Commissioning, characterization and first high dose rate irradiations at a compact X-ray tube for microbeam and minibeam radiation therapy
Authors:
Christian Petrich,
Johanna Winter,
Anton Dimroth,
Thomas Beiser,
Monika Dehn,
Jessica Stolz,
Jacopo Frignani,
Stephanie E. Combs,
Franz Schilling,
Ghaleb Natour,
Kurt Aulenbacher,
Thomas E. Schmid,
Jan J. Wilkens,
Stefan Bartzsch
Abstract:
Minibeam and microbeam radiation therapy promise improved treatment outcomes through reduced normal tissue toxicity at better tumor control rates. The lack of suitable compact radiation sources limits the clinical application of minibeams to superficial tumors and renders it impossible for microbeams. We developed and constructed the first prototype of a compact line-focus X-ray tube (LFXT) with t…
▽ More
Minibeam and microbeam radiation therapy promise improved treatment outcomes through reduced normal tissue toxicity at better tumor control rates. The lack of suitable compact radiation sources limits the clinical application of minibeams to superficial tumors and renders it impossible for microbeams. We developed and constructed the first prototype of a compact line-focus X-ray tube (LFXT) with technology potentially suitable for clinical translation of minibeams and microbeams. We give an overview of the commissioning process preceding the first operation, present optical and radiological focal spot characterization methods, and dosimetric measurements. Additionally, we report on first preclinical in vitro cell and in vivo mouse brain irradiations conducted with the LFXT prototype. The focal spot characterization resulted in a strongly eccentric electron distribution with a width of 72.3 $μ$m. Dosimetry showed sharp microbeam dose profiles with steep lateral penumbras and a peak-to-valley dose ratio above 10 throughout a 70 mm thick PMMA phantom. An open-field dose rate of 4.3 Gy/s was measured at an acceleration voltage of 150 kV and a beam current of 17.4 mA at 150 mm distance from the focal spot. In vitro and in vivo experiments demonstrated the feasibility of the LFXT for minibeam and microbeam applications with field sizes of 1.5-2 cm. The mice displayed no observable side effects throughout the follow-up period after whole-brain 260 $μ$m-minibeam irradiation. We successfully constructed and commissioned the first proof-of-concept LFXT prototype. Dosimetric characterizations of the achieved microbeam field showed the superiority of the LFXT compared to conventional X-ray tubes in terms of beam quality. In future developments, the remaining limitations of the prototype will be addressed for improved minibeam and first ever microbeam radiation therapy in a clinical setting.
△ Less
Submitted 11 June, 2025;
originally announced June 2025.
-
Characterization of a novel time-resolved, real-time scintillation dosimetry system for ultra-high dose rate radiation therapy applications
Authors:
Alexander Baikalov,
Daline Tho,
Kevin Liu,
Stefan Bartzsch,
Sam Beddar,
Emil Schüler
Abstract:
Background: Scintillation dosimetry has promising qualities for ultra-high dose rate (UHDR) radiotherapy (RT), but no system has shown compatibility with mean dose rates ($\bar{DR}$) above 100 Gy/s and doses per pulse ($D_p$) exceeding 1.5 Gy typical of UHDR (FLASH)-RT. The aim of this study was to characterize a novel scintillator dosimetry system with the potential of accommodating UHDRs. Method…
▽ More
Background: Scintillation dosimetry has promising qualities for ultra-high dose rate (UHDR) radiotherapy (RT), but no system has shown compatibility with mean dose rates ($\bar{DR}$) above 100 Gy/s and doses per pulse ($D_p$) exceeding 1.5 Gy typical of UHDR (FLASH)-RT. The aim of this study was to characterize a novel scintillator dosimetry system with the potential of accommodating UHDRs. Methods: A thorough dosimetric characterization of the system was performed on an UHDR electron beamline. The system's response as a function of dose, $\bar{DR}$, $D_p$, and the pulse dose rate ${DR}_p$ was investigated, together with the system's dose sensitivity (signal per unit dose) as a function of dose history. The capabilities of the system for time-resolved dosimetric readout were also evaluated. Results: Within a tolerance of $\pm$3% the system exhibited dose linearity and was independent of $\bar{DR}$ and $D_p$ within the tested ranges of 1.8-1341 Gy/s and 0.005-7.68 Gy, respectively. A 6% reduction in the signal per unit dose was observed as ${DR}_p$ was increased from 8.9e4-1.8e6 Gy/s. Additionally, the dose delivered per integration window of the continuously sampling photodetector had to remain between 0.028 and 11.64 Gy to preserve a stable signal response per unit dose. The system accurately measured $D_p$ of individual pulses delivered at up to 120 Hz. The day-to-day variation of the signal per unit dose at a reference setup varied by up to $\pm$13% but remained consistent (<$\pm$2%) within each day of measurements and showed no signal loss as a function of dose history. Conclusions: With daily calibrations and ${DR}_p$ specific correction factors, the system reliably provides real-time, millisecond-resolved dosimetric measurements of pulsed conventional and UHDR beams from typical electron linacs, marking an important advancement in UHDR dosimetry.
△ Less
Submitted 5 March, 2024;
originally announced March 2024.
-
Differential tissue sparing of FLASH ultra high dose rates: an {\it in-silico} study
Authors:
Ramin Abolfath,
Alexander Baikalov,
Sohrab Rahvar,
Alberto Fraile,
Stefan Bartzsch,
Emil Schüler,
Radhe Mohan
Abstract:
Purpose: To propose a theory for the differential tissue sparing of FLASH ultra high dose rates (UHDR) through inter-track reaction-diffusion mechanism. Methods: We calculate the time-evolution of particle track-structures using a system of coupled reaction-diffusion equations on a random network designed for molecular transport in porous and disordered media. The network is representative of the…
▽ More
Purpose: To propose a theory for the differential tissue sparing of FLASH ultra high dose rates (UHDR) through inter-track reaction-diffusion mechanism. Methods: We calculate the time-evolution of particle track-structures using a system of coupled reaction-diffusion equations on a random network designed for molecular transport in porous and disordered media. The network is representative of the intra- and inter-cellular diffusion channels in tissues. Spatial cellular heterogeneities over the scale of track spacing have been constructed by incorporating random fluctuations in the connectivity among network sites. Results: We demonstrate the occurrence of phase separation among the tracks as the complexity in intra- and inter-cellular structural increases. At the weak limit of disorder, such as in water and normal tissue, neighboring tracks melt into each other and form a percolated network of nonreactive species. In contrast, at the strong limit of disorder, tracks evolve individually like isolated islands with negligible inter-track overlap. Thus, the spatio-temporal correlation among the chemical domains decreases as the inter-cellular complexity of the tissue increases (e.g. from normal tissue to fractal-type malignant tissue). Conclusions: The differential sparing of FLASH UHDR in normal and tumor tissue may be explained by differences in inter- and intra-cellular structural complexities between the tissue types. The structural complexities of cancerous cells prevent clustering and chemical interaction of tracks, whereas this interaction prevails and thus leads to sparing in normal tissue.
△ Less
Submitted 7 October, 2022;
originally announced October 2022.
-
Modeling interspur interactions as a potential mechanism of the FLASH effect
Authors:
Alexander Baikalov,
Ramin Abolfath,
Radhe Mohan,
Emil Schüler,
Jan J. Wilkens,
Stefan Bartzsch
Abstract:
The mechanism responsible for the FLASH effect, normal tissue sparing by ultra-high dose rate (UHDR) irradiation with isoeffective tumor control compared to conventional dose rate (CDR) irradiation, remains undetermined. Here we investigate the contribution of interspur interactions (interactions between radiolytic species of individual particle tracks) to overall radiochemical interactions as a f…
▽ More
The mechanism responsible for the FLASH effect, normal tissue sparing by ultra-high dose rate (UHDR) irradiation with isoeffective tumor control compared to conventional dose rate (CDR) irradiation, remains undetermined. Here we investigate the contribution of interspur interactions (interactions between radiolytic species of individual particle tracks) to overall radiochemical interactions as a function of irradiation parameters, and suggest an increase in interspur interaction as a potential mechanism for tissue sparing in FLASH radiation therapy. We construct a model that analytically represents the spatiotemporal distribution of spurs in a target volume as a function of irradiation parameters (e.g. dose, dose rate, linear energy transfer), and quantifies the effect of interspur interactions on the ongoing radiochemistry. Spurs evolve under a simplified reaction-diffusion equation with parameters based on Monte Carlo simulations, and interspur interaction is quantified by calculating the expected values of interspur overlap in the target.
The model demonstrates that for any set of irradiation parameters, a minimum critical dose and dose rate are necessary to induce significant interspur interaction, and that interspur interactions correlate negatively with beam linear energy transfer at a fixed dose. The model suggests optimal beam parameters, including dose, dose rate, linear energy transfer, and pulse structure, to maximize interspur interactions. Depending on the rate of radical scavenging in the target, which limits interspur interaction, this model predicts that the irradiation parameters necessary to elicit the FLASH effect may coincide with an onset of significant interspur interactions, suggesting that interspur interaction may be the underlying mechanism of the FLASH effect.
△ Less
Submitted 25 July, 2022;
originally announced July 2022.
-
The effect of non-ionizing excitations on the diffusion of ion species and inter-track correlations in FLASH ultra-high dose rate radiotherapy
Authors:
Ramin Abolfath,
Alexander Baikalov,
Stefan Bartzsch,
Niayesh Afshordi,
Radhe Mohan
Abstract:
We present a microscopic mechanism that accounts for the outward burst of "cold" ion species (IS) in a high-energy particle track due to coupling with ''hot" non-ion species (NIS). IS refers to radiolysis products of ionized molecules, whereas NIS refers to non-ionized excitations of molecules in a medium. The interaction is mediated by a quantized field of acoustic phonons, a channel that allows…
▽ More
We present a microscopic mechanism that accounts for the outward burst of "cold" ion species (IS) in a high-energy particle track due to coupling with ''hot" non-ion species (NIS). IS refers to radiolysis products of ionized molecules, whereas NIS refers to non-ionized excitations of molecules in a medium. The interaction is mediated by a quantized field of acoustic phonons, a channel that allows conversion of thermal energy of NIS to kinetic energy of IS, a flow of heat from the outer to the inner core of the track structure. We demonstrate the coexistence of "hot" NIS with "cold" IS in the radiation track structures right after their generation. NIS, concentrated within nano-scales volumes wrapping around IS, are the main source of intensive heat-waves and the outward burst of IS due to femto-second time scale IS-NIS coupling. By comparing the transport of IS coupled to NIS with identical configurations of non-interacting IS in thermal equilibrium at room temperature, we demonstrate that the energy gain of IS due to the surrounding hot nanoscopic volumes of NIS significantly increases their effective diffusion constants. The much higher diffusion constants predicted in the present model suggest higher inter-track chemical reaction rates at FLASH-UHDR, as well as lower intra-track reaction rates.
△ Less
Submitted 19 January, 2022;
originally announced January 2022.
-
Deep Learning Based HPV Status Prediction for Oropharyngeal Cancer Patients
Authors:
Daniel M. Lang,
Jan C. Peeken,
Stephanie E. Combs,
Jan J. Wilkens,
Stefan Bartzsch
Abstract:
We investigated the ability of deep learning models for imaging based HPV status detection. To overcome the problem of small medical datasets we used a transfer learning approach. A 3D convolutional network pre-trained on sports video clips was fine tuned such that full 3D information in the CT images could be exploited. The video pre-trained model was able to differentiate HPV-positive from HPV-n…
▽ More
We investigated the ability of deep learning models for imaging based HPV status detection. To overcome the problem of small medical datasets we used a transfer learning approach. A 3D convolutional network pre-trained on sports video clips was fine tuned such that full 3D information in the CT images could be exploited. The video pre-trained model was able to differentiate HPV-positive from HPV-negative cases with an area under the receiver operating characteristic curve (AUC) of 0.81 for an external test set. In comparison to a 3D convolutional neural network (CNN) trained from scratch and a 2D architecture pre-trained on ImageNet the video pre-trained model performed best.
△ Less
Submitted 17 November, 2020;
originally announced November 2020.
-
Shape optimization in laminar flow with a label-guided variational autoencoder
Authors:
Stephan Eismann,
Stefan Bartzsch,
Stefano Ermon
Abstract:
Computational design optimization in fluid dynamics usually requires to solve non-linear partial differential equations numerically. In this work, we explore a Bayesian optimization approach to minimize an object's drag coefficient in laminar flow based on predicting drag directly from the object shape. Jointly training an architecture combining a variational autoencoder mapping shapes to latent r…
▽ More
Computational design optimization in fluid dynamics usually requires to solve non-linear partial differential equations numerically. In this work, we explore a Bayesian optimization approach to minimize an object's drag coefficient in laminar flow based on predicting drag directly from the object shape. Jointly training an architecture combining a variational autoencoder mapping shapes to latent representations and Gaussian process regression allows us to generate improved shapes in the two dimensional case we consider.
△ Less
Submitted 10 December, 2017;
originally announced December 2017.