which are both functions of the deposited recoil energy and the incident particle type. We present a new precision measurement of the relative scintillation yield \leff and the absolute ionization yield Q_y, for nuclear recoils in xenon. A dark matter particle is expected to deposit energy by scattering from a xenon nucleus. Knowledge of \leff is therefore crucial for establishing the energy threshold of the experiment; this in turn determines the sensitivity to particle dark matter. Our \leff measurement is in agreement with recent theoretical predictions above 15 keV nuclear recoil energy, and the energy threshold of the measurement is 4 keV. A knowledge of the ionization yield \Qy is necessary to establish the trigger threshold of the experiment. The ionization yield \Qy is measured in two ways, both in agreement with previous measurements and with a factor of 10 lower energy threshold.
△ Less
Submitted 30 December, 2008; v1 submitted 3 July, 2008;
originally announced July 2008.
-
Limits on spin-dependent WIMP-nucleon cross-sections from the XENON10 experiment
Authors:
J. Angle,
E. Aprile,
F. Arneodo,
L. Baudis,
A. Bernstein,
A. Bolozdynya,
P. Brusov,
L. C. C. Coelho,
C. E. Dahl,
L. DeViveiros,
A. D. Ferella,
L. M. P. Fernandes,
S. Fiorucci,
R. J. Gaitskell,
K. L. Giboni,
R. Gomez,
R. Hasty,
L. Kastens,
J. Kwong,
J. A. M. Lopes,
N. Madden,
A. Manalaysay,
A. Manzur,
D. N. McKinsey,
M. E. Monzani
, et al. (12 additional authors not shown)
Abstract:
XENON10 is an experiment to directly detect weakly interacting massive particle (WIMPs), which may comprise the bulk of the non-baryonic dark matter in our Universe. We report new results for spin-dependent WIMP-nucleon interactions with 129-Xe and 131-Xe from 58.6 live-days of operation at the Laboratori Nazionali del Gran Sasso (LNGS). Based on the non-observation of a WIMP signal in 5.4 kg of…
▽ More
XENON10 is an experiment to directly detect weakly interacting massive particle (WIMPs), which may comprise the bulk of the non-baryonic dark matter in our Universe. We report new results for spin-dependent WIMP-nucleon interactions with 129-Xe and 131-Xe from 58.6 live-days of operation at the Laboratori Nazionali del Gran Sasso (LNGS). Based on the non-observation of a WIMP signal in 5.4 kg of fiducial liquid xenon mass, we exclude previously unexplored regions in the theoretically allowed parameter space for neutralinos. We also exclude a heavy Majorana neutrino with a mass in the range of 10 GeV -2 TeV as a dark matter candidate under standard assumptions for its density and distribution in the galactic halo.
△ Less
Submitted 10 September, 2008; v1 submitted 19 May, 2008;
originally announced May 2008.
-
Compton Imaging of MeV Gamma-Rays with the Liquid Xenon Gamma-Ray Imaging Telescope (LXeGRIT)
Authors:
E. Aprile,
A. Curioni,
K. L. Giboni,
M. Kobayashi,
U. G. Oberlack,
S. Zhang
Abstract:
The Liquid Xenon Gamma-Ray Imaging Telescope (LXeGRIT) is the first realization of a liquid xenon time projection chamber for Compton imaging of MeV gamma-ray sources in astrophysics. By measuring the energy deposit and the three spatial coordinates of individual gamma-ray scattering points, the location of the source in the sky is inferred with Compton kinematics reconstruction. The angular res…
▽ More
The Liquid Xenon Gamma-Ray Imaging Telescope (LXeGRIT) is the first realization of a liquid xenon time projection chamber for Compton imaging of MeV gamma-ray sources in astrophysics. By measuring the energy deposit and the three spatial coordinates of individual gamma-ray scattering points, the location of the source in the sky is inferred with Compton kinematics reconstruction. The angular resolution is determined by the detector's energy and spatial resolutions, as well as by the separation in space between the first and second scattering. The imaging response of LXeGRIT was established with gamma-rays from radioactive sources, during calibration and integration at the Columbia Astrophysics Laboratory, prior to the 2000 balloon flight mission. In this paper we describe in detail the various steps involved in imaging sources with LXeGRIT and present experimental results on angular resolution and other parameters which characterize its performance as a Compton telescope.
△ Less
Submitted 2 May, 2008;
originally announced May 2008.
-
First Results from the XENON10 Dark Matter Experiment at the Gran Sasso National Laboratory
Authors:
J. Angle,
E. Aprile,
F. Arneodo,
L. Baudis,
A. Bernstein,
A. Bolozdynya,
P. Brusov,
L. C. C. Coelho,
C. E. Dahl,
L. DeViveiros,
A. D. Ferella,
L. M. P. Fernandes,
S. Fiorucci,
R. J. Gaitskell,
K. L. Giboni,
R. Gomez,
R. Hasty,
L. Kastens,
J. Kwong,
J. A. M. Lopes,
N. Madden,
A. Manalaysay,
A. Manzur,
D. N. McKinsey,
M. E. Monzani
, et al. (12 additional authors not shown)
Abstract:
The XENON10 experiment at the Gran Sasso National Laboratory uses a 15 kg xenon dual phase time projection chamber (XeTPC) to search for dark matter weakly interacting massive particles (WIMPs). The detector measures simultaneously the scintillation and the ionization produced by radiation in pure liquid xenon, to discriminate signal from background down to 4.5 keV nuclear recoil energy. A blind…
▽ More
The XENON10 experiment at the Gran Sasso National Laboratory uses a 15 kg xenon dual phase time projection chamber (XeTPC) to search for dark matter weakly interacting massive particles (WIMPs). The detector measures simultaneously the scintillation and the ionization produced by radiation in pure liquid xenon, to discriminate signal from background down to 4.5 keV nuclear recoil energy. A blind analysis of 58.6 live days of data, acquired between October 6, 2006 and February 14, 2007, and using a fiducial mass of 5.4 kg, excludes previously unexplored parameter space, setting a new 90% C.L. upper limit for the WIMP-nucleon spin-independent cross-section of 8.8 x 10^{-44} cm^2 for a WIMP mass of 100 GeV/c^2, and 4.5 x 10^{-44} cm^2 for a WIMP mass of 30 GeV/c^2. This result further constrains predictions of supersymmetric models.
△ Less
Submitted 3 December, 2007; v1 submitted 31 May, 2007;
originally announced June 2007.
-
Observation of Anti-correlation between Scintillation and Ionization for MeV Gamma-Rays in Liquid Xenon
Authors:
E. Aprile,
K. L. Giboni,
P. Majewski,
K. Ni,
M. Yamashita
Abstract:
A strong anti-correlation between ionization and scintillation signals produced by MeV gamma-rays in liquid xenon has been measured and used to improve the energy resolution by combining the two signals. The improvement is explained by reduced electron-ion recombination fluctuations of the combined signal compared to fluctuations of the individual signals. Simultaneous measurements of ionization…
▽ More
A strong anti-correlation between ionization and scintillation signals produced by MeV gamma-rays in liquid xenon has been measured and used to improve the energy resolution by combining the two signals. The improvement is explained by reduced electron-ion recombination fluctuations of the combined signal compared to fluctuations of the individual signals. Simultaneous measurements of ionization and scintillation signals were carried out with Cs-137, Na-22 and Co-60 gamma rays, as a function of electric field in the liquid. A resolution of 1.7%(sigma) at 662 keV was measured at 1 kV/cm, significantly better than the resolution from either scintillation or ionization alone. A detailed analysis indicates that further improvement to less than 1%(sigma) is possible with higher light collection efficiency and lower electronic noise.
△ Less
Submitted 9 April, 2007;
originally announced April 2007.
-
A Study of the LXeGRIT Detection Efficiency for MeV Gamma-Rays during the 2000 Balloon Flight Campaign
Authors:
A. Curioni,
E. Aprile,
T. Doke,
K. L. Giboni,
M. Kobayashi,
U. G. Oberlack
Abstract:
LXeGRIT - Liquid Xenon Gamma-Ray Imaging Telescope - is the first prototype of a Compton telescope for \MeV \g-ray astrophysics based on a LXe time projection chamber. One of the most relevant figures of merit for a Compton telescope is the detection efficiency for \g-rays, which depends on diverse contributions such as detector geometry and passive materials, trigger efficiency, dead time, etc.…
▽ More
LXeGRIT - Liquid Xenon Gamma-Ray Imaging Telescope - is the first prototype of a Compton telescope for \MeV \g-ray astrophysics based on a LXe time projection chamber. One of the most relevant figures of merit for a Compton telescope is the detection efficiency for \g-rays, which depends on diverse contributions such as detector geometry and passive materials, trigger efficiency, dead time, etc. A detailed study of the efficiency of the LXeGRIT instrument, based both on laboratory measurements and Monte Carlo simulations, is presented in this paper.
△ Less
Submitted 9 February, 2007; v1 submitted 9 February, 2007;
originally announced February 2007.
-
Gamma Ray Spectroscopy with Scintillation Light in Liquid Xenon
Authors:
K. Ni,
E. Aprile,
K. L. Giboni,
P. Majewski,
M. Yamashita
Abstract:
Scintillation light from gamma ray irradiation in liquid xenon is detected by two Hamamatsu R9288 photomultiplier tubes (PMTs) immersed in the liquid. UV light reflector material, PTFE, is used to optimize the light collection efficiency. The detector gives a high light yield of 6 photoelectron per keV (pe/keV), which allows efficient detection of the 122 keV gamma-ray line from Co-57, with a me…
▽ More
Scintillation light from gamma ray irradiation in liquid xenon is detected by two Hamamatsu R9288 photomultiplier tubes (PMTs) immersed in the liquid. UV light reflector material, PTFE, is used to optimize the light collection efficiency. The detector gives a high light yield of 6 photoelectron per keV (pe/keV), which allows efficient detection of the 122 keV gamma-ray line from Co-57, with a measured energy resolution of (8.8+/-0.6)% (sigma). The best achievable energy resolution, by removing the instrumental fluctuations, from liquid xenon scintillation light is estimated to be around 6-8% (sigma) for gamma-ray with energy between 662 keV and 122 keV.
△ Less
Submitted 4 August, 2006; v1 submitted 3 August, 2006;
originally announced August 2006.
-
Simultaneous Measurement of Ionization and Scintillation from Nuclear Recoils in Liquid Xenon as Target for a Dark Matter Experiment
Authors:
E. Aprile,
C. E. Dahl,
L. DeViveiros,
R. Gaitskell,
K. L. Giboni,
J. Kwong,
P. Majewski,
K. Ni,
T. Shutt,
M. Yamashita
Abstract:
We report the first measurements of the absolute ionization yield of nuclear recoils in liquid xenon, as a function of energy and electric-field. Independent experiments were carried out with two dual-phase time projection chamber prototypes, developed for the XENON Dark Matter project. We find that the charge yield increases with decreasing recoil energy, and exhibits only a weak field dependen…
▽ More
We report the first measurements of the absolute ionization yield of nuclear recoils in liquid xenon, as a function of energy and electric-field. Independent experiments were carried out with two dual-phase time projection chamber prototypes, developed for the XENON Dark Matter project. We find that the charge yield increases with decreasing recoil energy, and exhibits only a weak field dependence. These results are a first demonstration of the capability of dual phase xenon detectors to discriminate between electron and nuclear recoils, a key requirement for a sensitive dark matter search at recoil energies down to 20 keV.
△ Less
Submitted 24 January, 2006;
originally announced January 2006.
-
Scintillation Response of Liquid Xenon to Low Energy Nuclear Recoils
Authors:
E. Aprile,
K. L. Giboni,
P. Majewski,
K. Ni,
M. Yamashita,
R. Hasty,
A. Manzur,
D. N. McKinsey
Abstract:
Liquid Xenon (LXe) is expected to be an excellent target and detector medium to search for dark matter in the form of Weakly Interacting Massive Particles (WIMPs). Knowledge of LXe ionization and scintillation response to low energy nuclear recoils expected from the scattering of WIMPs by Xe nuclei is important for determining the sensitivity of LXe direct detection experiments. Here we report o…
▽ More
Liquid Xenon (LXe) is expected to be an excellent target and detector medium to search for dark matter in the form of Weakly Interacting Massive Particles (WIMPs). Knowledge of LXe ionization and scintillation response to low energy nuclear recoils expected from the scattering of WIMPs by Xe nuclei is important for determining the sensitivity of LXe direct detection experiments. Here we report on new measurements of the scintillation yield of Xe recoils with kinetic energy as low as 10 keV. The dependence of the scintillation yield on applied electric field was also measured in the range of 0 to 4 kV/cm. Results are in good agreement with recent theoretical predictions that take into account the effect of biexcitonic collisions in addition to the nuclear quenching effect.
△ Less
Submitted 29 March, 2005;
originally announced March 2005.
-
Performance of a Large Area Avalanche Photodiode in a Liquid Xenon Ionization and Scintillation Chamber
Authors:
K. Ni,
E. Aprile,
D. Day,
K. L. Giboni,
J. A. M. Lopes,
P. Majewski,
M. Yamashita
Abstract:
Scintillation light produced in liquid xenon (LXe) by alpha particles, electrons and gamma-rays was detected with a large area avalanche photodiode (LAAPD) immersed in the liquid. The alpha scintillation yield was measured as a function of applied electric field. We estimate the quantum efficiency of the LAAPD to be 45%. The best energy resolution from the light measurement at zero electric fiel…
▽ More
Scintillation light produced in liquid xenon (LXe) by alpha particles, electrons and gamma-rays was detected with a large area avalanche photodiode (LAAPD) immersed in the liquid. The alpha scintillation yield was measured as a function of applied electric field. We estimate the quantum efficiency of the LAAPD to be 45%. The best energy resolution from the light measurement at zero electric field is 7.5%(sigma) for 976 keV internal conversion electrons from Bi-207 and 2.6%(sigma) for 5.5 MeV alpha particles from Am-241. The detector used for these measurements was also operated as a gridded ionization chamber to measure the charge yield. We confirm that using a LAAPD in LXe does not introduce impurities which inhibit the drifting of free electrons.
△ Less
Submitted 14 February, 2005;
originally announced February 2005.
-
The XENON Dark Matter Search Experiment
Authors:
E. Aprile,
K. L. Giboni,
P. Majewski,
K. Ni,
M. Yamashita,
R. Gaitskell,
P. Sorensen,
L. DeViveiros,
L. Baudis,
A. Bernstein,
C. Hagmann,
C. Winant,
T. Shutt,
J. Kwong,
U. Oberlack,
D. McKinsey,
R. Hasty
Abstract:
The XENON experiment aims at the direct detection of dark matter in the form of WIMPs (Weakly Interacting Massive Particles) via their elastic scattering off Xe nuclei. A fiducial mass of 1000 kg, distributed in ten independent liquid xenon time projection chambers(LXeTPCs) will be used to probe thelowest interaction cross section predicted by SUSY models. The TPCs are operated in dual (liquid/g…
▽ More
The XENON experiment aims at the direct detection of dark matter in the form of WIMPs (Weakly Interacting Massive Particles) via their elastic scattering off Xe nuclei. A fiducial mass of 1000 kg, distributed in ten independent liquid xenon time projection chambers(LXeTPCs) will be used to probe thelowest interaction cross section predicted by SUSY models. The TPCs are operated in dual (liquid/gas) phase, to allow a measurement of nuclear recoils down to 16keV energy, via simultaneous detection of the ionization, through secondary scintillation in the gas, and primary scintillation in the liquid. Thedistinct ratio of primary to secondary scintillation for nuclear recoils from WIMPs (or neutrons), and for electron recoils from background, iskey to the event-by-event discrimination capability of XENON. A dual phase xenon prototype has been realized and is currently being tested, along with otherprototypes dedicated to other measurements relevant to the XENON program. As part of the R&D phase, we will realize and move underground a first XENON module (XENON10) with at least 10 kg fiducial mass to measure the background rejection capability and to optimize the conditions for continuous and stable detector operation underground. We present some of the results from the ongoing R&D and summarize the expected performance of the 10 kg experiment, from MonteCarlo simulations. The main design features of the 100 kg detector unit(XENON100), with which we envisage to make up the 1 tonne sensitive mass ofXENON1T will also be presented.
△ Less
Submitted 28 July, 2004;
originally announced July 2004.
-
The LXeGRIT Compton Telescope Prototype: Current Status and Future Prospects
Authors:
E. Aprile,
A. Curioni,
K. L. Giboni,
M. Kobayashi,
U. G. Oberlack,
E. L. Chupp,
P. P. Dunphy,
T. Doke,
J. Kikuchi,
S. Ventura
Abstract:
LXeGRIT is the first prototype of a novel concept of Compton telescope, based on the complete 3-dimensional reconstruction of the sequence of interactions of individual gamma rays in one position sensitive detector. This balloon-borne telescope consists of an unshielded time projection chamber with an active volume of 400 cm$^2 \times 7$ cm filled with high purity liquid xenon. Four VUV PMTs det…
▽ More
LXeGRIT is the first prototype of a novel concept of Compton telescope, based on the complete 3-dimensional reconstruction of the sequence of interactions of individual gamma rays in one position sensitive detector. This balloon-borne telescope consists of an unshielded time projection chamber with an active volume of 400 cm$^2 \times 7$ cm filled with high purity liquid xenon. Four VUV PMTs detect the fast xenon scintillation light signal, providing the event trigger. 124 wires and 4 anodes detect the ionization signals, providing the event spatial coordinates and total energy. In the period 1999 -- 2001, LXeGRIT has been extensively tested both in the laboratory and at balloon altitude, and its response in the MeV region has been thoroughly characterized. Here we summarize some of the results on pre-flight calibration, event reconstruction techniques, and performance during a 27 hour balloon flight on October 4 -- 5. We further present briefly the on-going efforts directed to improve the performance of this prototype towards the requirements for a base module of a next-generation Compton telescope.
△ Less
Submitted 4 December, 2002; v1 submitted 30 November, 2002;
originally announced December 2002.
-
On the Background Rate in the LXeGRIT Instrument during the 2000 Balloon Flight
Authors:
A. Curioni,
E. Aprile,
K. L. Giboni,
M. Kobayashi,
U. G. Oberlack,
E. L. Chupp,
P. P. Dunphy,
T. Doke,
J. Kikuchi,
S. Ventura
Abstract:
LXeGRIT is the first prototype of a novel Compton telescope for MeV gamma-ray astrophysics based on a Liquid Xenon Time Projection Chamber (LXeTPC), sensitive in the energy band of 0.15-10 MeV. In this homogeneous, 3D position sensitive detector, gamma rays with at least two interactions in the sensitive volume of 2800 cm$^{3}$, are imaged as in a standard Compton telescope. Gamma-rays with a si…
▽ More
LXeGRIT is the first prototype of a novel Compton telescope for MeV gamma-ray astrophysics based on a Liquid Xenon Time Projection Chamber (LXeTPC), sensitive in the energy band of 0.15-10 MeV. In this homogeneous, 3D position sensitive detector, gamma rays with at least two interactions in the sensitive volume of 2800 cm$^{3}$, are imaged as in a standard Compton telescope. Gamma-rays with a single interaction cannot be imaged and constitute a background which can be easily identified and rejected. Charged particles and localized beta-particles background is also easily suppressed based on the TPC localization capability with millimeter resolution. A measurement of the total gamma-ray background rate in near space conditions and the background rejection power of the LXeTPC was a primary goal of the LXeGRIT balloon flight program. We present here a preliminary analysis addressing this question, based on balloon flight data acquired during the Oct 4-5, 2000 LXeGRIT balloon flight from Ft. Sumner, NM. In this long duration (27 hr) balloon experiment, the LXeGRIT TPC was not surrounded by any gamma-ray or charged particle shield. Single site events and charged particles were mostly rejected on-line at the first and second trigger level. The remaining count rate of single-site \g-ray events, at an average atmospheric depth of 3.2 g cm$^{-2}$, is consistent with that expected from atmospheric and diffuse gamma-ray background, taking into account the instrument mass model and response.
△ Less
Submitted 4 December, 2002; v1 submitted 26 November, 2002;
originally announced November 2002.
-
Compton scattering sequence reconstruction algorithm for the liquid xenon gamma-ray imaging telescope (LXeGRIT)
Authors:
U. G. Oberlack,
E. Aprile,
A. Curioni,
V. Egorov,
K. L. Giboni
Abstract:
The Liquid Xenon Gamma-Ray Imaging Telescope (LXeGRIT) is a balloon born experiment sensitive to \g -rays in the energy band of 0.2-20 MeV. The main detector is a time projection chamber filled with high purity liquid xenon (LXeTPC), in which the three-dimensional location and energy deposit of individual \g -ray interactions are accurately measured in one homogeneous volume. To determine the \g…
▽ More
The Liquid Xenon Gamma-Ray Imaging Telescope (LXeGRIT) is a balloon born experiment sensitive to \g -rays in the energy band of 0.2-20 MeV. The main detector is a time projection chamber filled with high purity liquid xenon (LXeTPC), in which the three-dimensional location and energy deposit of individual \g -ray interactions are accurately measured in one homogeneous volume. To determine the \g -ray initial direction (Compton imaging), as well as to reject background, the correct sequence of interactions has to be determined. Here we report the development and optimization of an algorithm to reconstruct the Compton scattering sequence and show its performance on Monte Carlo events and LXeGRIT data.
△ Less
Submitted 13 December, 2000;
originally announced December 2000.