Characterization and ex vivo application of flexible 2D scintillating coatings in ultra-high dose rate electron beams for FLASH radiotherapy
Authors:
Verdi Vanreusel,
Stephen Brown,
Shujat Ali,
Thomas De Kerf,
Anthony J. Doemer,
Paul Leblans,
Benjamin Movsas,
Humza Nusrat,
Behzad Shirmard,
Kundan Thind,
Steve Vanlanduit,
Dirk Verellen,
Alessia Gasparini,
Luana de Freitas Nascimento
Abstract:
The increasing interest in FLASH-RT has lead to the conversion of linear accelerators to enable ultra-high dose rate (UHDR) beams for preclinical research. Dosimetry hereof remains challenging with several crucial aspects missing. This work shows the challenges for real-time 2D UHDR dosimetry, and aims to present a solution in the context of preclinical irradiations in non-homogeneous UHDR electro…
▽ More
The increasing interest in FLASH-RT has lead to the conversion of linear accelerators to enable ultra-high dose rate (UHDR) beams for preclinical research. Dosimetry hereof remains challenging with several crucial aspects missing. This work shows the challenges for real-time 2D UHDR dosimetry, and aims to present a solution in the context of preclinical irradiations in non-homogeneous UHDR electron beams. An experimental camera-scintillation sheet combination, was used to investigate the spatial dose distribution of a converted UHDR Varian Trilogy. The dosimetric system was characterized by variation of the number of pulses and source to surface distance (SSD) and its application was investigated by variation of bolus thickness and ambient light intensity. The challenges of prelcinical real time 2D dosimetry with scintillating coatings were assessed by ex vivo irradiations of a rat brain, mouse hindlimb and whole body mouse. Radiochromic EBT XD film was used as passive reference dosimeter. The coating showed a linear response with the number of pulses in the absence and presence of transparent bolus, up to 3 cm thick, and with the inverse squared SSD. The presence of ambient light reduces the signal-background ratio. The sheet showed to have sufficient flexibility to be molded on the subjects' surface, following its curvatures. Linearity with number of pulses was preserved in a preclinical setting. For small field sizes the light output became too low, resulting in noisy dose maps. The system showed robust within 5% for camera set up differences. Calibration of the system was complicated due to set up variations and the inhomogeneity of the beam. We showed the need for 2D real-time dosimetry to determine beam characteristics in non-homogeneous UHDR beams using a preclinical setting. We presented one solution to meet this need with scintillating based dosimetry.
△ Less
Submitted 22 April, 2025;
originally announced April 2025.
-
Oil Spill Drone: A Dataset of Drone-Captured, Segmented RGB Images for Oil Spill Detection in Port Environments
Authors:
T. De Kerf,
S. Sels,
S. Samsonova,
S. Vanlanduit
Abstract:
The high incidence of oil spills in port areas poses a serious threat to the environment, prompting the need for efficient detection mechanisms. Utilizing automated drones for this purpose can significantly improve the speed and accuracy of oil spill detection. Such advancements not only expedite cleanup operations, reducing environmental harm but also enhance polluter accountability, potentially…
▽ More
The high incidence of oil spills in port areas poses a serious threat to the environment, prompting the need for efficient detection mechanisms. Utilizing automated drones for this purpose can significantly improve the speed and accuracy of oil spill detection. Such advancements not only expedite cleanup operations, reducing environmental harm but also enhance polluter accountability, potentially deterring future incidents. Currently, there's a scarcity of datasets employing RGB images for oil spill detection in maritime settings. This paper presents a unique, annotated dataset aimed at addressing this gap, leveraging a neural network for analysis on both desktop and edge computing platforms. The dataset, captured via drone, comprises 1268 images categorized into oil, water, and other, with a convolutional neural network trained using an Unet model architecture achieving an F1 score of 0.71 for oil detection. This underscores the dataset's practicality for real-world applications, offering crucial resources for environmental conservation in port environments.
△ Less
Submitted 28 February, 2024;
originally announced February 2024.
-
An innovative dose rate independent 2D Ce-doped YAG scintillating dosimetry system for time resolved beam monitoring in ultra-high dose rate electron FLASH radiation therapy
Authors:
Verdi Vanreusel,
Sophie Heinrich,
Thomas De Kerf,
Paul Leblans,
Dirk Vandenbroucke,
Steve Vanlanduit,
Dirk Verellen,
Alessia Gasparini,
Luana de Freitas Nascimento
Abstract:
FLASH-RT has proven beneficial in preclinical studies. However, the lack of accurate real time 2D dosimetry is a limiting factor. In this work, an innovative solution for 2D real time dosimetry in UHDR electron beams is presented. The in-house developed ImageDosis system consists of a scientific camera, with high temporal resolution, and a coating, containing 12% of Y$_3$Al$_5$O$_{12}$:Ce$^{3+}$ a…
▽ More
FLASH-RT has proven beneficial in preclinical studies. However, the lack of accurate real time 2D dosimetry is a limiting factor. In this work, an innovative solution for 2D real time dosimetry in UHDR electron beams is presented. The in-house developed ImageDosis system consists of a scientific camera, with high temporal resolution, and a coating, containing 12% of Y$_3$Al$_5$O$_{12}$:Ce$^{3+}$ as scintillating material. Reference dosimetry was performed by means of radiochromic film, and a (C$_{38}$H$_{34}$P$_2$)MnBr$_4$ point scintillator was used to validate the pulse discrimination properties of the ImageDosis system. Irradiations were performed in two centers (Antwerp and Orsay), with an ElectronFlash accelerator, with varying number of pulses, pulse length, pulse repetition frequency (PRF) and energy. Also, the temporal resolution and 2D properties were investigated. For doses > 3.5 Gy, the ImageDosis showed a linear dose response up to at least 13 Gy. No dose rate dependence was found for an average dose rate up to 140 Gy/s, a dose per pulse up to 2 Gy and a PRF up to 300 Hz. The ImageDosis system showed capable of measuring the dose of the individual pulses up to a PRF of 250 Hz, but did not detect 3% of the pulses, because these pulses were delivered during the dead time of the camera. The maximal difference in FWHM of the field size between the ImageDosis system and the reference was 3.6%. For a nominal field size of 100 mm and 120 mm, a decreased output was observed on the superior part of the field. The ImageDosis system showed linear dose response and no dose rate nor energy dependence. It showed capable of discriminating and measuring the dose of individual pulses and promising 2D characteristics that need further optimization.
△ Less
Submitted 29 January, 2024; v1 submitted 21 November, 2023;
originally announced November 2023.
-
A Multisensor Hyperspectral Benchmark Dataset For Unmixing of Intimate Mixtures
Authors:
Bikram Koirala,
Behnood Rasti,
Zakaria Bnoulkacem,
Andrea de Lima Ribeiro,
Yuleika Madriz,
Erik Herrmann,
Arthur Gestels,
Thomas De Kerf,
Sandra Lorenz,
Margret Fuchs,
Koen Janssens,
Gunther Steenackers,
Richard Gloaguen,
Paul Scheunders
Abstract:
Optical hyperspectral cameras capture the spectral reflectance of materials. Since many materials behave as heterogeneous intimate mixtures with which each photon interacts differently, the relationship between spectral reflectance and material composition is very complex. Quantitative validation of spectral unmixing algorithms requires high-quality ground truth fractional abundance data, which ar…
▽ More
Optical hyperspectral cameras capture the spectral reflectance of materials. Since many materials behave as heterogeneous intimate mixtures with which each photon interacts differently, the relationship between spectral reflectance and material composition is very complex. Quantitative validation of spectral unmixing algorithms requires high-quality ground truth fractional abundance data, which are very difficult to obtain. In this work, we generated a comprehensive laboratory ground truth dataset of intimately mixed mineral powders. For this, five clay powders (Kaolin, Roof clay, Red clay, mixed clay, and Calcium hydroxide) were mixed homogeneously to prepare 325 samples of 60 binary, 150 ternary, 100 quaternary, and 15 quinary mixtures. Thirteen different hyperspectral sensors have been used to acquire the reflectance spectra of these mixtures in the visible, near, short, mid, and long-wavelength infrared regions (350-15385) nm. {\color{black} Overlaps in wavelength regions due to the operational ranges of each sensor} and variations in acquisition conditions {\color{black} resulted in} a large amount of spectral variability. Ground truth composition is given by construction, but to verify that the generated samples are sufficiently homogeneous, XRD and XRF elemental analysis is performed. We believe these data will be beneficial for validating advanced methods for nonlinear unmixing and material composition estimation, including studying spectral variability and training supervised unmixing approaches. The datasets can be downloaded from the following link: https://github.com/VisionlabUA/Multisensor_datasets.
△ Less
Submitted 30 August, 2023;
originally announced September 2023.
-
Corrosion Monitoring On Zinc Electroplated Steel Using Shortwave Infrared Hyperspectral Imaging
Authors:
Thomas De Kerf,
Zohreh Zahiri,
Paul Scheunders,
Steve Vanlanduit
Abstract:
In this study, we investigate the use of hyperspectral imaging (HSI) to inspect the formation of corrosion products on galvanised carbon steel samples. Ten samples were subjected to an accelerated corrosion test with different exposure times. The analysis is performed in a two-step procedure: First, the different corrosion minerals are identified by microscopic Fourier transform infrared spectrosc…
▽ More
In this study, we investigate the use of hyperspectral imaging (HSI) to inspect the formation of corrosion products on galvanised carbon steel samples. Ten samples were subjected to an accelerated corrosion test with different exposure times. The analysis is performed in a two-step procedure: First, the different corrosion minerals are identified by microscopic Fourier transform infrared spectroscopy (FTIR) at specific locations on the samples. The following corrosion minerals are identified: ZnO (zincite/zinc oxide) Zn5(OH)8Cl2 H2O (simonkolleite), ZnCO3 (smithsonite), Zn5(CO3)2(OH)6 (marionite/ hydrozincite). Second, the identified corrosion minerals are correlated with the HSI spectra for these specific locations. This correlation provides us with the spectra in the SWIR region and allows us to construct a classification map for the different corrosion minerals. The results show that we are able to identify the different minerals using HSI camera. This proposed methodology allows us to speed up the inspection process, compared to FTIR, while still accurately distinguishing between the different corrosion minerals.
△ Less
Submitted 23 June, 2022;
originally announced June 2022.